AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
TGT73.100 1
TGT73.100
On order
WIKA · Temperature measurement · Family TGT73

Gas-actuated thermometer with electrical output signal

Gas-actuated thermometer with integrated electrical output signal (4-20 mA or 0-10 V) for local indication and signal transmission, stainless steel case.
Range
-200…700 °C
Output
4-20mA · 0-10V
Accuracy
0.2 % of full scale value
Protection
IP65
RoHSEMCEN 13190CE4 ... 20 mA(Ex version)
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

Wherever the process temperature has to be indicated on-site and, at the same time, a signal transmission to the central control or remote centre is desired, the model TGT73 intelliTHERM ® can be used.

Through the combination of a mechanical measuring system and electronic signal processing, the process temperature can be read reliably, even if the voltage supply is lost.

Due to the wide variety of possible designs, the model TGT73 gas-actuated thermometers can be perfectly adapted to any process connection or location. With the adjustable stem and dial version, the case can be adjusted precisely to the desired viewing angle.

Special features

  • Cost-effective “2 in 1” temperature measurement
  • Compact design
  • Scale ranges from -200 ... +700 °C [-328 ... +1,292 °F]
  • Plug-and-play, thus no transmitter configuration necessary
Identification
ModelTGT73.100
Product familyTGT73
SubcategoryGas-actuated thermometer with electrical output
Measuring range
Temperature ranges−80…+60 °C · −60…+40 °C · −40…+60 °C · −30…+50 °C · −20…+80 °C · −20…+120 °C · −20…+140 °C · +0…+60 °C · +0…+80 °C · +0…+100 °C · +0…+120 °C · +0…+160 °C · +0…+200 °C · +0…+250 °C · +0…+300 °C · +0…+400 °C · +0…+500 °C · +0…+600 °C · +0…+700 °C · −20…+60 °C · −200…+700 °C · +50…+350 °C · +50…+450 °C · +100…+500 °C · +100…+600 °C
Accuracy
Accuracy class0.2 % of full scale value
Process conditions
Process connection
G ½ B½ NPTG ½ female½ NPT femaleM20 x 1.5M24 x 1.5 femaleG ¾B1/2NPT3/4NPTM18x1.5M24x1.5
Process temperature−200…+700 °C
Electrical characteristics
Output signal4-20mA · 0-10V · 4 ... 20 mA, 2-wire, passive, per NAMUR NE43 · 4 ... 20 mA (Ex version) · 0 ... 10 V, 3-wire
Power supplyDC 12 ≤ U ≤ 30 V
Electrical connectionAngular connector incl. strain relief, M20 x 1.5 cable gland, max. 1.5 mm²
Materials
Probe materialStainless steel 316
Operating environment
Ambient temperature−40…+60 °C
IP ratingIP65
Certifications and approvals
Certifications
RoHSEMCEN 13190CE
Explosion protection
Ex ia Zone 1 gas II 2G Ex ia IIC T6/T5/T4 GbEx ia Zone 1 gas Ex ia IIC T6/T5/T4 GbZone 20 dust II 2D Ex ia IIIB T85°C/T95°C/T100°C/T135°C DbEx ia IIIB T85°C/T95°C/T100°C/T135°C Db4 ... 20 mA(Ex version)Zone 20 dust II 2D Ex ia IIIB T85℃/T95℃/T100℃/T135℃ DbZone 20 dust Ex ia IIIB T85℃/T95℃/T100℃/T135℃ Db
Additional details
Sensor typebimetal
Sensor classClass 1 per EN 13190
Probe length200 mm
Probe diameter6 mm
ThermowellNo
Datasheet languagesEN
Tables from datasheet
Basic information
StandardEN 13190
