AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
CTM9350-165 1
CTM9350-165
On order
WIKA · Calibration technology · Family CTM9350

Temperature multi-function calibrator

Temperature multi-function calibrator with touchscreen, dry-well, micro bath, surface, and infrared black body functions for laboratory and on-site use.
Range
-30…165 °C
Output
RTD simulation · TC simulation
Protection
IP20
CERoHS
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

Whether in laboratories, workshops or on-site, the CTM9350 series of temperature multi-function calibrators can meet any calibration requirement. All instruments can be fitted, with an integrated measuring instrument. This enables the measurement of resistances, thermoelectric voltages and also current signals (from thermometers with a 0/4 ... 20 mA transmitter) and their direct display in the selected unit.

Using a dry-well calibrator or a micro calibration bath to calibrate either surface thermometers or non-contact thermometers does not reflect the application and can result in false values. In these cases, the model CTM9350-165 multi-function calibrator should be used.

Fig. left: with integrated measuring instrument

Fig. right: without integrated measuring instrument

With this temperature multi-function calibrator, in the temperature range from -30 ... +165 °C [-22 ... +329 °F], you can cover not only the common functions, but also, with special inserts, this can be used as a surface temperature calibrator and an infrared black body.

It is operated via a large colour touchscreen. Test tasks can be created and automated via the user interface, saving the user a lot of time. Operation is intuitive and fast.

