AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
F1122 1
F1122
On order
WIKA · Force measurement · Family F1122

Hydraulic compression force transducer

NS 3 x 33, 3-jaw clamping force test instrument up to 1,000 kN
Range
0…1000 kN
Accuracy
±0.5 % Fnom (with digital pressure gauge or pressure sensor)
Protection
IP65
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The hydraulic force transducer model F1122, version NS 3 x 33, is suited for regular testing of the clamping force in 3-jaw chucks. It thereby ensures optimum use of the clamping jaws.

Hydraulic force measurement is a simple way to capture and display the forces occurring in various applications. The force is measured using the principle of hydraulics: The force acting on a piston leads to a pressure increase that can be visualised on a connected display instrument. The scale of the display instrument can be defined in various units (e.g. N, kN, kg, t).

Special features

  • Measuring ranges 0 ... 6 kN to 0 ... 1,000 kN
  • Relative linearity error ±1.0 ... 1.6 % F nom with analogue pressure gauge, ±0.5 % F nom with digital pressure gauge or pressure sensor (For rated loads below 500 N, the accuracy is ±1.6 % F nom for all connected measuring instruments.)
  • Piston stroke ≤ 0.5 mm, force introduction as total clamping force
  • Operates without supply voltage
  • 5-year leak-tightness warranty (Use of the force measuring unit as intended is a prerequisite for the extended 5-year warranty.)
Identification
ModelF1122
SubcategoryHydraulic force transducer
Measuring range
Force ranges0…6 kN · 0…1000 kN
Accuracy
Accuracy class±0.5 % Fnom (with digital pressure gauge or pressure sensor), ±1.6 % Fnom (analogue)
Process conditions
Process temperature−10…+50 °C
Mounting typeCompression
Materials
Cell materialStainless steel
Operating environment
Ambient temperature−10…+50 °C
IP ratingIP65
Weight6.7 kg
Additional details
Overload protection130 %
Datasheet languagesEN
Tables from datasheet
Model F1122
Rated force F nom0 ... 6 kN to 0 ... 1,000 kN
Nominal sizeNS 3 x 16
Display ■ Standard ■ OptionPressure gauge 232.57 (NS 100) Digital pressure gauge DG-10 Pressure sensor (on request)
Relative linearity error d lin ■ Standard ■ Option≤ ±1.6%F nom (analogue display) 1) ≤ ±0.5%F nom (pressure sensor/digital pressure gauge) 1)
Limit force F L100%F nom
Breaking force F B> 130%F nom
Force introduction ■ Standard ■ OptionTotal clamping force Clamping force per jaw
Rated displacement s nom< 0.5mm
Rated temperature range B T, nom-10 …+50 °C
Ingress protection (per EN/IEC 60529)IP65
CaseStainless steel
PistonStainless steel
Mounting type ■ Standard ■ OptionAdapterL=50mm Capillary Measuring hose for &#x27;separation without any losses&#x27;
Fill fluidGlycerine/water 70 %/30%
Weight in kg ■ with pressure gauge 232.57 (NS 100) ■ with digital pressure gauge DG-106.9 6.7
VersionDisplayOptions
Rated forceSystem pressure232.57DG-10Measuring hose DN 2 [max. L 1) ]Capillary [max. L 1) ]
kNbarm
66-0.51.0
1010-1.02.0
1616-1.02.0
2020- 2)1.52.0
2525-1.52.0
4040-1.52.0
5050-2.02.0
6060-2.02.0
1001002.02.0
1601602.04.0
2502503.24.0
4004003.26.0
6006003.26.0
1,0001,000--6.0