AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
CTR6500 1
CTR6500
On order
WIKA · Calibration technology · Family CTR

AC resistance thermometry bridge

High-performance AC resistance thermometry bridge for very accurate temperature measurements in a range of -200 ... +962 °C
Range
-200…962 °C
CE
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The performance of the model CTR6500 resistance thermometry bridge is improved significantly by using the proven AC bridge technology. Due to the elimination of thermal EMF errors and the automatic cancellation of probe and cable reactance effects, the CTR6500 provides exceptional stability with time and ambient temperature and has an excellent resolution thanks to inherently low noise.

These make the CTR6500 perfectly suited to high-accuracy temperature measurement and calibration.

AC resistance thermometry bridge model CTR6500

The CTR6500 has 25 Ω and 100 Ω internal standard resistors and can also be used with external standard resistors. It has a ratio range of 0 ... 4.9999999 or 0 ... 500 Ω resistance and is capable of measuring temperature ranges to meet ITS 90, CVD and EN 60751 standards.

The results are shown on the large-scale, multi-function VFD screen. The mean, min and max value, standard deviation and the sampling rate can also be displayed with a separate graphical display. To ensure the long-term reliability the CTR6500 uses surface mount technology with no mechanical relays.

Special features

  • Resolution 0.1 ppm / 0.001 mΩ / 0.01 mK
  • 25 Ω and 100 Ω internal reference resistors
  • Channels expandable from 1 to 60 via multiplexers
  • Multifunction VFD with numeric, statistical or graphical information
Identification
ModelCTR6500
Product familyCTR
SubcategoryResistance thermometry bridge
Measuring range
Temperature ranges−200…+962 °C
Accuracy
Uncertainty±0.01 % (uncalibrated); Ratio range 0 ... 1 accuracy 0.4 ppm equivalent to 0.1 mK at 0 °C
Process conditions
Calibration typetemperature_bath
Operating environment
Ambient temperature+15…+25 °C
Weight9 kg
Certifications and approvals
CertificationsCE
Additional details
PortableNo
Datasheet languagesEN
Tables from datasheet
SpecificationsModel CTR6500
Probe typesIndustrial platinum resistance thermometers (PRTs) and standard platinum resistance thermometers (SPRTs) with Ro = 25 Ωand 100 Ωup to an alpha of 0.00392
Input channels2 on the main device (one PRT, SPRT or resistor + one reference resistor) 60 over multiplexer CTS9000
Input connections4 x BNC + shield (front panel)
Data entry formatITS 90 and CVD for calibrated probes; or EN 60751 for uncalibrated probes
Measuring ranges
Sense current100 µA, 200 µA, 500 µA, 1 mA, 2 mA, 5 mA, 10mA
Sense current mulitpliersx √2
Temperature range-200 ...+962 °C, depending on thermometer probe
Resistance range0…500Ω
Internal resistors
Values25 Ω, 100Ω
Thermal stabilityTCR±0.1 ppm/°C
Accuracy±0.01 %(uncalibrated)
Accuracy 1)Ratio range 0 ... 1 accuracy 0.4 ppm equivalent to 0.1 mKat 0 °C Ratio range 1 ... 2 accuracy 0.6 ppm equivalent to 0.3 mKfull range Ratio range 2 ... 3 accuracy 0.8 ppm equivalent to 0.8 mKfull range Ratio range 3 ... 5 accuracy 1 ppm equivalent to 1.25 mKfull range
Display
Screenlarge graphicVFD display screen (adjustable brightness)
UnitsRatio, °C, °F, K orΩ
Resolution0.1 ppm 0.01mK 0.001mΩ
Functions
Real-time clockintegrated clock with date
Voltage supply
Power supplyAC 90 ... 264 V, 47 ... 63 Hz; universal rear input on rear panel
Power consumptionmax.95VA
Permissible ambient conditions
Operating temperature15 ... 25 °C
Storage temperature-20 ...+50 °C
Communication
InterfaceUSB, RS-232 or IEEE-488.2
Case
Dimensions455 x 150 x 450mm(WxHxD)
Weight9 kg
CE conformity
EMCdirective2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (portable test and measuring equipment)
ModelDimensionsTemperature rangeDetector length
CTP5000-652Pt100, d = 6 mm, l = 450 mm(without spring strain relief, 100 mmhandle)-70 ...+650 °C30mm
CTP5000-651Pt100, d = 7.5 mm, l = 450 mm(125 mmhandle)-189 ...+650 °C50mm
ModelDimensionsTemperature rangeDetector length
CTP5000-T25Pt25, d = 6.5 ... 7.5 mm, l = 480mm-189 ...+660 °C45mm
View all (7)
KeyDescriptionFunction
Instrument function keys
DispSelect display typeAlternates numerical, graphical, standby display
ChanSelect input channelOpens and closes the Rt /R s channel select menu
ZeroZero display measurementOpens and closes the display zero (null) menu
HoldHold display measurementStarts and stops display (measurement continues)
CurrSelect operating currentOpens and closes the sensor-current menu
GainSelect instrument gainOpens and closes the instrument-gain menu
FilterSelect filter valueOpens and closes measurement-bandwidth menu
UnitsSelect display unitsOpens and closes the display-units menu
Shift keyShift keySelects lower function keys (and, Exp or +)
Instrument function shift keys
FuncSelect function menuSelects function menu
ProbeSelect probe menuOpens and closes probe menu
=0=Zero checkSelects zero check
=1=Unity checkSelects unity check
√2Set currentSets root 2 current
Auto/ManReserved turnsAutomatic gain selection ON/OFF
RsSelect reference resistorOpens the reference resistor menu
Menu function keys
BackClear data entry/returnClears any data entry errors or returns from a menu
MenuMenu selectionDisplays other submenus
▲▼◀▶Arrow keysUsed to navigate through the menus
OKSave entrySaves data entry and returns to previous menu
KeyDescriptionFunction
Numeric keys
0 ... 9Numerical data entryEnters a numerical digit or selects a numeric menu
-Minus keyUsed during numerical data entry
.Decimal pointUsed during numerical data entry
Numeric shift keys
ExpExponent keyUsed for numerical data entry (with the shift key)
+Plus keyUsed for numerical data entry (with the shift key)
SpecificationsPrecision resistor module, model FR4
Accuracy1 ... 10 Ω±0.005 %(50 ppm) > 10 ... 100 Ω±0.0025 %(25 ppm) > 100 ... 1,000 Ω±0.001 %(10 ppm)
Stability< 2 ppm per year
Temperature coefficient< 0.6 ppm/°C