AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
TIF50, TIF52 1
TIF50, TIF52
On order
WIKA · Temperature measurement · Family TIF

HART ® field temperature transmitter

Field temperature transmitters with HART® protocol, available in various field case variants for general use in process engineering.
Range
-270…1820 °C
Output
HART · 4-20mA
Accuracy
±0.03 % of measuring span
Protection
IP66
CERoHSEACATEXIECExREACHII 2G Ex db IIC T4/T5/T6 GbEx ia [ia Ga] IIC T4/T5/T6 GbEx ia IIIC T135 °C Db1 Ex d IIC T6 ... T40 Ex ia IIC T4/T5/T61 Ex ib [ia ] IIC T4/T5/T6DIP A20 Ta 120 °CDIP A21 Ta 120 °CEx d IIC T6...T4II 2D Ex ia IIC T135℃ DbDIP A20 Ta 120℃DIP A21 Ta 120℃
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The TIF series field temperature transmitters, consisting of a rugged field case, model T32 temperature transmitter and a model DIH display, have been designed for general use in process engineering.

They offer high accuracy, galvanic isolation and excellent protection against electromagnetic influences (EMI). Via HART ® protocol, the TIFxx is configurable (interoperable) with a variety of open configuration tools.

In addition to the different sensor types, e.g. sensors in accordance with DIN EN 60751, JIS C1606, DIN 43760, IEC 60584 or DIN 43710, customer-specific sensor characteristics can also be defined, through the input of value pairs (user-defined linearisation). Through the configuration of a sensor with redundancy (dual sensor), on a sensor failure it will automatically change over to the working sensor.

Furthermore, there is the possibility to activate sensor drift detection. With this, an error signalling occurs when the magnitude of the temperature difference between sensor 1 and sensor 2 exceeds a user-selectable value.

for further approvals see page 10

Field temperature transmitters models TIF50, TIF52

The field temperature transmitter also has additional sophisticated supervisory functionality such as monitoring of the sensor lead resistance and sensor-break detection in accordance with NAMUR NE89 as well as monitoring of the measuring range. Moreover, these transmitters have comprehensive cyclic self-monitoring functionality.

Via the display it is possible to show range alarms as well as MIN and MAX values.

The field temperature transmitter is available in various field case variants. Stainless steel and aluminium can be specified.

It can be mounted directly on a wall. A pipe mounting kit is also available for fitting to pipes with a diameter of 1 ... 2".

