Miniature resistance thermometer
Description
Resistance thermometers of these series are used as universal thermometers for the measurement of liquid and gaseous media in the range -50 ... +250 °C [-58 ... +482 °F].
They can be used for pressures up to 140 bar [2,030 psi] with 3 mm [0.12 in] protection tube diameters and up to 270 bar [3,916 psi] with 6 mm [0.24 in] protection tube diameters, depending on the instrument version. All electrical components are protected against humidity (IP67 or IP69K) and designed to withstand vibration (20 g, depending on instrument version).
The resistance thermometer is available with direct sensor output or integrated transmitter, which can be configured individually via the WIKAsoft-TT PC configuration software. Measuring range, dampening, error signalling per NAMUR NE 043 and tag no. can be adjusted.
for further approvals see page 6
Fig. left: Resistance thermometer, model TR33 Fig. right: M12 x 1 adapter to DIN EN 175301-803 angular connector
Insertion length, process connection, sensor and connection method can each be selected for the respective application within the ordering information. The model TR33 resistance thermometer consists of a protection tube with a fixed process connection and is screwed directly into the process. The electrical connection is made via an M12 x 1 circular connector. An adapter for electrical connection with angular connector per DIN EN 175301-803 is optionally available (patent, industrial property right: 001370985).
Special features
- Very compact design, high vibration resistance and fast response time
- With direct sensor output (Pt100, Pt1000 in 2-, 3- or 4-wire connection) or integrated transmitter with 4 ... 20 mA output signal
- Integrated transmitter is individually parameterisable with free-of-charge WIKAsoft-TT PC configuration software
- Sensor element with accuracy class A per IEC 60751
| Model | TR33 |
| Subcategory | Resistance thermometer |
| Temperature ranges | −50…+250 °C · −30…+250 °C · −20…+125 °C |
| Accuracy class | ±0.25 K |
| Process connection | G¼BG⅜BG½B¼NPT½NPTM12 x 1.5M20 x 1.57/16-20 UNF-2A |
| Process temperature | −50…+250 °C |
| Output signal | 4-20mA · HART |
| Power supply | DC 10 ... 30V |
| Electrical connection | M12 x 1 circular connector (4-pin) |
| Probe material | Stainless steel 1.4571 |
| Ambient temperature | −40…+85 °C |
| IP rating | IP67 |
| Certifications | CERoHSREACH |
| Sensor type | Pt100, Pt1000 |
| Sensor class | A, B |
| Probe diameter | 3 mm |
| Thermowell | No |
| Datasheet languages | EN, DE |
| Type of measuring element | ||
|---|---|---|
| Version 4 ... 20 mA(modelTR33-Z-TT) | Pt1000 (measuring current < 0.3 mA;self-heating can be ignored) | |
| Version Pt100 (model TR33-Z-Px) / Pt1000 (model TR33-Z-Sx) | ■ Pt100 (measuring current: 0.1 ... 1.0 mA) ■ Pt1000 (measuring current: 0.1 ... 0.3 mA) | |
| → For detailed specifications for Pt sensors, see Technical information IN 00.17 at www.wika.com. | ||
| Connection method | ||
| Version 4 ... 20 mA(modelTR33-Z-TT) | 2-wire | |
| Version Pt100 (model TR33-Z-Px) / Pt1000 (model TR33-Z-Sx) | 2-wire | The lead resistance is recorded as an error in the measurement |
| 3-wire | With a cable length of 30 mor longer, measuring deviations can occur | |
