Range
-1…211 bar
Protection
IP20
CERoHSUKCA
Price
On request
Depends on configuration
Special features
- Pressure ranges: -1 ... 210 bar [-15 ... 3,045 psi]
- Control speed 15 s
- Control stability < 0.003 % FS
- Accuracy down to 0.008 % IS (IntelliScale)
- Precision 0.004 % FS
Identification
| Model | CPC6050 |
| Product family | CPC |
| Subcategory | Pressure controller |
Measuring range
| Pressure ranges | -1…210 bar · 0…211 bar · 0…0.025 bar · 0…210 bar · -0.012…0.012 bar · 0…0.5 bar |
| Temperature ranges | +0…+50 °C · −20…+70 °C · +15…+45 °C |
Accuracy
| Uncertainty | 0.008 % IS (IntelliScale) |
Process conditions
| Medium | air |
| Calibration type | electronic |
Operating environment
| Ambient temperature | +0…+50 °C |
| IP rating | IP20 |
| Weight | 22.7 kg |
Certifications and approvals
| Certifications | CERoHSUKCA |
| Traceability | A2LA · DAkkS |
Additional details
| Portable | No |
| Datasheet languages | EN, DE, ES, FR, IT, NL, PL, PT, RO, SV, RU, EL, JA, ZH |
Tables from datasheet
Reference pressure transducer model CPR6050
| Pressure range | Standard | ||
|---|---|---|---|
| Accuracy 1) | 0.01%FS 2) | ||
| Gauge pressure 6) | 0 ... 0.025 to 0 ... 210 bar [0 ... 0.36 to 0 ... 3,045 psi] | ||
| Bidirectional pressure 6) | -0.012 ... +0.012 to -1 ... 210 bar [-0.18 ... +0.18 to -15 ... 3,045 psi] | ||
| Absolute pressure 7) | 0 ... 0.5 to 0 ... 211 bar abs. [0 ... 7.5 to 0 ... 3,060 psi abs.] | ||
| Precision 8) | 0.004%FS | ||
| Calibration interval | 365 days 9) | ||
| Pressure range | Optional | ||
| Accuracy 1) | 0.008%FS | ■ 0.008 %IS-50 3) ■ 0.01 %IS-50 4) | 0.008 %IS-33 5) |
| Gauge pressure 6) | 0 ... 0.025 to 0 ... 210 bar [0 ... 0.36 to 0 ... 3,045 psi] | 0 ... 1 to 0 ... 210 bar [0 ... 15 to 0 ... 3,045 psi] | 0 ... 1 to 0 ... 100 bar [0 ... 15 to 0 ... 1,500 psi] |
| Bidirectional pressure 6) | -0.012 ... +0.012 to -1 ... 210 bar [-0.18 ... +0.18 to -15 ... 3,045 psi] | -1 ... 10 to -1 ... 210 bar [-15 ... 145 to -15 ... 3,045 psi] | -1 ... 10 to -1 ... 100 bar [-15 ... 145 to -15 ... 1,500 psi] |
| Absolute pressure 7) | 0 ... 0.5 to 0 ... 211 bar abs. [0 ... 7.5 to 0 ... 3,060 psi abs.] | 0 ... 1 to 0 ... 211 bar abs. [0 ... 15 to 0 ... 3,060 psi abs.] | 0 ... 1 to 0 ... 101 bar [0 ... 15 to 0 ... 1,515 psi] |
| Calibration interval | 365 days | 365 days | 365 days |
| Precision 8) | 0.004%FS | 0.004%FS | 0.004%FS |
Barometric reference, optional
| Measuring range | ■ 552 ... 1,172 mbar abs. ■ 8 ... 17 psi abs. ■ 552 ... 1,172 hPa abs. |
|---|---|
| Accuracy 1) | 0.01 %of reading |
| Function | The barometric reference can be used to switch pressure types 2) , absolute <=> gauge.With gauge pressure transducers, the measuring range of the transducers must begin with -1 bar [-15 psi] in order to carry out a complete absolute pressure emulation. |
Pressure controller CPC6050
| Instrument | |
|---|---|
| Instrument version | ■ Desktop case ■ 19" rack-mounting kit |
| Dimensions | See technical drawings |
| Weight | Approx. 22.7 kg [50 lb] incl. all internal options |
| Warm-up time | Approx. 15 min |
| Digital display | |
| Type of display | 10.1" colour LC display with capacitive touchscreen |
| Display resolution | 4 ... 6 digits depending on range and units |
| Measuring range | -0.012 ... +0.012 to -1 ... 210 bar [-0.18 ... +0.18 to -15 ... 3,045 psi] Depending on the reference pressure transducer model CPR6050 |
| Pressure type | ■ Gauge ■ Absolute ■ Bidirectional |
| Unit | 39 and two freely programmable pressure units |
| Pressure limitation (vacuum and overpressure) | |
| Measure/Control port | 100 %of the primary sensor range |
| Supply port | 110 %of primary sensor range |
| Reference port | Atmosphere |
| Vent port | Atmosphere |
| Exhaust port | Ambient pressure down to full vacuum |
| Overpressure limit | |
| Measure/Control port | 105 %of primary sensor range |
| Supply port | 110 %of CPMrange |