Special features

  • Setting of units and measuring range possible on site (only model TIF52)
  • Different hazardous area approvals
  • Duplex sensor, redundant measurement possible
  • Customer-specific characteristic curves programmable
Identification
ModelTIF50, TIF52
Product familyTIF
SubcategoryField temperature transmitter
Measuring range
Temperature ranges−200…+850 °C · −270…+1300 °C · −200…+500 °C · −60…+250 °C · −210…+1200 °C · −200…+900 °C · −270…+1000 °C · −270…+400 °C · −200…+600 °C · −50…+1768 °C · +0…+1820 °C
Accuracy
Accuracy class±0.03 % of measuring span
Electrical characteristics
Output signalHART · 4-20mA
Power supply14.5 ... 42 V DC
Electrical connection2-wire
Operating environment
Ambient temperature−60…+85 °C
IP ratingIP66
Weight1.5 kg
Certifications and approvals
Certifications
CERoHSEACATEXIECExREACH
Explosion protection
II 2G Ex db IIC T4/T5/T6 GbEx ia [ia Ga] IIC T4/T5/T6 GbEx ia IIIC T135 °C Db1 Ex d IIC T6 ... T40 Ex ia IIC T4/T5/T61 Ex ib [ia ] IIC T4/T5/T6DIP A20 Ta 120 °CDIP A21 Ta 120 °CATEXIECExEx d IIC T6...T4II(1)2G Ex ia[ia Ga]IIC T4/T5/T6 GbII 2D Ex ia IIC T135℃ DbDIP A20 Ta 120℃DIP A21 Ta 120℃
Additional details
Sensor typePt100, Pt(x), JPt100, Ni100, Type J, K, L, E, N, T, U, R, S, B, mV sensor
Sensor classIEC 60751:2008, JIS C1606:1989, DIN 43760:1987, DIN 43710:1985
Datasheet languagesEN, DE
Tables from datasheet
Field temperatureMax. configurable measuring range 1)Standardα valuesMinimum measuring span 14)Typical measuring deviation 2)Temperature coefficient per °C typical 3)
Resistance sensorPt100-200 ...+850 °CIEC 60751 :2008α = 0.0038510 K or 3.8Ω (greater value applies)≤ ±0.12 °C 5)≤ ±0.0094 °C 6) 7)
Pt(x) 4) 10 ... 1000-200 ...+850 °CIEC 60751:2008α = 0.00385≤ ±0.12 °C 5)≤ ±0.0094 °C 6) 7)
JPt100-200 ...+500 °CJIS C1606: 1989α = 0.003916≤ ±0.12 °C 5)≤ ±0.0094 °C 6) 7)
Ni100-60 ...+250 °CDIN 43760: 1987α = 0.00618≤ ±0.12 °C 5)≤ ±0.0094 °C 6) 7)
Resistance sensor0 ...8,370Ω4Ω≤ ±1.68Ω 8)≤ ±0.1584Ω 8)
Potentiometer 9)0 ...100%10%≤ 0.50% 10)≤ ±0.0100% 10)
Measuring current during measurementMax. 0.3 mA(Pt100)
Connection methods1 sensor 2-/4-/3-wire or 2 sensors 2-wire (for further information, please refer to 'Designation of connection terminals')
Max. lead resistance50 Ωeach wire, 3-/4-wire
Thermo - coupleType J (Fe-CuNi)-210 ... +1,200 °CIEC 60584-1: 199550 K or2mV (greater value applies)≤ ±0.91 °C 11)≤ ±0.0217 °C 7) 11)
Type K (NiCr-Ni)-270 ... +1,300 °CIEC 60584-1: 1995≤ ±0.98 °C 11)≤ ±0.0238 °C 7) 11)
Type L (Fe-CuNi)-200 ...+900 °CDIN 43760: 1987≤ ±0.91 °C 11)≤ ±0.0203 °C 7) 11)
Type E (NiCr-Cu)-270 ... +1,000 °CIEC 60584-1: 1995≤ ±0.91 °C 11)≤ ±0.0224 °C 7) 11)
Type N (NiCrSi-NiSi)-270 ... +1,300 °CIEC 60584-1: 1995≤ ±1.02 °C 11)≤ ±0.0238 °C 7) 11)
TypeT (Cu-CuNi)-270 ...+400 °CIEC 60584-1: 1995≤ ±0.92 °C 11)≤ ±0.0191 °C 7) 11)
Type U (Cu-CuNi)-200 ...+600 °CDIN 43710: 1985≤ ±0.92 °C 11)≤ ±0.0191 °C 7) 11)
Type R (PtRh-Pt)-50 ... +1,768 °CIEC 60584-1: 1995150 K≤ ±1.66 °C 11)≤ ±0.0338 °C 7) 11)
Type S (PtRh-Pt)-50 ... +1,768 °CIEC 60584-1: 1995150 K≤ ±1.66 °C 11)≤ ±0.0338 °C 7) 11)
Type B (PtRh-Pt)0 ... +1,820 °C 15)IEC 60584-1: 1995200 K≤ ±1.73 °C 11)≤ ±0.0500 °C 7) 12)
mVsensor-500 ...+1,800mV4mV≤ ±0.33mV 13)≤ ±0.0311mV 7) 13)
Connection methods1 sensor or 2 sensors (for further information, please refer to 'Designation of connection terminals')
