WIKA · Calibration technology · Family CPC
Pressure controller
High-end pressure controller with premium accuracy, suitable as a factory or working standard for calibration of pressure measuring instruments, featuring patented valve technology and modular transducer design.
Range
-1…401 bar
Protection
None
RoHSCEUKCA
Price
On request
Depends on configuration
Special features
- Pressure ranges: -1 ... 400 bar [-15 ... 6,000 psi]
- Control stability 0.002 % of the span
- Accuracy down to 0.008 % IS (IntelliScale)
- Precision 0.004 % FS
Identification
| Model | CPC8000 |
| Product family | CPC |
| Subcategory | Pressure controller |
Measuring range
| Pressure ranges | -1…400 bar · 0…400 bar · 0…0.35 bar · 0…0.5 bar · 0…401 bar · 0…100 bar · -1…1 bar |
| Temperature ranges | +15…+45 °C · +0…+70 °C |
Accuracy
| Uncertainty | 0.008 % of reading (IS) |
Process conditions
| Medium | dry, clean air, nitrogen |
| Calibration type | deadweight |
Operating environment
| Ambient temperature | +15…+40 °C |
| Weight | 22.2 kg |
Certifications and approvals
| Certifications | RoHSCEUKCA |
| Traceability | A2LA · DAkkS |
Additional details
| Portable | No |
| Datasheet languages | EN, DE, IT, FR, NL, PL, PT, RO, ES, SV, RU, EL, JA, ZH |
Tables from datasheet
Reference pressure transducer model CPR8000
| Pressure range | |||
|---|---|---|---|
| Accuracy 1) | 0.008%FS 2) | 0.008 %IS-50 3) | 0.008 %IS-33 4) |
| Gauge pressure 5) | 0 ... 0.35 up to 0 ... 400 bar [0 ... 5 up to 0 ... 6,000 psi] | 0 ... 1 up to 0 ... 400 bar [0 ... 15 up to 0 ... 6,000 psi] | 0 ... 1 up to 0 ... 100 bar [0 ... 15 up to 0 ... 1,500 psi] |
| Bidirectional | -1 ...+1 up to -1 ... 400 bar [-15 ...+15 up to -15 ... 6,000 psi] | -1 ... 10 up to -1 ... 400 bar [-15 ... 145 up to -15 ... 6,000 psi] | -1 ... 10 up to -1 ... 100 bar [-15 ... 145 up to -15 ... 1,500 psi] |
| Absolute pressure 6) | 0 ... 0.5 up to 0 ... 401 bar abs. [0 ... 7.5 up to 0 ... 6,015 psi abs.] | 0 ... 1 up to 0 ... 401 bar abs. [0 ... 15 up to 0 ... 6,015 psi abs.] | 0 ... 1 up to 0 ... 101 bar abs. [0 ... 15 up to 0 ... 1,515 psi abs.] |
| Precision 7) | 0.004%FS | ||
| Calibration interval | 365 days 8) | 365 days | 365 days |
Barometric reference
| Measuring range | ■ 552 ... 1.172 mbar abs. ■ 8 ... 17 psi abs. ■ 552 ... 1.172 hPa abs. |
|---|---|
| Accuracy 2) | 0.01 %of reading |
| Function | The barometric reference can be used to switch pressure types 2) , absolute <=> gauge. With gauge pressure transducers, the measuring range of the transducers must begin with -1 bar [-15 psi] in order to carry out an absolute pressure emulation. |
| Communication | |
|---|---|
| Interface | ■ Ethernet ■ IEEE-488 ■ USB ■ RS-232 |
| Baud rate | ■ 9600 ■ 19200 ■ 38400 ■ 57600 ■ 115200 |
| Command sets | ■ Mensor ■ WIKA SCPI ■ Others on request |
| Response time | < 100 ms |
| Digital I/O Digital input | DC 3.3V or DC 5 V; current limited by 330 Ωresistor |
| Digital output | 0.5 A at AC 125V |
Voltage supply
| Operating voltage | ■ AC 100 ... 120 V, 50/60 Hz ■ AC 220 ... 240 V, 50/60 Hz |
|---|---|
| Power consumption | Max.130VA |
| Supply voltage fluctuation | ±10% |
| Fuse | 1.6 A, 250V;SLO-BLO5x20mm |
View all (12)
Operating conditions
| Place of use | Indoor Not for wet locations |
|---|---|
| Altitude | Up to 3,048 m[10,000 ft] above sea level |
| Operating temperature | 15 ... 40 °C [59 ... 104 °F] |
| Compensated temperature range | 15 ... 45 °C [59 ... 113 °F] |
| Storage temperature | 0 ... 70 °C [32 ... 158 °F] |
| Relative humidity, condensation | 0 ... 95 %r.h. (non-condensing) |
| Mounting position of transducers | Horizontal or slightly tilted |
| Permissible pollution degree | Degree 2 |
| EMC (HF field) | EN 61326-1 emission (group 1, class A) and immunity (industrial application) |
| Logo | Description | Region |
|---|---|---|
| EU declaration of conformity | European Union | |
| EMCdirective 1) EN 61326 emission (group 1, class A) and immunity (industrial application) | ||
| LowVoltage Directive | ||
| RoHS directive | ||
| UKCA | United Kingdom | |
| Electromagnetic compatibility regulations | ||