Special features

  • Easy operation via intuitive, user-friendly menus
  • Large, easy-to-read touchscreen
  • Short response times due to optimised control
  • Multi-function instrument with four controller parameter sets
  • Creation of calibration tasks incl. preparation of a certificate
Identification
ModelCTM9350-165
Product familyCTM9350
Subcategorytemperature multi-function calibrator
Measuring range
Temperature ranges−30…+165 °C · −30…+160 °C · −30…+155 °C · −25…+150 °C
Accuracy
Uncertainty±0.07 K (dry-well), ±0.10 K (micro bath), ±0.5 K (surface and IR)
Process conditions
Mediumair, silicone oil, mineral oil, water
Calibration typemultifunction
Electrical characteristics
Output signalRTD simulation · TC simulation · mA · V · voltage · current · resistance · thermocouple
Operating environment
Ambient temperature+0…+50 °C
IP ratingIP20
Weight13 kg
Certifications and approvals
Certifications
CERoHS
TraceabilityDAkkS
Additional details
PortableNo
Datasheet languagesEN, DE
Tables from datasheet
Basic information
Temperature rangeApplication as:
Dry-well calibrator-30 ...+160 °C [-22 ...+320 °F]
Micro calibration bath-30 ...+155 °C [-22 ...+311 °F]
Surface temperature calibrator-25 ...+150 °C [-13 ...+302 °F]
Infrared black body-30 ...+165 °C [-22 ...+329 °F]
AccuracyApplication as:
Dry-well calibrator±0.07 K
Micro calibration bath±0.10 K
Surface temperature calibrator±0.5 K
Infrared black body±0.5 K
Defined the measuring deviation between the measured value and the reference value.
Temperature stabilityApplication as:
Dry-well calibrator±0.005 K
Micro calibration bath±0.01 K
Surface temperature calibrator±0.150 K
Infrared black body±0.020 K
Maximum temperature difference at a stable temperature over 30 minutes.
Temperature control
Heating timeFrom +20 °C to +155 °C [from +68 °F to +311 °F]27 min
From -20 °C to +155 °C [from 4 °F to +311 °F]34 min
Cooling timeFrom +20 °C to -25 °C [from +68 °F to -13 °F]35 min
From +165 °C to +30 °C [from +329 °F to 86 °F]17 min
Stabilisation timeDependent on temperature and temperature probe
Metal block
Dimension for calibration insertØ60 x 170 mm[Ø2.36 x 6.69 in]
Dry-well materialAluminium
Basic information
Digital display instrument
DisplayBright colour touchscreen (7"), laminated safety glass
Display range-199.9 ... +999.9 °C [-199.9 ... +999.9 °F]
Resolution0.001 °C
Units°C, °F, K
Menu languagesEnglish German
Functions
Menu functionsCalibration without certificate Calibration with automatic certificate generation Remote control Data export to a USB stick
User settingsUser-defined data is indicated on the test certificate
Dimensions (W x D x H)
Calibrator without carrying handle210 x 300 x 430 mm[8.27 x 11.81 x 16.93 in]
Height of carrying handle50 mm[1.97 in]
Weight13 kg [28.67 lb]
Accuracy specificationsApplication as
Dry-well calibratorMicro calibration bath
Temperature range-30 ...+160 °C [-22 ...+320 °F]-30 ...+155 °C [-22 ...+311 °F]
Accuracy 1)±0.07 K±0.10 K
Temperature stability 2)±0.005 K±0.01 K
Influence due to loading 1)
External reference temperature probe±0.01 K±0.02 K
Temperature distributionDetermined in accordance with current calibration guideline in a standard insert sleeve.
Axial homogeneity±0.06 K±0.1 K
Radial homogeneity±0.01 K±0.08 K
Hysteresis±0.010 K±0.010 K
Communication
Interface3 x USB Ethernet
ConnectivityOPC UA Serial communication HTTP
Baud rate2400
Measuring rate1 measured value per second
Internal programTest items, test tasks and test points can be applied without limit
View all (17)
Voltage supply and performance data
Operating voltage 1)AC 90 ... 240 V, 50/60 Hz
Power consumption375W
FuseMicro fuse, 6.3 AH 250V slow blow fuse
Voltage supply and performance data
Power cord■ Europe ■ USA/Canada ■ Switzerland ■ UK ■ China
Operating conditions
Place of useFor indoor use only
Operating altitude≤ 2,000 m[≤ 6,562 ft] above sea level
Operating temperature0 ... 50 °C [32 ... 122 °F] The ambient temperature influences the heating/cooling behaviour
Storage and transport temperature range-10 ...+60 °C [14 ... 140 °F]
Humidity< 80 %relative humidity to 31 °C [88 °F] Decreasing linearly to 50 %relative humidity at 40 °C [104 °F]
CondensationNon-condensing
Electrical safety
Overvoltage categoryII
Pollution degree2 per IEC-61010-1
Output signal
Analogue output
Voltage supplyDC 24V (can be activated via menu)
LoadMax.24mA
Switching functionNC,NO
Electrical connection
Number of channels
Resistance thermometer2
Thermocouple2
Current signal1
Voltage signal1
Switch test2
Connection type
Resistance thermometer4 x 4-mm banana jacks
Thermocouple2 x thermocouple terminal (mini)
Current and voltage signal4 mmbanana jacks
Pin assignment
Resistance thermometer2-wire connection 3-wire connection 4-wire connection
Measuring range
Resistance thermometerPt100 0 ...400Ω
Pt1000 0 ...4,000Ω
Thermocouple-10 ...+100mV
Current signalDC 0 ...24mA
Voltage signalDC 0 ...12V
AccuraciesMeasuring rangeAccuracy
Resistance thermometer
Pt100-200 ...+850 °C [-328 ... +1,562 °F]±0.03 K
Pt500-200 ...+850 °C [-328 ... +1,562 °F]±0.12 K
Pt1000-200 ...+850 °C [-328 ... +1,562 °F]±0.06 K
Ni100-60 ...+180 °C [-76 ...+356 °F]±0.02 K
Ni500-60 ...+200 °C [-76 ...+392 °F]±0.08 K
Ni1000-60 ...+200 °C [-76 ...+392 °F]±0.04 K
Cold junction-200 ... +1,820 °C [-328 ... +3,308 °F]±0.3 K
Thermocouple
Type K-160 ... +1,260 °C [-256 ... +2,300 °F]±0.08 K
Type J-190 ... +1,200 °C [-310 ... +2,192 °F]±0.07 K
Type N0 ... 1,300 °C [32 ... 2,372 °F]±0.13 K
Type E-200 ... +1,000 °C [-328 ... +1,832 °F]±0.06 K
TypeT-130 ...+400 °C [-202 ...+752 °F]±0.09 K
Type R160 ... 1,760 °C [320 ... 3,200 °F]±0.78 K
Type S170 ... 1,760 °C [338 ... 3,200 °F]±0.73 K
Type B920 ... 1,820 °C [1,688 ... 3,308 °F]±0.5 K
Direct current0 ...24mA0.01 %of end value
DC voltage0 ...12V0.01 %of end value
LogoDescriptionRegion
