AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
CTD9100-ZERO 1
CTD9100-ZERO
On order
WIKA · Calibration technology · Family CTD9100

Temperature dry-well calibrator

Portable temperature dry-well calibrator for simple calibration of thermometers at 0 °C and tempering of thermocouple cold junctions.
Range
-10…100 °C
RoHSCE
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Special features

  • Low-cost
  • Low weight and compact design
  • Simple operation
Identification
ModelCTD9100-ZERO
Product familyCTD9100
SubcategoryDry-well calibrator
Measuring range
Temperature ranges−10…+100 °C
Accuracy
Uncertainty0.05 K at 0 °C, otherwise 0.1 K
Process conditions
Calibration typedry_block
Operating environment
Ambient temperature−10…+100 °C
Weight7 kg
Certifications and approvals
Certifications
RoHSCE
TraceabilityDKD · DAkkS
Additional details
PortableYes
Datasheet languagesEN
Tables from datasheet
CTD9100-ZERO
Indication
Display4-digit LED display
Temperature range-10 …0…+100°C[14 ... 32 ... 212 °F]
Accuracy 1)0.05 K at 0 °C [32 °F] otherwise 0.1 K
Stability 2)< 0.05 K
Display resolution0.1 °C
Temperature distribution
Axial homogeneity 3)dependent on temperature, temperature probes and their quantity
Temperature control
Heating time15 min from -10 °C to +100 °C (from 14 °F to 212 °F)
Cooling time5 min from 23 °C to 0 °C (from 73 °F to 14 °F) 10 min from 100 °C to 0 °C (from 212 °F to 14 °F)
Stabilisation time 4)dependent on temperature and temperature probe
Insert
Immersion depth150 mm[5.91 in]
Test bores7 bores with Ø6.5 mm[0.26 in]
Voltage supply
Power supply 5)AC 100 ... 240 V, 50/60 Hz
Power consumption225VA
Fuse3.15 A slow blow fuse (at AC 250 V)
Power cordAC 230 V; for Europe
Communication
InterfaceRS-485
Case
Dimensions (W x D x H)160 x 230 x 320+50 mm[6.3 x 9.1 x 12.6+2 in]
Weight7 kg [15.5 lbs]
LogoDescriptionCountry
EU declaration of conformity ■ EMCdirective ■ Low voltage directive ■ RoHS directiveEuropean Union
GOST (option) Metrology, measurement technologyRussia
KazInMetr (option) Metrology, measurement technologyKazakhstan
-MTSCHS (option) Permission for commissioningKazakhstan
BelGIM (option) Metrology, measurement technologyBelarus
Certificate
CalibrationStandard: 3.1 calibration certificate per DIN EN 10204 Option:DKD/DAkkS calibration certificate
Recommended recalibration interval1 year (dependent on conditions of use)
AccessoriesOrder code
DescriptionCTX-A-K6
Power cord for EU-EU-
for Switzerland-CH-
for UK-UK-
for USA/Canada-US-
Calibration software-CS-
Interface converter RS-485 to USB 2.0-RC-
Ordering information for your enquiry:
1. Order code: CTX-A-K6 2. Option:⇓ [ ]