AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
F1136 1
F1136
On order
WIKA · Force measurement · Family F1136

Hydraulic compression force transducer

Clamping force test instrument up to 500 kN
Range
0…500 kN
Accuracy
±1.6 % Fnom (analogue)
Protection
IP65
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The model F1136 (NS 80) hydraulic force transducer enables the simple and economical force measurement in vices and clamping systems and display of forces of measuring ranges up to 500 kN. The flattened case of this compression force transducer enables a stable force measurement.

The force is measured using the principle of hydraulics: The force acting on a piston leads to a pressure increase that can be visualised on a connected display instrument. The scale of the display instrument can be defined in various units (e.g. N, kN, kg, t).

Special features

  • Measuring ranges 0 ... 1.2 kN to 0 ... 500 kN
  • Flattened case for stable measurement
  • Relative linearity error ±1.0 ... 1.6% F nom with analogue pressure gauge, ±0.5% Fnom with digital pressure gauge or pressure sensor (For rated forces below 500 N, the relative linearity error is ±1.6 % F nom for all connected measuring instruments.)
  • Operates without supply voltage
  • 5-year leak-tightness warranty (Use of the force measuring unit as intended is a prerequisite for the extended 5-year warranty.)
Identification
ModelF1136
SubcategoryHydraulic compression force transducer
Measuring range
Force ranges0…1.2 kN · 0…500 kN · 1.2…500 kN
Accuracy
Accuracy class±1.6 % Fnom (analogue), ±0.5 % Fnom (digital)
Process conditions
Process temperature−25…+50 °C
Mounting typeDirect, Adapter, capillary, measuring hose
Materials
Cell materialStainless steel
Operating environment
Ambient temperature−25…+50 °C
IP ratingIP65
Weight4.3 kg
Additional details
Overload protection100 %
Datasheet languagesEN
Tables from datasheet
Model F1136
Rated force F nom0 ... 1.2 kN to 0 ... 500 kN
Nominal sizeNS 80
Display ■ Standard ■ OptionPressure gauge 213.40 (NS 63) Digital pressure gauge DG-10 Pressure sensor (on request)
Relative linearity error d lin ■ Standard ■ Option≤ ±1.6%F nom (analogue display) 1) ≤ ±0.5%F nom (pressure sensor/digital pressure gauge) 1)
Limit force F L100%F nom
Breaking force F B> 130%F nom
Rated displacement s nom< 0.5mm
Rated temperature range B T, nom-25 …+50 °C
Ingress protectionIP65 per EN/IEC 60529
CaseStainless steel
PistonStainless steel
Mounting type ■ Standard ■ OptionDirect Adapter, capillary, measuring hose for &#x27;separation without any losses&#x27;
Fill fluidGlycerine 70 %, water30%
Weight in kg ■ with pressure gauge 213.40 (NS 63) ■ with digital pressure gauge DG-104.3 4.5
VersionDisplayOptions
Rated forceSystem pressure213.40DG-10Measuring hose DN 2 [max. L]Capillary [max. L]
kNbarm
1.21.6---
22.5---
3.24--1.0
56-0.51.0
810-1.02.0
1216-1.02.0
1620- 1)1.52.0
2025-1.52.0
3240-1.52.0
4050-2.02.0
5060-2.02.0
801002.02.0
1201602.04.0
2002503.24.0
250315-3.24.0
3204003.26.0
5006003.26.0
Other rated loads and versions on request