EU declaration of conformityEuropean Union
EMCdirective EN 61326 emission (group 1, class A) and immunity (industrial environment)
LowVoltage Directive EN 61010, safety requirements for electrical equipment for measurement, control and laboratory use
RoHS directive
Description
Calibration
Integrated measuring instrument■ Without ■ 3.1 inspection certificate per EN 10204 for RTD, TC, mAandV ■ DAkkS calibration certificate for RTD, TC, mAandV
Instrument■ 3.1 inspection certificate per EN 10204 ■ DAkkS calibration certificate as micro calibration bath (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate as temperature dry-well calibrator (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate as micro calibration bath and as temperature dry-well calibrator (traceable and accredited in accordance with ISO/IEC 17025)
Recommended calibration interval1 year (dependent on conditions of use)
Inserts The insert has several additional reference either heated or cooledDescription 1) been reached, the thermometer.The
bores into which the temperature probes being calibrated and one thermometer, for comparative calibration, can be inserted.The block is to the desired calibration temperature.Once a stable temperature has temperature probes to be calibrated can be compared with the reference documentation of this comparison represents the calibration.
Insert for surface measurement The calibration of surface temperature probes is very difficult and not without controversy. Temperature probes mounted on surfaces dissipate heat from the surface and create a cold zone on the surface being measured. In the temperature multi-function calibrator, the calibration temperature is created in a specially designed surface insert and measured directly external reference thermometer.under the surface by an
measurementInsert for infrared
measuring spot of the pyrometer being calibrated must be smaller than the diameter of infrared insert.The sleeve has been specifically manufactured with regard to its design and in order to achieve a defined emissivity for the measurement. the hollow and specially designed insert into the block using a special replacement tool.The also has three bores in the edge with 2 x 3.5 mmand 1 x 4.5 mm[2 x 0.14 in and 1 x in], for the accurate monitoring of the temperature via external reference probes. insert has a special design and surface finish on the inside.Through this, an emissivity of achieved.The the surface Fit sleeve 0.18 The 0.9994 (black body) is
Insert for liquids and probe basket for tank Angled probes, large-diameter probes or probes with special designs cannot be calibrated with a dry-well calibrator. For this reason, the temperature multi-function calibrator also has the possibility to function as a stirred liquid bath.The liquid is circulated using a magnetic stirrer, and thus provides very good temperature distribution within the bath.The liquids are selected calibration temperature.depending upon the desired
Reference thermometer Angled temperature probes are supplied to match the insert.
Accessories for model CTM9350-165Order code
Description 1)CTX-A-KE
Transport case with trolley frame-3-
External reference probe up to max. 255 °C [528 °F] Dimension:Ø3x300 mm[Ø0.12 x 11.81 in] Angled: 190 mm[7.48 in] - length from the the sensor tip-E-
Insert replacement tool-A-
Insert replacement tool For insert for surface measurement-B-
Drain pump-C-
Standard sleeve for the surface measurement operating mode Dimensions:Ø60 x 205 mm[Ø2.36 x 8.07 in] Material: aluminium 3.4365-D-
Standard sleeve for the infrared measurement operating mode Dimensions:Ø60 x 150 mm[Ø2.36 x 5.91 in] Material: aluminium 3.4365-F-
Standard sleeve for the operating mode as temperature dry-well calibrator Dimensions:Ø60 x 150 mm[Ø2.36 x 5.91 in] Material: aluminium 3.4365-G-
Silicone oil DC 200.10 In 1 litre plastic bottle For temperature range -35 ...+160 °C [-31 ...+320 °F];FP = 163 °C [325.4 °F]-H-
Replaceable insert for liquids New adjustment required-I-
Screw-on lid Material: stainless steel-J-
Screw-on lid with 6 G ¼ bores Material: plastic-K-
Accessories for model CTM9350-165Order code
Description 1)CTX-A-KE
Power cord For the EU-L-
For Switzerland-M-
For UK-N-
For USA/Canada-O-
Electrical connection set Consisting of: ■ Clamp connectors (4 x red, 4 x black and 1 x white) ■ 2 x thermocouple adapters ■ 2 x split ferrite cores ■ 2 x ferrite keys-P-
PC and network cable-Q-
Ordering information for your enquiry:
1. Order code: 2. Option:⇓ [ ]
Inserts for model CTM9350-165Order code
Description 1)CTA9I-4U
Undrilled insert Dimensions:Ø60 x 150 mm[Ø2.36 x 6.69 in] Material: aluminium-N-
Drilled insert Dimensions:Ø60 x 150 mm[Ø2.36 x 6.69 in] Drilling depth: 145 mm[5.71 in] Material: aluminium-
Bore diameter: 1 x 3.2 mmand 1 x 6.3mm [1 x 0.13 in and 1 x 0.25 in]-A-
Bore diameter: 2 x 3.2 mm, 1 x 4.2 mm, 1 x 6.3 mm, 1 x 8.4 mmand 1 x 9.9mm [2 x 0.13 in, 1 x 0.17 in, 1 x 0.25 in, 1 x 0.33 in and 1 x 0.39 in]-B-
Bore diameter: 1 x 3.2 mm, 1 x 5.0 mm, 1 x 6.5 mmand 1 x 10.5mm [1 x 0.13 in, 1 x 0.20 in, 1 x 0.26 in and 1 x 0.41 in]-U-
Bore diameter: 1 x 3.2 mm, 1 x 5.0 mm, 1 x 7.0 mmand 1 x 10.5mm [1 x 0.13 in, 1 x 0.20 in, 1 x 0.28 in and 1 x 0.41 in]-V-
Bore diameter: 1 x 3.3 mm, 1 x 4.8 mmand 2 x 6.4mm [1 x 0.13 in, 1 x 0.19 in and 2 x 0.25 in]-W-
Bore diameter: 2 x 3.2 mm, 2 x 4.3 mm, 3 x 6.3 mmand 2 x 8.5mm [2 x 0.13 in, 1 x 0.17 in, 3 x 0.25 in and 2 x 0.33 in]-M-
Bore diameter: 1 x 2.0 mm, 3 x 3.2 mm, 2 x 4.3 mm, 1 x 5.0 mmund 2 x 6.5mm [1 x 0.08 in, 3 x 0.13 in, 2 x 0.17 in, 1 x 0.2 in and 2 x 0.26 in]-Y-
Customer-specific, special probes are possible on request.-?-
Inserts for model CTM9350-165Order code
Description 1)CTA9I-4U
Insert replacement tool-J-
Ordering information for your enquiry:
1. Order code: CTA9I-4U 2. Option:⇓ [ ]