ILT-C05 1
ILT-C05
On order
WIKA · Level measurement · Family ILT

Industrial radar level sensor

Industrial radar level sensor for continuous non-contact level measurement of liquids, highly viscous media, and solids up to 150 °C, suitable for mobile working machines and industrial applications.
Range
0.1…5 m
Output
4...20 mA · 0...5 V
Accuracy
±5 mm
Protection
IP67
RoHSFCCISEDACMARSMCE
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The ILT free-beam radar level sensor enables non-contact level measurement, even through plastic vessel walls.

The free-beam radar for level measurement emits electromagnetic waves that are reflected by the surface of the material to be measured. The reflected signals are received and analysed, providing information about the level. Up to two additional switch points can be integrated to enable further switching outputs, e.g. for overflow and/or dry run protection.

Thanks to its compact design and easy handling, the model ILT radar sensor is ideal for measuring applications in industrial environments and in mobile working machines.

Radar level sensor, model ILT-C05

It has been prepared for the requirements of mobile working machines and offers high-precision measurements at temperatures of -40 ... +150 °C [-40 ... +302 °F]. Thanks to metal shielding, the sensor operates, interference-free, at field strengths of up to 100 V/m and remains reliable even with vibrations of up to 40g and shocks of up to 100g. The model ILT is particularly safe in operation, even under difficult conditions such as dust, moisture and heat. The non-contact measuring principle, combined with a maintenance-free design, ensures low total cost of ownership.

Special features

  • Developed for the extreme operating conditions in mobile working machines
  • Continuous, accurate measuring signal
  • Compact design for easy integration
  • Can also be installed outside a container (plastic)
  • Adjustment function for complex container geometries possible
Identification
ModelILT-C05
Product familyILT
SubcategoryFree-beam radar, high-temperature version
Measuring range
Level ranges0.1…5 m
Process conditions
Process connection
G ½ AG1/2A1/2"NPT
Process temperature−40…+150 °C
Max process pressure25 bar
Measurement technologyradar
Electrical characteristics
Output signal4...20 mA · 0...5 V · PNP switching output · DC0...5V · PNP · NPN
Power supplyDC 8...36 V
Electrical connectionCircular connector M12 x 1, 4-pin or 5-pin; Cable outlet unshielded
Materials
Wetted parts materialPEEK, FKM, Stainless steel 1.4404
Operating environment
Ambient temperature−40…+100 °C
IP ratingIP67
Certifications and approvals
Certifications
RoHSFCCISEDACMARSMCE
Additional details
Datasheet languagesEN, DE
Tables from datasheet
Circular connector M12 x 1
4-pinLevel + 1 switch point
1U+
2S+
3U-
4SP1
Cable outlet
UnshieldedLevel + 2 switch points
BrownU+
WhiteU-
GreenS+
YellowSP1
PinkSP2
Operating conditions
Medium temperature limit-40 … +150 °C [-40 … +302 °F]
Ambient temperature limit-40 … +100 °C [-40 … +212 °F]
Storage temperature limit-40 … +70 °C [-40 … +158 °F]
Pressure limit of medium0 … 25 bar [0 … 362 psi]
Vibration resistance per IEC 60068-2-640g, 10 ... 2,000 Hz
Permanent vibration resistance per IEC 60068-2-610g, 10 ... 2,000 Hz
Shock resistance per IEC 60068-2-27100g, 11 ms
Free fall in line with IEC 60068-2-31
Single instrument1 m [3.28 ft]
Multiple packaging0.5 m [1.64 ft]
Mounting positionVertical
Ingress protection (IP code) per IEC 60529IP67
EMC
ESD per ISO 10605±8 kV contact discharge, ±15 kV air discharge
HF field per ISO 11452-2100 V/m
BCI per ISO 11452-4200 mA
Pulse 1 per ISO 7637-2 1)Level III
Pulse 2a per ISO 7637-2 1)Level III
Pulse 2b per ISO 7637-2 1)Level III
Pulse 3a per ISO 7637-2 1)Level III
Pulse 3b per ISO 7637-2 1)Level III
Fast transient pulses per ISO 7637-3Level IV
Radio disturbance per CISPR 2530 ... 1,000 MHz
Packaging and instrument labelling
Packaging■ Individual packaging ■ Multiple packaging (up to 50 pieces possible)
Instrument labelling (product label)■ WIKA product label, adhesive foil ■ Customised product label on request
View all (7)
LogoDescriptionRegion
EU declaration of conformityEuropean Union
EMC Directive EN 61326 emission (group 1, class B) and immunity (industrial environments) RED - Radio Equipment Directive ETSI EN 305 550 The instrument may be used without restrictions RoHS directive Federal Communications Commissionin the EU and in the EFTA countries. (FCC) for US USA
- Innovation, Science and EconomicDevelopment (ISED) for Canada Canada
ACMA Radio(Australian Communication and Media Authority) approvalAustralia
RSM (Radio RadioSpectrum Management) approvalNew Zealand
DescriptionTemperature rangeCable diameterCable lengthOrder num- ber
Circular connector M12 x 1 with moulded cable
Straight version, cut to length, 4-pin, PUR cable,-20 ... +80 °C [-4 ... +176 °F]4.5 mm [0.18 in]2 m [6.6 ft]14086880
UL Listed, IP675 m [16.4 ft]14086883
10 m [32.8 ft]14086884
Angled version, cut to length, 4-pin, PUR cable, UL Listed, IP67-20 ... +80 °C [-4 ... +176 °F]4.5 mm [0.18 in]2 m [6.6 ft]14086889
5 m [16.4 ft]14086891
10 m [32.8 ft]14086892
Connection cable M12 x 1 with integrated LED display
Connection cable, 4-pin, PUR cable, UL Listed, IP67 1 x LED green, 2 x LED yellow-20 ... +80 °C [-4 ... +176 °F]4.5 mm [0.18 in]2 m [6.6 ft]14252834
Connection cable, 4-pin, PUR cable, UL listed, IP67 1 x LED green, 2 x LED yellow-20 ... +80 °C [-4 ... +176 °F]4.5 mm [0.18 in]5 m [16.4 ft]14252835
DescriptionOrder number
Adjustment magnet14760395