Nominal size in mm[in]100 [4"]
WindowLaminated safety glass
Connection location■ Back mount (axial) ■ Lower mount (radial) ■ Back mount, adjustable stem and dial ■ Version with capillary
Connection design→ For drawings, see page 7
SStandard (threaded connection) 1)
1Plain stem (without thread)
2Male nut
3Union nut
4Compression fitting (sliding on stem)
5Union nut and loose threaded connection
6Compression fitting (can be adjusted on either capillary or spiral protective sleeve)
7Compression fitting at the case
'Adjustable stem and dial' case versionSwivelling 90° and rotatable 360°
Damping, case filling■ Without ■ With liquid damping
Reverse polarity protectionYes
Material (non-wetted)
Case, ringStainless steel 304
PointerAluminium, black
Articulated joint 'adjustable stem and dial'Stainless steel 304
DialAluminium
Measuring element
Type of measuring elementInert gas expansion system
Working range
Constant loading (1 year)Measuring range (EN 13190)
Short time (max. 24 h)Scale range (EN 13190)
Capillary
Diameter2 mm[0.08 in]
Spiral protective sleeve7 mm[0.28 in]
LengthCapillary without spiral protective sleeveMax. 60 m[196.85 in]
Capillary with spiral protective sleeveMax. 40 m[131.23 in]
Capillary with spiral protective sleeve and PVC coatingMax. 20 m[65.62 in]
Minimum bending radiusCapillary without spiral protective sleeve6 mm[0.24 in]
Capillary with spiral protective sleeve20 mm[0.79 in]
Capillary with spiral protective sleeve and PVC coating30 mm[1.18 in]
Material (non-wetted)Stainless steel 316
Mounting options■ Surface mounting flange, stainless steel ■ Instrument mounting bracket, aluminium die-casting ■ Panel mounting flange, stainless steel
Contact bulb
Dimensions120 x 22 x 12 mm[4.72 x 0.87 x 0.47 in]
Mounting types■ Mounting on pipes ■ Mounting on tanks
Material (non-wetted)Stainless steel 316
Accuracy specifications
Accuracy classClass 1 per EN 13190 at 23 °C ±10 °C ambient temperature
Accuracy of output signal0.2 %of full scale value (only electronics)
Influence of auxiliary power≤ 0.1 %of full scale/10V
Influence of load≤ 0.1 %of full scale
Temperature error (electronics)< 0.3 %of full scale value/10 K (in overall temperature range)
Long-term stability (electronics)< 0.3 %of full scale/a
Resolution0.15 %of full scale (10-bit resolution at 360°)
Scale range in °CMeasuring range 1) in °CScale interval in °CError limit in °C
-80 ... +60-60 ...+4022
-60 ... +40-50 ...+3011
-40 ... +60-30 ...+5011
-30 ... +50-20 ...+4011
-20 ... +60-10 ...+5011
-20 ... +80-10 ...+7011
-20 ... +1200 ... 10023
-20 ... +1400 ... 12023
0 ... 6010 ... 5011
0 ... 8010 ... 7011
0 ... 10010 ... 9011
0 ... 12010 ... 11022
0 ... 16020 ... 14022
0 ... 20020 ... 18022
0 ... 25030 ... 22052.5
0 ... 30030 ... 27055
View all (20)
Scale range in °CMeasuring range 1) in °CScale interval in °CError limit in °C
0 ... 40050 ... 35055
0 ... 50050 ... 45055
0 ... 600100 ... 5001010
0 ... 700100 ... 6001010
Output signal
Measuring rate (refresh rate)> 1/s
Voltage supply
Auxiliary power U B4 ...20mADC 12 ≤ U B ≤ 30V
4 ... 20 mA(Ex version)DC 14 ≤ U B ≤ 30V