| Electrical equipment designed for use within certain voltage limits in support of the electrical equipment (safety) regulations | ||
| Restriction of hazardous substances (RoHS) regulations |
| Logo | Description | Region |
|---|---|---|
| - | MChS Permission for commissioning | Kazakhstan |
| - | PAC China Metrology, measurement technology | China |
| Logo | Description |
|---|---|
| - | China RoHS directive |
Certificates
| Calibration 1) | |
|---|---|
| Barometric reference | ■ Without ■ A2LA calibration certificate (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate for barometric reference (traceable and accredited in accordance with ISO/IEC 17025) |
| Reference pressure transducer model CPR8000 | ■ A2LA calibration certificate (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate - gauge pressure (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate - absolute pressure (traceable and accredited in accordance with ISO/IEC 17025) |
| Recommended calibration interval | 1 year (dependent on conditions of use) |
| Cal-Template (demo version) | Cal-Template (light version) | Cal-Template (full version) | Log-Template (full version) |
|---|---|---|---|
| Fully automatic calibration | Semi-automatic calibration | Fully automatic calibration | ■ Live measured value recording for a certain period of time with selectable interval, duration and start time ■ Creation of logger protocols with graphic and/or tabular representation of the measuring results in PDF format ■ Possibility of exporting measuring results as CSV file |
| Limitation to two measuring points | No limitation of the measuring points approached | ||
| ■ Creation of 3.1 inspection certificates per DIN EN 10204 ■ Calibration data can be exported to Excel ® template or XML file ■ Calibration of pressure measuring instruments | |||
| Ordering information for a single licence | |||
| Is available for a cost-free download | WIKA-CAL-LZ-Z-Z | WIKA-CAL-CZ-Z-Z | WIKA-CAL-ZZ-L-Z |
| Ordering information for a pair licence | |||
| Cal-Template (light version) together with Log-Template (full version) | WIKA-CAL-LZ-L-Z | ||
| Cal-Template (full version) together with Log-Template (full version) | WIKA-CAL-CZ-L-Z |
| Description 1) | Order code | |
|---|---|---|
| CPX-A-C8 | ||
| - | Desktop case | -D- |
| - | 19" rack mount case With side pieces, EU With side pieces,NAM | -R- -U- |
| Barometric reference Measuring range: 8 ... 17 psi abs. Accuracy to 0.01 %of reading | -3- | |
| Measuring range: 552 ... 1,172 mbar abs. Accuracy to 0.01 %of reading | -K- | |
| Measuring range: 552 ... 1,172 hPa abs. Accuracy to 0.01 %of reading | -L- | |
| Calibration adapter For reference pressure transducers, voltage supply and software | -4- | |
| Calibration adapter For barometric reference, voltage supply and software | -5- | |
| Transport case | -6- | |
| Silencer | -7- | |
| RS-232 interface cable | -9- | |
| - | Vacuum pump | -2- |
| Adapter set 6 mmSwagelok ® male thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -M- | |
| Adapter set 6 mmSwagelok ® male thread (4 adapters) Max. 400 bar [6,000 psi] Material: stainless steel | -C- | |
| Adapter set ¼" tube fitting (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -I- | |
| Adapter set ¼" tube fitting (4 adapters) Max. 400 bar [6,000 psi] Material: stainless steel | -E- |
| Description 1) | Order code | |
|---|---|---|
| CPX-A-C8 | ||
| Adapter set ⅛BSPG, female thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -B- | |
| Adapter set ¼NPT, female thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -N- | |
| Adapter set ¼NPT, female thread (4 adapters) Max. 400 bar [6,000 psi] Material: stainless steel | -A- | |
| Adapter set ⅛NPT, female thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -S- | |
| Adapter set ⅛NPT, female thread (4 adapters) Max. 400 bar [6,000 psi] Material: stainless steel | -F- | |
| Ordering information for your enquiry: | ||
| 1. Order code: CPX-A-C8 2. Option: | ⇓ [ ] |
Industry
Machinery
Transmitter and pressure gauge manufacturers
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →