AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
DI35-M 1
DI35-M
On order
WIKA · Pressure measurement · Family DI35

High-quality digital indicator for panel mounting

With multi-function input or with two inputs for standard signals
Output
4 ... 20 mA · 0 ... 20 mA
Accuracy
≤ ±0.01 ... 0.1 % of span ±1 digit
Protection
IP65/IP00
CEEACRoHSGOSTBelGIM
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The model DI35 digital indicator is a multi-function and very accurate instrument for a wide variety of measuring tasks. It is available in two different versions:

Special features

  • Multi-function input (29 calibrated input configurations) or double input (0/4 ... 20 mA, DC 0 ... 10 V) with calculation function
  • Accuracy ±0.01 ... 0.1 % of the span ±1 digit (depends on the input configuration)
  • Transmitter power supply, MIN/MAX memory, HOLD/TARE/Totaliser function
  • Linearisation with up to 30 programmable points
  • Up to four freely programmable switch contacts (optionally)
Identification
ModelDI35-M
Product familyDI35
SubcategoryDigital indicator
Accuracy
Accuracy class≤ ±0.01 ... 0.1 % of span ±1 digit
Accuracy standardEN 61326
Electrical characteristics
Output signal4 ... 20 mA · 0 ... 20 mA · DC 0 ... 10 V · 0-10V · RS-232 · RS-485
Power supplyAC 100 ... 240 V, 50/60 Hz, DC 100 ... 240 V or DC 10 ... 40 V, AC 18 ... 30 V, 50/60 Hz
Electrical connectionRemovable plug-in terminal
Switching functionFreely programmable
Switching contacts4
Materials
Case materialGlass-fibre reinforced polycarbonate, black, Glass-fibre reinforced polycarbonate
Operating environment
Ambient temperature+0…+50 °C
IP ratingIP65/IP00
Weight0.35 kg
Certifications and approvals
Certifications
CEEACRoHSGOSTBelGIM
Additional details
Datasheet languagesEN
Tables from datasheet
Selectable versions
StandardAC 100 ... 240 V, 50/60 Hz, DC 100 ...240V
OptionDC 10 ... 40 V, AC 18 ... 30 V, 50/60 Hz
Selectable versions
Option 1RS-232 (not galvanically isolated)
Option 2RS-232 (galvanically isolated)
Option 3RS-485 (not galvanically isolated, only for point- to-point connection)
Option 4RS-485 (galvanically isolated, only for point-to- point connection)
Input signalsMeasuring spanMeasuring error in %of the measuring span 1)Minimum measuring time
DI35-MDI35-D
1-channel measurement2-channel measurement
Current signals0 ...20mA≤ ±0.02 %±1digit0.02 s0.02 s0.04 s
4 ...20mA≤ ±0.02 %±1digit0.02 s0.02 s0.04 s
Voltage signalsDC 0 ...18mV≤ ±0.06 %±1digit0.02 s--
DC 0 ...35mV≤ ±0.06 %±1digit0.02 s--
DC 0 ...75mV≤ ±0.04 %±1digit0.02 s--
DC 0 ...150mV≤ ±0.03 %±1digit0.02 s--
DC 0 ...300mV≤ ±0.03 %±1digit0.02 s--
DC 0 ...600mV≤ ±0.03 %±1digit0.02 s--
DC 0 ...1,250mV≤ ±0.03 %±1digit0.02 s--
DC 0 ...2,500mV≤ ±0.03 %±1digit0.02 s--
DC 0 ...5V≤ ±0.02 %±1digit0.02 s--
DC 0 ...10V≤ ±0.01 %±1digit0.02 s0.02 s0.04 s
Thermocouples
Type B, PtRh-PtRh-100 ... +1,810 °C≤ ±0.10 %±1digit0.04 s--
Type E, NiCr-CuNi-260 ... +1,000 °C≤ ±0.06 %±1digit0.04 s--
Type J, Fe-CuNi-210 ... +1,200 °C≤ ±0.05 %±1digit0.04 s--
Type K, NiCr-Ni-250 ... +1,271 °C≤ ±0.05 %±1digit0.04 s--
Type L, Fe-CuNi-200 ...+900 °C≤ ±0.06 %±1digit0.04 s--
Type N, NiCrSi-NiSI-250 ... +1,300 °C≤ ±0.06 %±1digit0.04 s--
Type R, PtRh-Pt0 ... 1,760 °C≤ ±0.07 %±1digit0.04 s--
Type S, PtRh-Pt0 ... 1,760 °C≤ ±0.06 %±1digit0.04 s--
Type T, Cu-CuNi-240 ...+400 °C≤ ±0.07 %±1digit0.04 s--
Resistance thermometer 2)
Pt100 (2-/4-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.04 s--
Pt100 (3-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.06 s--