Max. lead resistance5 kΩ each wire
Cold junction compensation, configurableinternal compensation or external with Pt100, with thermostat or off
Display, operating unitModel TIF50Model TIF52
Display principleLCD, rotatable in 10° steps
Measured value of display7-segment LCD, 5-digit, character size9mm
Bar graph20-segment LCD
Info line14-segment LCD, 6-digit, character size 5.5mm
Status indicators♥ :HART ® mode (signalling of HART ® parameter adoption) : Unit lock :Warnings or error messages
Indication range-9999 ... 99999
Measuring rateapprox. 4/s
Accuracy±0.1 %of measuring span±0.05 %of measuring span
Temperature coefficient±0.1 %of measuring span / 10 K
HART ® functionality
■ Access control-Secondary master
■ Automatically set parameters
■ Available commands-Unit, measuring range start/end, format, zero point, span, damping, polling address
■ Identified commandsGeneric mode: 1, 15, 35, 44Generic mode: 0, 1, 6, 15, 34, 35, 36, 37, 44
■ MultidropNot supportedMeasured values are automatically taken from the HART ® digital data and displayed
Rise time, damping, measuring rate
Rise time t 90Approx. 0.8 s
Damping , configurableOff ; configurable between 1 s and 60 s
Switch-on time (time to get the first measured value)Max. 15 s
Measuring rate 1)Measured value update approx. 3/s
Analogue output, output limits, signalling, insulation resistance
Analogue output , configurableLinear to temperature per IEC 60751 / JIS C1606 / DIN 43760 (for resistance sensors) or linear to temperature per IEC 584 / DIN 43710 (for thermocouples) 4 ... 20 mAor 20 ... 4 mA, 2-wire
Output limits , configurable per NAMUR NE43 customer-specifically adjustableLower limit 3.8 mA 3.6 ...4.0mAupper limit 20.5 mA 20.0 ...21.5mA
Current value for signalling, configurable per NAMUR NE43 Substitute valueDownscale < 3.6 mA (3.5 mA) 3.5 ...12.0mAupscale > 21.0 mA(21.5 mA) 12.0 ...23.0mA
In simulation mode, independent from input signal, simulation value configurable from 3.5 ...23.0mA
Load R A (without HART ® )R A ≤ (U B - 13.5 V) / 0.023 A with R A in Ωand U B inV
Load R A (with HART ® )R A ≤ (U B - 14.5 V) / 0.023 A with R A in Ωand U B inV
Insulation voltage (input to analogue output)AC 1,200V (50 Hz / 60 Hz); 1 s
Insulation specification to DIN EN 60664-1:2003Overvoltage category III
View all (15)
Explosion protection, power supply
ModelApprovalsPermissible ambient/storage temperature (in accordance with the relevant temperatureSafety-related maximum valuesPower supply U B (DC)
classes)Sensor (Connections 1 - 4)Current loop (Connections ±)
TIF50-S, TIF52-Swithout{-50} -40 ...+85 °C--14.5 ...42V
TIF50-F, TIF52-FFlameproof enclosure BVS 10 ATEX E 158 IECEx BVS 10.0103 II 2G Ex db IIC T4/T5/T6 Gb-40 ...+85 °C at T4 -40 ...+75 °C at T5 -40 ...+60 °C at T6-U M = 30V P M =2W14.5 ...30V
Ex db IIC T4/T5/T6 Gb
TIF50-F, TIF52-FFlameproof enclosure TC RU C-DE.BH02.B.00466/20 1 Ex d IIC T6 ...T4-60 2) / -40 ...+85 °C at T4 -60 2) / -40 ...+75 °C at T5 -60 2) / -40 ...+60 °C at T6-U M = 30V P M =2W14.5 ...30V
TIF50-I, TIF52-IIntrinsically safe equipment 1) BVS 16 ATEX E 112 X IECEx BVS 16.0075Xsee installation drawing in the operating instructions atsee installation drawing in the operating instructions at14.5 ...29V
