DC resistance thermometry bridge
Description
The model CTR6000 resistance thermometry bridge is a high-performance DC resistance thermometry bridge. The CTR6000 provides the latest technology in DC measurement. This way of measurement has always been a hallmark of ASL instruments and the CTR6000 continues that tradition.
The CTR6000 series features a large graphic display, combining versatility with ease of use. As well as the flexible menu options on the display, the CTR6000 also includes buttons dedicated to select common functions. For multichannel applications, the CTR6000 can be used with ASL's switch box systems CTS9000.
DC resistance thermometry bridge model CTR6000
Data can be recorded and stored on a PC through the ULog temperature measurement and data acquisition software.
The DC version provides a switched DC sensor current to simulate the advantages of AC measurement. In addition to measuring platinum resistance thermometers (PRTs), the CTR6000 also measures NTC thermistors.
Special features
- DC accuracy ±0.8 mK at 0 °C [32 °F]
- Resolution 1 ppm, 0.1 mK
- Channels expandable from 1 to 60 via multiplexers
- Measures ratio, Ω,°C, °F or K
- Large multi-function display
| Model | CTR6000 |
| Product family | CTR |
| Subcategory | Resistance thermometry bridge |
| Temperature ranges | −200…+962 °C |
| Uncertainty | ±0.8 mK at 0 °C, ±3 mK full range (temperature) |
| Calibration type | temperature_bath |
| Ambient temperature | +15…+25 °C |
| Weight | 7 kg |
| Certifications | CE |
| Portable | No |
| Datasheet languages | EN |
| Specifications | Model CTR6000 |
|---|---|
| Input channels | 2 on the main device (one platinum resistance thermometer (PRT) or thermistor + one reference resistor) 60 over multiplexer CTS9000 |
| Input connections | 4 x BNC + shield (front panel) |
| Data entry format | ITS 90 and CVD for calibrated probes; or EN 60751 for uncalibrated probes Steinhart and Hart for NTC thermistors |
| Measuring ranges | |
| Sense current | 10 µA, 20 µA, 100 µA, 200 µA, 1 mA, 2 mA, 10mA |
| Sense current mulitpliers | 0.5 and √2 |
| Temperature range | -200 ...+962 °C, depending on thermometer probe |
| Resistance range | 0 …500 kΩ |
| Internal resistors | |
| Values | 25 Ω, 100 Ω, 10 kΩ, 100 kΩ |
| Thermal stability | ±0.1 ppm/°C |
| Accuracy | ±0.01 %(uncalibrated) |
| Accuracy | |
| PRT 1) | ±3 ppm full range (ratio) ±3 ppm of reading or 0.0003 Ωwhichever greater (resistance) ±0.8 mKat 0 °C, ±3 mKfull range (temperature) |
| Thermistor | 0 ... 10 kΩ±10 ppm of reading or 0.01 Ω(whichever value is greater) 10 ... 50 kΩ ±20 ppm of reading 50 ... 100 kΩ ±40 ppm of reading 100 ... 500 kΩ ±100 ppm of reading |
| Display | |
| Screen | large graphicVFD display screen |
| Units | ratio, °C, °F, K orΩ |
| Resolution | 0.1mK |
| Functions | |
| Real-time clock | integrated clock with date |
| Voltage supply | |
| Power supply | AC 90 ... 264 V, 47 ... 63 Hz; universal rear input on rear panel |
| Power consumption | max.95VA |
| Permissible ambient conditions | |
| Operating temperature | 15 ... 25 °C |
| Storage temperature | -20 ...+50 °C |
| Communication | |
| Interface | USB, RS-232 or IEEE |
| Case | |
| Dimensions | 455 x 150 x 450mm(WxHxD) |
| Weight | 7 kg |
| Model | Dimensions | Temperature range | Detector length |
|---|---|---|---|
| CTP5000-652 | Pt100, d = 6 mm, l = 450 mm(without spring strain relief, 100 mmhandle) | -70 ...+650 °C | 30mm |
| CTP5000-651 | Pt100, d = 7.5 mm, l = 450 mm(125 mmhandle) | -189 ...+650 °C | 50mm |
| Model | Dimensions | Temperature range | Detector length |
|---|---|---|---|
| CTP5000-T25 | Pt25, d = 6.5 ... 7.5 mm, l = 480mm | -189 ...+660 °C | 45mm |
| Key | Description | Function |
|---|---|---|
| Function keys | ||
| Disp | Select display type | Alternates numerical, graphical, standby display |
| Chan | Select input channel | Opens and closes the Rs channel select menu |
| Zero | Zero display measurement | Opens and closes the display zero (null) menu |
| Hold | Hold display measurement | Starts and stops display (measurement continues) |
| Curr | Select operating current | Opens and closes the sensor-current menu |
| Gain | Select instrument gain | Opens and closes the instrument-gain menu |
| Filter | Select filter value | Opens and closes measurement-bandwidth menu |
| Units | Select display units | Opens and closes the display-units menu |
| Shift key | Shift key | Selects lower function keys (and, Exp or +) |
| Function shift keys | ||
| Func | Select function menu | Selects function menu |
| Probe | Select probe menu | Opens and closes probe menu |
| √2 | Set current | Sets root 2 current |
| Auto/Man | Reserved turns | Automatic gain selection ON/OFF |
| Rs | Select reference resistor | Opens the reference resistor menu |
| Menu function keys | ||
| Back | Clear data entry/return | Clears any data entry errors or returns from a menu |
| Menu | Menu selection | Displays other submenus |
| ▲▼◀▶ | Navigation keys | Used to navigate through the menus |
| OK | Save entry | Saves data entry and returns to previous menu |
View all (6)
| Key | Description | Function |
|---|---|---|
| Numeric keys | ||
| 0 ... 9 | Numerical data entry | Enters a numerical digit or selects a numeric menu |
| - | Minus key | Used during numerical data entry |
| . | Decimal point | Used during numerical data entry |
| Numeric shift keys | ||
| Exp | Exponent key | Used for numerical data entry (with the shift key) |
| + | Plus key | Used for numerical data entry (with the shift key) |
| Specifications | Precision resistor module, model FR4 |
|---|---|
| Value | 1, 10, 25 and 100Ω |
| Accuracy | 1 ... 10 Ω±0.005 %(50 ppm) > 10 ... 100 Ω±0.0025 %(25 ppm) > 100 ... 1,000 Ω±0.001 %(10 ppm) |
| Stability | < 2 ppm per year |
| Temperature coefficient | < 0.6 ppm/°C |
Couldn't find the answer in the documentation? An engineer will reply within a business day.