| Reference port | Atmosphere ±350 mbar [±5 psi] |
| Vent port | Atmosphere |
| Exhaust port | Full vacuum |
| Control parameter | SVR module 1) | LPPump module |
|---|---|---|
| Control stability | < 0.003 %FSof the active range (typical 0.001%FS 2) ) | |
| Control mode | Precision, high speed and customised | External supply ON/OFF |
| Control speed | 15 s 3) | 25 s 3) |
| Control range | 0 ...100%FS | |
| Minimum control pressure | 0.0017 bar [0.025 psi] over exhaust pressure or 0.05 %FSwhichever is greater | 0.0034 bar [0.05 psi] over exhaust pressure or 0.05 %FSwhichever is greater |
| Overshoots | < 1 %FSin high-speed control mode (typical < 0.05 %FSin precision control mode) | < 1 %FSin high-speed control mode (< 0.1 %FSin pump only mode) |
| Test volume | 50 ... 1,000 ccm | 50 ... 300 ccm |
View all (15)
Communication
| Interface | ■ Ethernet ■ IEEE-488 ■ USB ■ RS-232 |
|---|---|
| Communication protocol | 10/100Based-T |
| Baud rate | ■ 9600 ■ 19200 ■ 38400 ■ 57600 ■ 115200 |
| Command sets | ■ Mensor ■ WIKA SCPI ■ Others on request |
| Response time | Approx. 100 ms |
| Measuring rate | 30 ... 60 ms |
| Internal program | Up to 24 sequences with up to 99 steps each |
Pressure connection on the CPC6050
| Connections | ■ Up to 8 ports with 7/16"- 20 F SAE ■ Up to 2 ports with 1/8" F NPT ■ 1 port with 10-32 UNF female | ||
|---|---|---|---|
| Filter elements | All pressure ports have 40-micron filters. | ||
| Pressure port adapters | ■ Without ■ 6 mmtube fitting ■ ¼" tube fitting | ■ ¼NPT, female thread ■ ⅛NPT, female thread ■ ⅛BSP, female thread | |
| Barometer port adapters | ■ Barb fitting ■ 6 mmtube fitting ■ ¼" tube fitting | ||
| Wetted parts | ■ Aluminium ■ Brass ■ Buna N ■ Urethane ■ FKM/FPM | ■ PCTFE ■ PEEK ■ PTFE ■ PPS | ■ RTV ■ Ceramic ■ Silicone ■ Silicone grease ■ Stainless steel 316 and 316L ■ Glass-filled epoxy |
| Overpressure protection | Safety relief valve fixed to reference pressure transducer and adjusted to customised |
Voltage supply
| Operating voltage | ■ AC 100 ... 120 V, 50/60 Hz ■ AC 220 ... 240 V, 50/60 Hz |
|---|---|
| Power consumption | Max.210VA |
| Overvoltage resistance | Category II |
| Electrical safety | Protection class 1 (PE connected) |
| Fuse | 1.6 A, 250V;SLO-BLO5x20mm |
| Power cord | ■ For Europe ■ For USA/Canada ■ For UK ■ For India ■ For China |
Operating conditions
| Altitude | Up to 3,048 m[10,000 ft] above sea level |
|---|---|
| Place of use | Indoor |
| Operating temperature | 0 ... 50 °C [32 ... 122 °F] |
| Compensated temperature range | 15 ... 45 °C [59 ... 113 °F] |
| Storage temperature range | -20 ...+70 °C [-4 ...+158 °F] |
| Relative humidity, condensation | 5 …95%r.h.(non-condensing) |
| Permissible media | ■ Dry, clean air ■ Nitrogen (ISO 8573-1:2010 class 5.5.4 or better) |
| Mounting position | Horizontal |
| EMC (HF field) | EN 61326-1 emission (group 1, class A) and immunity (industrial application) |
| Logo | Description | Region |
|---|---|---|
| EU declaration of conformity | European Union | |
| EMCdirective 1) EN 61326-1 emission (group 1, class A) and immunity (industrial application) | ||
| Low voltage directive | ||
| RoHS directive | ||
| United Kingdom | UKCA | |
| Electromagnetic compatibility regulations | ||
| Electrical equipment designed for use within certain voltage limits in support of the electrical equipment (safety) regulations | ||
| Restriction of hazardous substances (RoHS) | regulations |
| Logo | Description | Region |
|---|---|---|
| - | MChS Permission for commissioning | Kazakhstan |
Certificate
| Calibration 1) | |
|---|---|
| Reference pressure transducer model CPR6050 | ■ A2LA calibration certificate (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate - gauge pressure (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate - absolute pressure (traceable and accredited in accordance with ISO/IEC 17025) |
