AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
CPB5000 1
CPB5000
On order
WIKA · Calibration technology · Family CPB

Pressure balance

Pneumatic version to 120 bar [1,600 lb/in 2 ]
Range
-1…120 bar
Protection
IP54
EACPressure Equipment DirectiveMachinery DirectiveMChS
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Special features

  • Total measurement uncertainty starting from 0.015 ... 0.006 % of reading
  • Four different piston-cylinder systems available for ranges ±1 bar, 0.1 ... 7 bar, 0.2 ... 25 bar, 1 ... 70 bar and 1 ... 120 bar, or also for ranges in lb/in 2 and kPa
  • Quick and safe replacement of the piston-cylinder system for measuring range changes via patented ConTect quick-release system possible
  • Masses manufactured from stainless steel and aluminium, can be adjusted to local gravity
  • Factory calibration included as standard, traceable to national standards; with UKAS calibration possible
Identification
ModelCPB5000
Product familyCPB
Subcategorypneumatic pressure balance
Measuring range
Pressure ranges-1…120 bar · 0.015…1 bar · 0.015…2 bar · 0.1…7 bar · 0.2…25 bar · 1…70 bar · 1…120 bar · 0.2…15 lb/in² · 0.2…30 lb/in² · 1…100 lb/in² · 3…400 lb/in² · 15…1000 lb/in² · 10…1600 lb/in² · 1.5…100 kPa · 1.5…200 kPa · 10…700 kPa · 20…2500 kPa · 100…7000 kPa · 100…12000 kPa · -14.5…1600 lb/in²
Accuracy
Uncertainty0.015 % of reading (standard), 0.006 % of reading (premium)
Process conditions
Mediumgas
Calibration typedeadweight
Operating environment
Ambient temperature+18…+28 °C
IP ratingIP54
Weight62 kg
Certifications and approvals
Certifications
EACPressure Equipment DirectiveMachinery DirectiveMChS
TraceabilityUKAS · DAkkS · NIST
Additional details
PortableYes
Datasheet languagesEN, DE, ES, FR, IT
Tables from datasheet
Measuring range [bar] 1)0.015 ... 10.015 ... 20.1 ... 70.2 ... 251 ... 701 ... 120
QuantityNominal pressure per piece [mbar]QuantityNominal pressure per piece [mbar]QuantityNominal pressure per piece [bar]QuantityNominal pressure per piece [bar]QuantityNominal pressure per piece [bar]QuantityNominal pressure per piece [bar]
Piston and make-up weight10.01510.01510.110.21111
Standard mass set10.00510.00510.410.32121
30.0230.0220.514.5510118
20.0120.01513519420
60.0560.0520.22224110
60.160.110.1111224
--1110.0510.510.512
--------10.5
Set of fine increment masses----20.0220.210.410.4
----10.0110.110.210.2
----10.00510.0510.110.1
------20.0220.0420.04
------10.0110.0210.02
Measuring range [lb/in²] 1)0.2 ... 150.2 ... 301 ... 1003 ... 40015 ... 1,00010 ... 1,600
QuantityNominal pressure per piece [lb/in²]QuantityNominal pressure per piece [lb/in²]QuantityNominal pressure per piece [lb/in²]QuantityNominal pressure per piece [lb/in²]QuantityNominal pressure per piece [lb/in²]QuantityNominal pressure per piece [lb/in²]
Piston10.210.21113110110
Standard mass set10.0510.051417210210
10.110.12519011801180
20.220.2810210032006200
10.510.52215011001100
10.810.811220240240
111110.5110120120
2222--151515
2424--------
--115--------
Set of fine increment masses----20.2221414
----10.1111212
----10.0510.51111
------20.220.420.4
------10.110.210.2
Measuring range [kPa] 1)1.5 ... 1001.5 ... 20010 ... 70020 ... 2,500100 ... 7,000100 ... 12,000
