Pressure gauge per EN 837-1 with mounted diaphragm seal
Description
Diaphragm seal systems are used to protect the pressure measuring instrument from aggressive, adhesive, crystallising, corrosive, highly viscous, environmentally hazardous or toxic media. The diaphragm made of stainless steel provides for the separation from the medium. The pressure is transmitted to the measuring instrument via the system fill fluid which is inside the Diaphragm seal system.
Due to its design, the DSS10M can be used universally. A replacement of the lower body is possible (e.g. when changing the process connection) without having to carry out modifications on the diaphragm seal system.
Special features
- Process connection with thread for direct threaded connection
- Version with internal diaphragm
- Diaphragm seal parts screwed together
- Universal application
Build a DSS10M for your task
The article number is assembled automatically. If unsure — ask an engineer; we'll match the line's parameters.
- Pressure measuring instrumentManometer according to EN 837-1 nominal size 100 mm
- Case filling—
- Unit of outer Scalebar
- Scale range—
- Assemblydirect assembly
- System fill fluidKN 2 Silicone oil for general applications
- Unit of temperatureCelsius
- Process temperature min.- 10 °C
- Process temperature max.+ 150°C
- Ambient temperature pressure meas. min+ 10 °C
- Ambient temperature pressure meas. max+ 40 °C
- Process connection standard—
- Threaded connection—
- Nominal pressure ratingPN 100
- Wetted parts—
- GasketPTFE max. 260°C
- Fastening partsRetainer flange and bolts in stainless steel
- Certificateswithout
Finish your configuration
Select every option to submit the request.
| Model | DSS10M |
| Product family | DSS10 |
| Subcategory | Diaphragm seal system with pressure gauge |
| Pressure ranges | 0…1 bar · 0…1.6 bar · 0…2.5 bar · 0…4 bar · 0…6 bar · 0…10 bar · 0…16 bar · 0…25 bar · 0…40 bar · 0…60 bar · -1…5 bar · -1…9 bar · -1…10 bar |
| Accuracy class | 1.0 |
| Accuracy standard | EN 837-1 |
| Process connection | G ½½ NPTG 1/21/2 NPT |
| Process temperature | −10…+150 °C |
| Case material | Stainless steel 1.4301 [304] |
| Wetted parts material | Stainless steel CrNi-Stahl 1.4435 [316L], Hastelloy C276, Diaphragm: Stainless steel CrNi-Stahl 1.4435 [316L] or Hastelloy C276, Diaphragm seal: CrNi-Stahl 1.4435 [316L] or Hastelloy C276, Diaphragm seal, pressure element: Stainless steel |
| Sealing material | Silicone oil KN 2 |
| Ambient temperature | +10…+40 °C |
| IP rating | IP65 |
| Certifications | EN 10204 3.13.1 inspection certificate per EN 10204 |
| Dial size | 100 mm |
| Datasheet languages | EN |
| Version | Pressure gauge with Bourdon tube per EN 837-1, screwed to diaphragm seal |
|---|---|
| Nominal size (NS) | 100 |
| Accuracy class | 1.0 |
| Scale range | 0 ... 1 to 0 ... 60 bar [0 ... 15 to 0 ... 800 psi] or all other equivalent vacuum or combined pressure and vacuum ranges |
| Pressure limitation | Steady: Full scale value Fluctuating: 0.9 x full scale value Short time: 1.3 x full scale value |
| Permissible temperature range Medium Ambient Storage | -10 ... 150 °C [14 ... 302 °F] 10 ... 40 °C [50 ... 104 °F] 10 ... 60 °C [50 ... 140 °F] |
| Ingress protection | IP65 per IEC/EN 60529 |
| Material wetted non-wetted | Diaphragm: Stainless steel CrNi-Stahl 1.4435 [316L] or Hastelloy C276 Diaphragm seal: CrNi-Stahl 1.4435 [316L] or Hastelloy C276 Dial, pointer: Aluminium Case: Stainless steel 1.4301 [304] Diaphragm seal, pressure element: Stainless steel Window: Laminated safety glass |
| Level of cleanliness of wetted parts | Oil and grease free per ASTM G93-03 level F and ISO 15001 (< 1,000 mg/m²) |
| System fill fluid | Silicone oil KN 2 for general applications |
| Wetted parts | G | Dimensions in mm[in] | ||||
|---|---|---|---|---|---|---|
| NG | H | D | Mb | h | ||
| Stainless steel | G½ | 100 [3.937] | 169 [6.654] | 95 [3.740] | 52 [2.047] | 20 [0.787] |
| 1.4404/1.4435 [316L] | ½NPT | 168 [6.614] | 19 [0.748] | |||
| Hastelloy C276 | G½ | 203 [7.992] | 20 [0.787] | |||
| ½NPT | 202 [7.953] | 19 [0.748] |
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →104 standard articles of family DSS10
If you know the code — pick directly. If not — use the configurator in mode “01”.