AC resistance thermometry bridge
Description
The performance of the model CTR6500 resistance thermometry bridge is improved significantly by using the proven AC bridge technology. Due to the elimination of thermal EMF errors and the automatic cancellation of probe and cable reactance effects, the CTR6500 provides exceptional stability with time and ambient temperature and has an excellent resolution thanks to inherently low noise.
These make the CTR6500 perfectly suited to high-accuracy temperature measurement and calibration.
AC resistance thermometry bridge model CTR6500
The CTR6500 has 25 Ω and 100 Ω internal standard resistors and can also be used with external standard resistors. It has a ratio range of 0 ... 4.9999999 or 0 ... 500 Ω resistance and is capable of measuring temperature ranges to meet ITS 90, CVD and EN 60751 standards.
The results are shown on the large-scale, multi-function VFD screen. The mean, min and max value, standard deviation and the sampling rate can also be displayed with a separate graphical display. To ensure the long-term reliability the CTR6500 uses surface mount technology with no mechanical relays.
Special features
- Resolution 0.1 ppm / 0.001 mΩ / 0.01 mK
- 25 Ω and 100 Ω internal reference resistors
- Channels expandable from 1 to 60 via multiplexers
- Multifunction VFD with numeric, statistical or graphical information
| Model | CTR6500 |
| Product family | CTR |
| Subcategory | Resistance thermometry bridge |
| Temperature ranges | −200…+962 °C |
| Uncertainty | ±0.01 % (uncalibrated); Ratio range 0 ... 1 accuracy 0.4 ppm equivalent to 0.1 mK at 0 °C |
| Calibration type | temperature_bath |
| Ambient temperature | +15…+25 °C |
| Weight | 9 kg |
| Certifications | CE |
| Portable | No |
| Datasheet languages | EN |
| Specifications | Model CTR6500 |
|---|---|
| Probe types | Industrial platinum resistance thermometers (PRTs) and standard platinum resistance thermometers (SPRTs) with Ro = 25 Ωand 100 Ωup to an alpha of 0.00392 |
| Input channels | 2 on the main device (one PRT, SPRT or resistor + one reference resistor) 60 over multiplexer CTS9000 |
| Input connections | 4 x BNC + shield (front panel) |
| Data entry format | ITS 90 and CVD for calibrated probes; or EN 60751 for uncalibrated probes |
| Measuring ranges | |
| Sense current | 100 µA, 200 µA, 500 µA, 1 mA, 2 mA, 5 mA, 10mA |
| Sense current mulitpliers | x √2 |
| Temperature range | -200 ...+962 °C, depending on thermometer probe |
| Resistance range | 0…500Ω |
| Internal resistors | |
| Values | 25 Ω, 100Ω |
| Thermal stabilityTCR | ±0.1 ppm/°C |
| Accuracy | ±0.01 %(uncalibrated) |
| Accuracy 1) | Ratio range 0 ... 1 accuracy 0.4 ppm equivalent to 0.1 mKat 0 °C Ratio range 1 ... 2 accuracy 0.6 ppm equivalent to 0.3 mKfull range Ratio range 2 ... 3 accuracy 0.8 ppm equivalent to 0.8 mKfull range Ratio range 3 ... 5 accuracy 1 ppm equivalent to 1.25 mKfull range |
| Display | |
| Screen | large graphicVFD display screen (adjustable brightness) |
| Units | Ratio, °C, °F, K orΩ |
| Resolution | 0.1 ppm 0.01mK 0.001mΩ |
| Functions | |
| Real-time clock | integrated clock with date |
| Voltage supply | |
| Power supply | AC 90 ... 264 V, 47 ... 63 Hz; universal rear input on rear panel |
| Power consumption | max.95VA |
| Permissible ambient conditions | |
| Operating temperature | 15 ... 25 °C |
| Storage temperature | -20 ...+50 °C |
| Communication | |
| Interface | USB, RS-232 or IEEE-488.2 |
| Case | |
| Dimensions | 455 x 150 x 450mm(WxHxD) |
| Weight | 9 kg |
| CE conformity | |
|---|---|
| EMCdirective | 2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (portable test and measuring equipment) |
| Model | Dimensions | Temperature range | Detector length |
|---|---|---|---|
| CTP5000-652 | Pt100, d = 6 mm, l = 450 mm(without spring strain relief, 100 mmhandle) | -70 ...+650 °C | 30mm |
| CTP5000-651 | Pt100, d = 7.5 mm, l = 450 mm(125 mmhandle) | -189 ...+650 °C | 50mm |
| Model | Dimensions | Temperature range | Detector length |
|---|---|---|---|
| CTP5000-T25 | Pt25, d = 6.5 ... 7.5 mm, l = 480mm | -189 ...+660 °C | 45mm |
View all (7)
| Key | Description | Function |
|---|---|---|
| Instrument function keys | ||
| Disp | Select display type | Alternates numerical, graphical, standby display |
| Chan | Select input channel | Opens and closes the Rt /R s channel select menu |
| Zero | Zero display measurement | Opens and closes the display zero (null) menu |
| Hold | Hold display measurement | Starts and stops display (measurement continues) |
| Curr | Select operating current | Opens and closes the sensor-current menu |
| Gain | Select instrument gain | Opens and closes the instrument-gain menu |
| Filter | Select filter value | Opens and closes measurement-bandwidth menu |
| Units | Select display units | Opens and closes the display-units menu |
| Shift key | Shift key | Selects lower function keys (and, Exp or +) |
| Instrument function shift keys | ||
| Func | Select function menu | Selects function menu |
| Probe | Select probe menu | Opens and closes probe menu |
| =0= | Zero check | Selects zero check |
| =1= | Unity check | Selects unity check |
| √2 | Set current | Sets root 2 current |
| Auto/Man | Reserved turns | Automatic gain selection ON/OFF |
| Rs | Select reference resistor | Opens the reference resistor menu |
| Menu function keys | ||
| Back | Clear data entry/return | Clears any data entry errors or returns from a menu |
| Menu | Menu selection | Displays other submenus |
| ▲▼◀▶ | Arrow keys | Used to navigate through the menus |
| OK | Save entry | Saves data entry and returns to previous menu |
| Key | Description | Function |
|---|---|---|
| Numeric keys | ||
| 0 ... 9 | Numerical data entry | Enters a numerical digit or selects a numeric menu |
| - | Minus key | Used during numerical data entry |
| . | Decimal point | Used during numerical data entry |
| Numeric shift keys | ||
| Exp | Exponent key | Used for numerical data entry (with the shift key) |
| + | Plus key | Used for numerical data entry (with the shift key) |
| Specifications | Precision resistor module, model FR4 |
|---|---|
| Accuracy | 1 ... 10 Ω±0.005 %(50 ppm) > 10 ... 100 Ω±0.0025 %(25 ppm) > 100 ... 1,000 Ω±0.001 %(10 ppm) |
| Stability | < 2 ppm per year |
| Temperature coefficient | < 0.6 ppm/°C |