| 4-wire | The lead resistance can be ignored | |
| Tolerance value of the measuring element 1) per IEC 60751 | ||
| Version 4 ... 20 mA(modelTR33-Z-TT) | Class A | |
| Version Pt100 (model TR33-Z-Px) / Pt1000 (model TR33-Z-Sx) | ■ Class A ■ Class B at | 2-wire |
| Tolerance value of the measuring element 1) per IEC 60751 | Class A |
|---|---|
| Measuring deviation of the transmitter per IEC 62828 | ±0.25 K |
| Total measuring deviation per IEC 62828 | Measuring deviation of the measuring element + transmitter |
| Influence of the ambient temperature | 0.1 %of the set measuring span / 10 KT a |
| Influence of supply voltage | ±0.025 %/V(depending on the supply voltage U B ) |
| Influence of the load | ±0.05 %/ 100Ω |
| Linearisation | Linear to temperature per IEC 60751 |
| Output error | ±0.1% 2) |
| Reference conditions | |
| Ambient temperatureT a ref | 23 °C |
| Supply voltage U B ref | DC 12V |
| Temperature range | ||
|---|---|---|
| Version 4 ... 20 mA(modelTR33-Z-TT) | Without neck tube -30 ...+150 °C [-22 ...+302 °F] With neck tube -30 ...+250 °C [-22 ...+482 °F] 1) Version with FKM O-ring: -20 ...+125 °C [-4 ...+257 °F] | |
| Version Pt100 (model TR33-Z-Px) / Pt1000 (model TR33-Z-Sx) | Class A | Without neck tube -30 ...+150 °C [-22 ...+302 °F] With neck tube -30 ...+250 °C [-22 ...+482 °F] Version with FKM O-ring: -20 ...+125 °C [-4 ...+257 °F] |
| Class B | Without neck tube -50 ...+150 °C [-58 ...+302 °F] With neck tube -50 ...+250 °C [-58 ...+482 °F] | |
| Unit (4 ... 20 mA version) | Configurable °C, °F, K | |
| Temperature at the connector (Pt100, Pt1000 version) | Max. 85 °C [185 °F] | |
| Measuring span (4 ... 20 mA version) | Minimum 20 K, maximum 300 K |
| Process connection | |
|---|---|
| Type of process connection | ■ G¼B ■ G⅜B ■ G½B ■ ¼NPT ■ ½NPT ■ M12 x 1.5 ■ M20 x 1.5 ■ 7/16-20 UNF-2A |
| Protection tube | |
| Protection tube diameter | ■ 3 mm[0.12 in] ■ 6 mm[0.24 in] |
| Insertion length U 1 | ■ 50 mm[1.97 in] ■ 75 mm[2.95 in] 1) ■ 100 mm[3.94 in] 1) ■ 120 mm[4.72 in] 1) ■ 150 mm[5.91 in] 1) ■ 200 mm[7.87 in] 1) ■ 250 mm[9.84 in] 1) ■ 300 mm[11.81 in] 1) ■ 350 mm[13.78 in] 1) ■ 400 mm[15.75 in] 1) Other insertion lengths on |
| request | |
| Material (wetted) | Stainless steel 1.4571 |
View all (14)
| Output signal (4 ... 20 mA version) | |
|---|---|
| Analogue output | 4 ... 20 mA, 2-wire |
| Load R A | R A ≤ (U B - 10 V) / 23 mAwith R A in Ωand U B inV The permissible load depends on the loop supply voltage. For communication with the instrument with programming unit PU-548, a max. load of 350 Ωis admissible. |
| Load diagram | Load R A inΩ Voltage U B inV 0 10 24 30 36 1083 833 583 |
| Factory configuration | |
| Measuring range | Measuring range 0 ... 150 °C [32 ... 302 °F] Other measuring ranges are adjustable |
| Current signals for error signalling | Configurable in accordance with NAMUR NE 043 downscale ≤ 3.6mA upscale ≥ 21.0mA |
| Current value for sensor short-circuit | Not configurable in accordance with NAMUR NE 043 downscale ≤ 3.6mA |
| Communication | |
| Info data | Tag no., description and user message can be stored in transmitter |
| Configuration and calibration data | Permanently stored |
| Configuration software | WIKAsoft-TT → Configuration software (multilingual) as a download from www.wika.com |