| Barometric reference | ■ Without ■ A2LA calibration certificate (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate for barometric reference (traceable and accredited in accordance with ISO/IEC 17025) |
| Recommended calibration interval | 1 year (dependent on conditions of use) |
Model A-CPS
| Operating conditions | |
|---|---|
| Maximum operating pressure | 211 bar abs. [3,065 psi abs.] |
| Maximum operating temperature | 80 °C [176 °F] |
| Voltage supply | |
| Power supply | DC 12V |
| Power consumption | 13VA |
| Pressure connection | |
| To the Measure/Control port of CPC6050 | 1 port with ¼" tube adapted to 7/16"- 20 F SAE |
| To the test item | 2 ports: ■ 7/16" - 20 F SAE ■ 6 mmtube fitting ■ ¼' tube fitting ■ ¼NPT, female thread ■ ⅛NPT, female thread ■ ⅛BSP, female thread |
| Dimensions | |
| Dimensions (W x H x D) | 139.7 x 266.7 x 139.7 mm[5.5 x 10.5 x 5.5 in] |
| Weight | 3.9 kg [8.8 lb] |
| Cal-Template (demo version) | Cal-Template (light version) | Cal-Template (full version) | Log-Template (full version) |
|---|---|---|---|
| Fully automatic calibration | Semi-automatic calibration | Fully automatic calibration | ■ Live measured value recording for a certain period of time with selectable interval, duration and start time ■ Creation of logger protocols with graphic and/or tabular representation of the measuring results in PDF format ■ Possibility of exporting measuring results as CSV file |
| Limitation to two measuring points | No limitation of the measuring points approached | ||
| ■ Creation of 3.1 inspection certificates per DIN EN 10204 ■ Calibration data can be exported to Excel ® template or XML file ■ Calibration of pressure measuring instruments | |||
| Ordering information for your enquiry for a single licence: | |||
| Is available for a cost-free download | WIKA-CAL-LZ-Z-Z | WIKA-CAL-CZ-Z-Z | WIKA-CAL-ZZ-L-Z |
| Ordering information for your enquiry for a pair licence: | |||
| Cal-Template (light version) together with Log-Template (full version) | WIKA-CAL-LZ-L-Z | ||
| Cal-Template (full version) together with Log-Template (full version) | WIKA-CAL-CZ-L-Z |
| Order code | |
|---|---|
| CPX-A-C5 | |
| 19" built-in case With side pieces,NAM With side pieces, EU | -U- -T- |
| Barometric reference Measuring range: 8 ... 17 psi abs. Accuracy to 0.01 %of reading | -3- |
| Measuring range: 552 ... 1,172 mbar abs. Accuracy to 0.01 %of reading | -K- |
| Measuring range: 552 ... 1,172 hPa abs. Accuracy to 0.01 %of reading | -L- |
| Calibration adapter For reference pressure sensor, voltage supply and software | -4- |
| Calibration adapter For barometric reference, voltage supply and software | -5- |
| Transport case | -6- |
| Adapter set Consisting of: 4 adapters with ⅛BSPG, female thread Material: brass | -B- |
| Adapter set Consisting of: 4 adapters with ¼" tube fitting Material: brass | -I- |
| Adapter set Consisting of: 4 adapters with 6 mmSwagelok ® male thread Material: brass | -M- |
| Adapter set Consisting of: 4 adapters with ¼NPT, female thread Material: brass | -N- |
| Adapter set Consisting of: 4 adapters with ⅛NPT, female thread Material: brass | -S- |
| Block and bleed valve Pressure range: ≤ 400 bar [6,000 psi] | -8- |
| Coalescing filter Pressure range: ≤ 240 bar [3,600 psi] | -9- |
| Accessories for CPC6050 1) | Order code | |
|---|---|---|
| Description | CPX-A-C5 | |
| Automatic contamination protection (A-CPS) Pressure range: ≤ 100 bar [1,500 psi] | -A- | |
| Pressure range: ≤ 210 bar [3,045 psi] | -O- | |
| Replacement filters for automatic CPS | -2- | |
| Vacuum regulator for low-pressure ranges | -1- | |
| Ordering information for your enquiry: | ||
| 1. Order code: CPX-A-C5 2. Option: | ⇓ [ ] |
Industry
Machinery
Transmitter and pressure measuring instrument manufacturers
Couldn't find the answer in the documentation? An engineer will reply within a business day.