0 ...10VDC 15 ≤ U B ≤ 30V
Permissible residual ripple of auxiliary power≤ 10 %ss
Impedance at voltage output0.5Ω
Load capacity2 …100 kΩ
Warm-up time≤ 5 min
Sampling rate sensor600 ms
Linear error≤ 1.0 %of span (terminal method)
Input signal, angle of rotation0 …270 ∢°
Electrical connection
Connection typeAngular connector (180° rotatable, max. 1.5 mm², cable protection, M20 x 1.5 cable gland, incl. strain relief)
Wire cross-sectionMax. 1.5 mm²
Outer cable diameter7 ... 13 mm[0.28 ... 0.51 in], see dimensions on page 19
Pin assignmentTerminal 1Terminal 2Terminal 3Terminal 4Terminal 4Terminal 6
Variant IGNDI +reservedreservedreservedreserved
Variant II (Ex version)GNDI +reservedreservedreservedreserved
Variant IIIGNDU B+U outreservedreservedreserved
MaterialPA 6 (polyamide)
Operating conditions
Ambient temperature range-40 …+60 °C [-4 ...+140 °F] without/with liquid damping
Storage temperature range-40 ...+60 °C [-40 ...+140 °F]
Max. operating pressure at stemMax. 25 bar [362.59 psi], static
Ingress protection (IP code) per IEC/EN 60529IP65
Safety-related characteristic values (Ex, variant II)
Max. auxiliary power U iDC 30V
Max. short-circuit current I i100mA
Max. power P i0.72W
Effective internal capacitance C i12 nF
Effective internal inductance L iNegligible
LogoDescriptionCountry
EU declaration of conformityEuropean Union
EMCdirective
RoHS directive
LogoDescriptionCountry
EU declaration of conformityEuropean Union
ATEX directive Hazardous areas
- Ex ia Zone1 gas II 2G Ex ia IIC T6/T5/T4 Gb
Zone20 dust II 2D Ex ia IIIB T85°C/T95°C/T100°C/T135°C
IECEx Hazardous areasInternational
- Ex ia Zone Zone1 gas Ex ia IIC T6/T5/T4 Gb 20 dust Ex ia IIIB T85°C/T95°C/T100°C/T135°C Db
EACEurasian Economic Community
EMCdirective
Hazardous areas
-MTSCHS Permission for commissioningKazakhstan
Certificates
Certificates■ 2.2 test report ■ 3.1 inspection certificate
CalibrationDAkkS calibration certificate
DesignF BM 1)Connection
S30 [1.18]G½-Male
113 [0.51]∅ 18
235 [1.38]G½-Male
315 [0.59]G1/2 - Female
453 [2.09]G½-Male
550 [1.97]G½-Male
753 [2.09]G½-Male
DesignF LM 1)Connection
S30 [1.18]G½-Male
113 [0.51]∅ 18
235 [1.38]G½-Male
315 [0.59]G1/2 - Female
453 [2.09]G½-Male
550 [1.97]G½-Male
753 [2.09]G½-Male
DesignF DSConnection
S17.5 [0.69]G½-Male
128 [1.10]∅ 18
238 [1.50]G½-Male
330 [1.18]G1/2 - Female
468 [2.68]G½-Male
68 [2.68]G½-Male
555 [2.68]G½-Male
768 [2.68]G½-Male
Process connectionDimensions in mm[in]
GiSWd 4
G½B14 [0.55]27 [1.06]26 [1.02]
G¾B16 [0.63]32 [1.26]32 [1.26]
½NPT19 [0.75]22 [0.87]-
¾NPT20 [0.79]30 [1.18]-
Process connectionDimensions in mm[in]
GiSW
G½B20 [0.79]27 [1.06]
M18 x 1.515 [0.59]22 [0.89]
Process connectionDimensions in mm[in]
Without threadØ d 1
-18 [0.7]
Process connec- tionDimensions in mm[in]
GiSW
G½B14 [0.55]27 [1.06]
G¾B16 [0.63]32 [1.26]
M24 x 1.513.5 [0.53]32 [1.26]
Process connec- tionDimensions in mm[in]
GiSWØ d 4
G½B14 [0.55]27 [1.06]26 [1.02]
G¾B16 [0.63]32 [1.26]32 [1.26]
M18 x 1.512 [0.47]24 [0.95]23 [0.91]
½NPT19 [0.75]22 [0.87]-
¾NPT20 [0.79]30 [1.18]-