| Voltage supply | |
| Supply voltage U B | DC 10 ...30V |
| Supply voltage input | Protected against reverse polarity |
| Permissible residual ripple of supply voltage | 10 %generated by U B < 3 %ripple of the output current |
| Time response | |
| Switch-on delay, electrical | Max. 4 s (time before the first measured value) |
| Warm-up time | After approx. 4 minutes, the instrument will function to the specifications (accuracy) given in the data sheet. |
| Connection type | M12 x 1 circular connector (4-pin) |
|---|---|
| Material | Stainless steel 1.4571 |
| Pin | Signal | Description |
|---|---|---|
| 1 | L+ | 10 ...30V |
| 2 | VQ | not connected |
| 3 | L- | 0V |
| 4 | C | not connected |
| Operating conditions | |
|---|---|
| Ambient temperature range | |
| Version 4 ... 20 mA(modelTR33-Z-TT) | -40 ...+85 °C [-40 ...+185 °F] Version with FKM O-ring: -20 °C [-4 °F] |
| Version Pt100 (model TR33-Z-Px) / Pt1000 (model TR33-Z-Sx) | -50 ...+85 °C [-58 ...+185 °F] Version with FKM O-ring: -20 °C [-4 °F] |
| Storage temperature range | -40 ...+85 °C [-40 ...+185 °F] Version with FKM O-ring: -20 °C [-4 °F] |
| Climate class per IEC 60654-1 | |
| Version 4 ... 20 mA(modelTR33-Z-TT) | Cx (-40 ...+85 °C [-40 ...+185 °F], 5 ... 95 %r.h.) Version with FKM O-ring: -20 °C [-4 °F] |
| Version Pt100 (model TR33-Z-Px) / Pt1000 (model TR33-Z-Sx) | Cx (-50 ...+85 °C [-58 ...+185 °F], 5 ... 95 %r.h.) Version with FKM O-ring: -20 °C [-4 °F] |
| Maximum permissible humidity, condensation | 100 %r.h., condensation allowed |
| Maximum operating pressure 1) 2) | |
| For protection tube diameter 3 mm[0.12 in] | 140 bar [2,030 psi] |
| For protection tube diameter 6 mm[0.24 in] | 270 bar [3,916 psi] |
| Salt fog | IEC 60068-2-11 |
| Vibration resistance per IEC 60751 | 10 ... 2,000 Hz, 20 g 1) |
| Shock resistance per IEC 60068-2-27 | 50 g, 6 ms, 3 axes, 3 directions, three times per direction |
| Maximum permissible autoclaving conditions | Max. 134 °C, 3 bar abs., 100 %r.h., duration 20 min., max. 50 cycles |
| Autoclavable with mounted protective cap at coupler connector | |
| Conditions for outdoor use (only applies to UL approval) | ■ The instrument is suitable for applications with pollution degree 3. ■ The power supply must be suitable for operation above 2,000m should the temperature transmitter be used at this altitude. ■ The instrument shall be installed in locations sheltered from the weather. ■ The instrument shall be installed sun/UV irradiation protected. |
| Ingress protection (IP code) | |
|---|---|
| Case with connected connector 3) | ■ IP67 per IEC/EN 60529 ■ IP69 per IEC/EN 60529 ■ IP69K per ISO 20653 |
| Coupler connector, not connected | IP67 per IEC/EN 60529 |
| Weight | approx. 0.2 ... 0.7 kg [0.44 ... 1.54 lbs] - depending on version |
| Logo | Description | Country |
|---|---|---|
| EU declaration of conformity | European Union | |
| CSA Safety (e.g. electr. safety, overpressure, ...) | USA and Canada | |
| UL Safety (e.g. electr. safety, overpressure, ...) | USA and Canada |
| Logo | Description | Country |
|---|---|---|
| EAC EMCdirective 1) | Eurasian Economic Community | |
| GOST Metrology, measurement technology | Russia | |
| KazInMetr Metrology, measurement technology | Kazakhstan | |
| - | MTSCHS Permission for commissioning | Kazakhstan |