II (1)2G II 2G Ex ia [ia Ga] IIC T4/T5/T6 Gb Ex ia IIC T4/T5/T6 Gb-40 ...+85 °C at T4 -40 ...+70 °C at T5 -40 ...+55 °C at T6www.wika.comwww.wika.com
II (1)2D II 2D Ex ia [ia Da] IIIC T135 °C Db Ex ia IIIC T135 °C Db-40 ...+40 °C (P i = 680 mW) -40 ...+70 °C (P i = 650 mW)
TIF50-I, TIF52-IIntrinsically safe equipment 1) TC RU C-DE.AЯ45.B.00918see installation drawing insee installation drawing in14.5 ...29V
0 Ex ia IIC T4/T5/T6 1 Ex ib [ia ] IIC T4/T5/T6-60 2) / -40 ...+85 °C at T4 -60 2) / -40 ...+70 °C at T5 -60 2) / -40 ...+55 °C at T6the operating instructions at www.wika.comthe operating instructions at www.wika.com
DIP A20Ta 120 °C DIP A21Ta 120 °C-60 2) / -40 ...+40 °C (P i = 680 mW) -60 2) / -40 ...+70 °C (P i = 650 mW)
Measuring deviation, temperature coefficient, long-term stability
Not measurable
Power supply effectNot measurable
Warm-up timeAfter approx. 5 minutes the instrument will function to the specifications (accuracy)
InputMeasuring deviation per DIN EN 60770, 23 °C ±3 KMean temperature coefficient (TC) for each 10 K change in ambient temperature in the range -40 ... +85 °C Lead resistance effectsLong-term stability after 1 year
■ Resistance thermometer Pt100/JPt100/ Ni100 1)-200 °C ≤ MV≤200 °C: ±0.10 K MV>200 °C: ±(0.1 K + 0.01 %|MW-200 K|) 2)±(0.06 K + 0.015%MV) 4-wire: no effect (0 to 50 Ωeach wire) 3-wire: ±0.02 Ω/ 10Ω 50 Ωeach wire)±60 mΩor 0.05 %of MV, greater value applies
■ Resistance sensor≤ 890 Ω: 0.053Ω 4) or 0.015%MV 5) ≤ 2,140 Ω: 0.128Ω 4) or 0.015%MV 5) ≤ 4,390 Ω: 0.263Ω 4) or 0.015%MV 5) ≤ 8,380 Ω: 0.503Ω 4) or 0.015%MV 5)(0 to 2-wire: Resistor of the ±(0.01 Ω+0.01%MV)
■ PotentiometerR part /R total is max.±0.5%±(0.1%MV)
■ Thermocouples Type E, J-150 °C < MV<0°C: ±(0.3 K + 0.2 %|MV|) MV>0°C: ±(0.3 K + 0.03%MV)Type E: MV>-150 °C: ±(0.1 K + 0.015 %|MV|) Type J: MV>-150 °C: ±(0.07 K + 0.02 %|MV|) 6 µV / 1,000Ω 6)±20 µV or 0.05 %of MV, greater value applies
Type T, U-150 °C < MV<0°C: ±(0.4 K + 0.2 %|MV|) MV>0°C: ±(0.4 K + 0.01%MV)-150 °C < MV<0°C: ±(0.07 K + 0.04%MV) MV>0°C: ±(0.07 K + 0.01%MV)
Type R, S50 °C < MV<400 °C: ±(1.45 K + 0.12 %|MV - 400 K|) 400 °C < MV<1,600 °C: ±(1.45 K + 0.01 %|MV - 400 K|)Type R: 50 °C < MV<1,600 °C: ±(0.3 K + 0.01 %|MV- 400 K|) Type S: 50 °C < MV<1,600 °C: ±(0.3 K + 0.015 %|MV- 400 K|)
Type B450 °C < MV<1,000 °C: ±(1.7 K + 0.2 %|MV - 1,000 K|) MV>1,000 °C: ±1.7 K450 °C < MV<1,000 °C: ±(0.4 K + 0.02 %|MV - 1,000 K|) MV>1,000 °C: ±(0.4 K + 0.005 %(MV- 1,000 K))
Type K-150 °C < MV<0°C: ±(0.4 K + 0.2 %|MV|) 0 °C < MV<1,300 °C:-150 °C < MV<1,300 °C: ±(0.1 K + 0.02 %|MV|)
Type L-150 °C < MV<0°C: ±(0.3 K + 0.1 %|MV|) MV>0°C:±(0.3 K + 0.03%MV)-150 °C < MV<0°C: ±(0.07 K + 0.02 %|MV|) MV>0°C:±(0.07 K + 0.015%MV)
Type N-150 °C < MV<0°C: ±(0.5 K + 0.2 %|MV|) MV>0°C:±(0.5 K + 0.03%MV)-150 °C < MV<0°C: ±(0.1 K + 0.05 %|MV|) MV>0°C:±(0.1 K + 0.02%MV)
■ mVsensor≤1,160 mV: 10 μV + 0.03 %|MV| >1,160 mV: 15 μV + 0.07% |MV|2 μV + 0.02 %|MV| 100 μV + 0.08 %|MV|
■ Cold junction 7)±0.8 K±0.1 K±0.2 K
Output±0.03 %of measuring span±0.03 %of measuring span±0.05 %of span
Pt100 / 4-wire / measuring range 0 ... 150 °C / ambient temperature 33 °C
Input Pt100, MV<200 °C±0.100 K
Output ±(0.03 %of 150 K)±0.045 K
TC 10 K - input ±(0.06 K + 0.015 %of 150 K)±0.083 K
TC 10 K - output ±(0.03 %of 150 K)±0.045 K
Measuring deviation (typical) √input² + output² +TC input ² +TC output ²±0.145 K
