Level sensor with reed measuring chain
Description
The model N-ALM level sensor is used to measure and transmit the level of liquids in conjunction with a customer's control unit outside of the containment area. It is based on the float principle with magnetic transmission in a 3-wire potentiometer circuit. The resistance measuring chain is closely stepped. Due to this assembly the generated voltage is approximately continuous. Signal transmission over large distances and use in hazardous areas are possible.
This instrument was designed without any organic materials or active electronic components in the sensors or cabling that could be affected by the accident conditions described. The sensor will continue to function even in case of a Loss of Coolant Accident (LOCA) and can be fitted with a filter for protection against coarse debris. An accident-proof connection using mineral-insulated metal-sheathed cable is also available and meets the same standards as the integrated sensor units.
Accident-proof level sensor with reed-chain technology, model N-ALM
This instrument is ideal for retrofitting the existing level measuring instrument as part of a wide range of post-accident monitoring systems.
The purpose of accident level measurement is to ensure functionality during and after a severe accident scenario. All materials used are inorganic and capable of withstanding a high dose of gamma radiation.
Special features
- Process- and system-specific solutions possible
- Functional under LOCA conditions
- High resistance to temperature, pressure, radiation and humidity
- Seismic qualification up to 2.5g (up to 5g on request)
| Model | N-ALM |
| Subcategory | float |
| Level ranges | 0…20 m |
| Process temperature | −10…+250 °C |
| Max process pressure | 10.5 bar |
| Measurement technology | float |
| Output signal | resistance (3-wire potentiometer circuit) |
| Electrical connection | mineral-insulated metal-sheathed cable |
| Wetted parts material | stainless steel |
| Ambient temperature | −10…+70 °C |
| IP rating | IP68 |
| Certifications | KTA 3505IEC/IEEE 60780-323:2016RCC-E edition 2019 |
| Datasheet languages | EN, DE |
| Operating temperature | T = -10 °C ... +70 °C [14 ... 158 °F] |
|---|---|
| Operating pressure | Ambient |
| Lengths | Up to 20 m [65 ft] |
| Humidity | 100 % |
| Operational radiation dose | ≤ 160 kGy [16 MRad] |
| Accident temperature | 250 °C [482 °F] |
|---|---|
| Accident pressure | 10.5 bar [1.05 MPa] |
| Accident mission time | 1 year |
| Accident radiation dose | Integrated 5.05 MGy [505 MRad] |
| Seismic acceleration | Up to 2.5g (higher accelerations up to 5g on demand) |
| Model 1) | Reed Switch Unit (RSU) | Magnetic Float Unit (MFU) | Stainless steel material | Float Guide Tube Unit (FGTU) | Filter box | Pt100 tem- perature sensor | Redundant measuring equipment | Modular design |
|---|---|---|---|---|---|---|---|---|
| ALM-D1 | x | x | x | x | x | - | - | - |
| ALM-D2 | x | x | x | - | - | - | - | - |
| ALM-D3 | x | x | x | - | - | - | - | x |
| ALM-D4 | x | x | x | x | x | x | - | - |
| ALM-D5 | x | x | x | x | x | - | x | - |
| ALM-D6 | x | x | x | x | - | x | - | x |
| BLR-ALM | x | x | x | x | x | - | - | - |
| BLR-NK | x | x | x | x | x | - | - | - |
| Qualification | |
|---|---|
| ALM-D1 | IEC/IEEE 60780-323:2016, KTA 3505 |
| ALM-D2 | |
| ALM-D3 | |
| ALM-D4 | |
| ALM-D5 | |
| ALM-D6 | |
| BLR-ALM | RCC-E edition 2019 |
| BLR-NK | IEC/IEEE 60780-323:2016 |
Couldn't find the answer in the documentation? An engineer will reply within a business day.