Pt200 (2-/4-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.04 s--
Pt200 (3-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.06 s--
Pt500 (2-/4-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.04 s--
Pt500 (3-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.06 s--
Pt1000 (2-/4-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.06 s--
Pt1000 (3-wire)-200 ...+850 °C≤ ±0.04 %±1digit0.04 s--
Input signalsMeasuring spanMeasuring error in %of the measuring span 1)Minimum measuring time
DI35-MDI35-D
1-channel measurement2-channel measurement
Current signals0 ...2mA≤ ±0.02 %±1digit0.02 s--
0 ...5mA≤ ±0.02 %±1digit0.02 s--
0 ...20mA≤ ±0.02 %±1digit0.02 s0.02 s0.04 s
4 ...20mA≤ ±0.02 %±1digit0.02 s0.02 s0.04 s
Voltage signalsDC -18 ...+18mV≤ ±0.06 %±1digit0.02 s--
DC -35 ...+35mV≤ ±0.06 %±1digit0.02 s--
DC -75 ...+75mV≤ ±0.04 %±1digit0.02 s--
DC -150 ...+150mV≤ ±0.03 %±1digit0.02 s--
DC -300 ...+300mV≤ ±0.03 %±1digit0.02 s--
DC -500 ...+600mV≤ ±0.03 %±1digit0.02 s--
DC -500 ...+1,250mV≤ ±0.03 %±1digit0.02 s--
DC -500 ...+2,500mV≤ ±0.03 %±1digit0.02 s--
DC -1 ...+5V≤ ±0.02 %±1digit0.02 s--
DC -1 ...+10V≤ ±0.01 %±1digit0.02 s0.02 s0.04 s
Resistance (2-, 3-, or 4-wire)0 ...100Ω≤ ±0.04 %±1digit0.04 s--
0 ... 1 kΩ≤ ±0.04 %±1digit0.04 s--
0 ... 10 kΩ≤ ±0.04 %±1digit0.04 s--
View all (11)
TerminalCase labellingMeaning
21S1Switch contact 1 1)Normally closed
22Normally open
23COM
24S2Switch contact 2 1)Normally closed
25Normally open
26COM
27S3Switch contact 3 1)Normally closed
28Normally open
29COM
30S4Switch contact 4 1)Normally closed
31Normally open
32COM
41GNDSerial interface RS232 1)
Serial interface RS485 1)
42TxDSerial interface RS232 1)
Data B(+)Serial interface RS485 1)
43RxDSerial interface RS232 1)
Data A(-)Serial interface RS485 1)
TerminalCase labellingMeaning
1Measuring input Resistance thermometer
2Measuring input Resistance thermometer
3Measuring input Resistance thermometer
+Voltage measuring signal ≤ 2.5V
+Thermocouple measuring signal
4Measuring input Resistance thermometer
-Voltage measuring signal ≤ 2.5V
-Thermocouple measuring signal
5+Voltage measuring signal
6+Current measuring signal
7-Voltage measuring signal
-Current measuring signal
8+Analogue output 1)
9-
10+Transmitter power supply
11-
12+Digital input
13LPower supply
14N
Plug A (input 1)
TerminalCase labellingMeaning
1+Current measuring signal Channel
2NCNot connected
3-Current measuring signal
4NCNot connected
Plug B (input 2)
TerminalCase labellingMeaning
1+Current measuring signal Channel
2NCNot connected
3-Current measuring signal
4NCNot connected
Plug C (switching outputs)
TerminalCase labellingMeaning
1S1Normally closed
2Normally open
3COM
4S2Normally closed
5Normally open
6COM
Plug D (digital output, analogue output, interface
TerminalMeaning
1Digital input, supply DC 24 V, ≤ 50mA
2Analogue output 0/4 ...20mA+
3Digital input +
4Analogue output 0/4 ... 20 mAand DC 0 ...10V -
5Analogue output 0 ...10V +
6GNDSerial interface RS232
Serial interface RS485
7TxDSerial interface RS232
Data B(+)Serial interface RS485
8RxDSerial interface RS232
Data A(-)Serial interface RS485
LogoDescriptionCountry
EU declaration of conformity ■ EMCdirective, EN 61326 emission (group 1, class B) and interference immunity (industrial application) ■ Low voltage directive ■ RoHS directiveEuropean union
EAC ■ Electromagnetic compatibility ■ Low voltage directiveEurasian Economic Com - munity
GOST Metrology, measurement technologyRussia
BelGIM Metrology, measurement technologyBelarus