| BelGIM Metrology, measurement technology | Belarus | |
| UkrSEPRO Metrology, measurement technology | Ukraine | |
| Uzstandard Metrology, measurement technology | Uzbekistan |
| Certification type | Measurement accuracy | Material certificate |
|---|---|---|
| 2.2 test report | x | x |
| 3.1 inspection certificate | x | x |
| DKD/DAkkS calibration certificate | x | - |
| Model | Description | Order no. | |
|---|---|---|---|
| Programming unit Model PU-548 | ■ Easy to use ■ LED status display ■ Compact design ■ No further voltage supply needed, neither for the programming unit nor for the transmitter | 14231581 | |
| Adapter cable M12 to PU-548 | Adapter cable for the connection of a modelTR33 resistance thermometer to the model PU-548 programming unit | 14003193 | |
| M12 x 1 transmitter adapter to DIN EN 175301-803 angular connector (yellow female connector element) | Adapter for the connection of a resistance thermometer with a DIN EN 175301-803 angular connector form A with a 4 ... 20 mAoutput signal → see data sheet AC 80.17 Case: PA Ambient temperature: -40 ...+115 °C [-40 ...+239 °F] Union nut: Zinc die-cast Contacts: Copper-zinc alloy, tinned Dielectric strength: 500V Ingress protection: IP65 4 ...20mA M12 x 1 connector Angular connector | 14069503 | |
| M12 x 1 Pt adapter to DIN EN 175301-803 angular connector (black female connector element) | Adapter for the connection of the resistance thermometer with a DIN EN 175301-803 form A angular connector with direct resistance output signal → see data sheet AC 80.17 Case: PA Ambient temperature: -40 ...+115 °C [-40 ...+239 °F] Union nut: Zinc die-cast Contacts: Copper-zinc alloy, tinned Dielectric strength: 500V Ingress protection: IP65 mA 4 ...20mA M12 x 1 connector Angular connector | 14061115 | |
| Angular connector | Per DIN EN 175301-803 form A | 11427567 | |
| Sealing for angular connector | For use with angular connector DIN EN 175301-803-A EPDM, brown | 11437902 |
| Model | Description | Order no. | ||
|---|---|---|---|---|
| - | M12 connection cable | Cable socket straight, 4-pin, ingress protection IP67 Temperature range -20 ...+80 °C [-4 ...+176 °F] | Cable length2m [6.56 ft] | 14086880 |
| Cable length5m [16.40 ft] | 14086883 | |||
| Cable socket straight, 4-pin, ingress protection IP69K, hygienic design Union nut from stainless steel Temperature range -40 ...+80 °C [-40 ...+176 °F] | Cable length3m [9.84 ft] | 14137167 | ||
| Cable length5m [16.40 ft] | 14137168 | |||
| Angled socket, 4-pin, ingress protection IP67 Temperature range -20 ...+80 °C [-4 ...+176 °F] | Cable length2m [6.56 ft] | 14086889 | ||
| Cable length5m [16.40 ft] | 14086891 | |||
| Angled socket, 4-pin, ingress protection IP69K, hygienic design Union nut from stainless steel Temperature range -40 ...+80 °C [-40 ...+176 °F] | Cable length3m [9.84 ft] | 14137169 | ||
| Cable length5m [16.40 ft] | 14137170 | |||
| - | M12 connector | Female angled, 4-pin, ingress protection IP67 Screw connection for conductor cross-section 0.25 …0.75 mm² [24 …18 AWG] Cable gland Pg7, outside diameter of cable 4 …6mm[0.16 ... 0.24 in] | 14136815 |
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →80 standard articles of family TR33
If you know the code — pick directly. If not — use the configurator in mode “01”.