Measuring deviation (maximum) (input + output +TC input +TC output )±0.273 K
Thermocouple type K / measuring range 0 ... 400 °C / internal compensation (cold junction) / ambient temperature 23 °C
Input type K, 0 °C < MV<1,300 °C ±(0.4 K + 0.04 %of 400 K)±0.56 K
Cold junction ±0.8 K±0.80 K
Output ±(0.03 %of 400 K)±0.12 K
Measuring deviation (typical) √input² + cold junction² + output²±0.98 K
Measuring deviation (maximum) (input + cold junction + output)±1.48 K
Monitoring
Test current for sensor monitoring 1)Nom.20 µA during test cycle, otherwise 0 µA
Monitoring NAMUR NE89 (monitoring of input lead resistance)
■ Resistance thermometer (Pt100, 4-wire)R L1 + R L4 > 100 Ωwith hysteresis5Ω R L2 + R L3 > 100 Ωwith hysteresis5Ω
■ ThermocoupleR L1 + R L4 + R thermocouple > 10 kΩ with hysteresis 100Ω
Sensor break monitoringAlways active
Self-monitoringActive permanently, e.g.RAM/ROM test, logical program operating checks and validity check
Measuring range monitoringMonitoring of the set measuring range for upper/lower deviations Standard: Deactivated
Monitoring of input lead resistance (3-wire)Monitoring of the resistance difference between lead 3 and 4; an error will be indicated if there is a difference of > 0.5 Ωbetween leads 3 and 4
Field case
Material■ Aluminium, window from polycarbonate ■ Stainless steel, window from polycarbonate
ColourAluminium: Night blue, RAL 5022Stainless steel: Silver
Cable bushings3 x M20 x 1.5 or3x½NPT
Ingress protectionIP66
WeightAluminium: approx. 1.5 kgStainless steel: approx. 3.7 kg
DimensionsSee drawing
Ambient conditions
Ambient temperature-60 1) / -40 ...+85 °C
Functional area of the display-20 2) ...+70 °C
Climate class per IEC 654-1: 1993Cx (-20 ...+85 °C, 35 ... 85 %r.h., non-condensing)
Maximum permissible humidity93 %r.h.±3%
Vibration resistance per IEC 60068-2-6:20073 g
Shock resistance per IEC 68-2-27: 198730 g
Electromagnetic compatibility (EMC)EN 61326 emission (group 1, class B) and interference immunity (industrial application), and also NAMUR NE21
ModelDescriptionOrder number
Programming unit, model PU-H
VIATOR ® HART ® USBHART ® modem for USB interface11025166
VIATOR ® HART ® USB PowerXpress™HART ® modem for USB interface14133234
VIATOR ® HART ® RS-232HART ® modem for RS-232 interface7957522
VIATOR ® HART ® Bluetooth ® ExHART ® modem for Bluetooth interface, Ex11364254
Magnetic quick connector magWIK■ Replacement for crocodile clips and HART ® terminals ■ Fast, safe and tight electrical connection ■ For all configuration and calibration processes14026893
LogoDescriptionRegion
EU declaration of conformityEuropean Union
EMCdirective EN 61326 emission (group 1, class B) and interference immunity (industrial application)
LogoDescriptionRegion
EU declaration of conformityEuropean Union
ATEX directive Hazardous areas
IECEx Hazardous areasInternational
EACEurasian Economic Community
EMCdirective
Hazardous areas 1)
PAC Russia Metrology, measurement technologyRussia
PAC Kazakhstan Metrology, measurement technologyKazakhstan
-MChS Permission for commissioningKazakhstan
PAC Belarus Metrology, measurement technologyBelarus
PAC Ukraine Metrology, measurement technologyUkraine
DNOP - MakNIIUkraine
Mining
Hazardous areas
-PESO Hazardous areasIndia
Certificates
Certificates■ 2.2 test report
■ 3.1 inspection certificate
Calibration■ DAkkS calibration certificate