QuantityNominal pressure per piece [kPa]QuantityNominal pressure per piece [kPa]QuantityNominal pressure per piece [kPa]QuantityNominal pressure per piece [kPa]QuantityNominal pressure per piece [kPa]QuantityNominal pressure per piece [kPa]
Piston and make-up weight11.511.511012011001100
Standard mass set10.510.514013021002100
3232250145051,00011,800
212151003500190042,000
65652202200240011,000
610610110110012002400
--1100151501501200
----------150
Set of fine increment masses----22220140140
----11110120120
----10.515110110
------222424
------111212
Piston-cylinder system
Measuring range in bar 1)0.015…10.015…20.1…70.2…251…701 …120
Required masses3.3 kg6.54 kg22.5 kg21 kg29 kg49.5 kg
Smallest step 2) (Standard mass set)0.005 bar0.005 bar0.05 bar0.3 bar0.5 bar0.5 bar
Smallest step 3) (Set of fine increment masses)----0.005 bar0.01 bar0.02 bar0.02 bar
Measuring range in lb/in² 1)0.2 ... 150.2 ... 301 …1003 …40015 …1,00010 …1,600
Required masses3.3 kg6.54 kg22.6 kg22.4 kg26.9 kg45.5 kg
Smallest step 2) (Standard mass set)0.05 lb/in²0.05 lb/in²0.5 lb/in²5 lb/in²5 lb/in²5 lb/in²
Smallest step 3) (Set of fine increment masses)----0.05 lb/in²0.1 lb/in²0.2 lb/in²0.2 lb/in²
Measuring range in kPa 1)1.5 …1001.5 …20010 ... 70020 ... 2,500100 ... 7,000100 ... 12,000
Required masses3.3 kg6.54 kg22.5 kg21 kg29 kg49.5 kg
Smallest step 2) (Standard mass set)0.5 kPa0.5 kPa5 kPa30 kPa50 kPa50 kPa
Smallest step 3) (Set of fine increment masses)----0.5 kPa1 kPa2 kPa2 kPa
Nominal effective area of the piston1/2 in²1/2 in²1/2 in²1/8 in²1/16 in²1/16 in²
Accuracies
Standard 4)0.015 %of reading
Premium 5)0.008 %of reading0.006 %of reading0.008 %of reading
Material
PistonHigh-chromium steelTungsten carbideHigh-chromium steel
CylinderHigh-alloy heat-treatable stainless steelTungsten carbideBronze
Mass setStainless steel,non-magnetic
Weight
Piston-cylinder system0.5 kg [1.1 lb]0.5 kg [1.1 lb]1 kg [2.2 lb]1 kg [2.2 lb]2 kg [4.4 lb]2 kg [4.4 lb]
bar mass set incl.mass carrier4 kg [8.8 lb]7.6 kg [16.8 lb]23 kg [50.8 lb]24 kg [53.0 lb]32 kg [70.7 lb]53 kg [117 lb]
kPa mass set incl.mass carrier4 kg [8.8 lb]7.6 kg [16.8 lb]23 kg [50.8 lb]24 kg [53.0 lb]32 kg [70.7 lb]53 kg [117 lb]
lb/in² mass set incl.mass carrier4 kg [8.8 lb]7.6 kg [16.8 lb]23 kg [50.8 lb]24 kg [53.0 lb]30 kg [66.3 lb]49 kg [108 lb]
lb/in² extension mass set, (incl. carrying case)11.0 kg [24.3 lb]18.5 kg [40.9 lb]
Storage case for mass set5.8 kg [12.8 lb]
Dimensions of the storage case (W x H x D)
For extension mass set215 x 310 x 310 mm[8.5 x 12.2 x 12.2 in]
For basic mass set400 x 310 x 310 mm[15.8 x 12.2 x 12.2 in]
For mass set300 x 265 x 205mm 6)400 x 310 x 310 mmand 215 x 310 x 310mm [15.8 x 12.2 x 12.2 in and 8.5 x 12.2 x 12.2 in]
View all (18)
Base
Base version
Pneumatic, low pressureUp to max. 10 bar [150 lb/in²]; with internal pressure generation
Pneumatic, high pressure/vacuumUp to max. 120 bar [1,600 lb/in²]; for external supply and vacuum
Connections
Connection for piston-cylinder system■ G1female thread with adaptation ■ ConTect quick-release connector
Test connection■ Quick connector G½Bfemale thread, freely rotating, changeable → Further threaded inserts, see accessories
External pressure connection■ 6mmSWAGELOK ® threaded pipe connection; max. 110 %of the assigned measuring range ■ With pneumatic high-pressure/vacuum base version and with separator
Pressure transmission medium
BasePneumatic based on clean, dry and non-corrosive gases (e.g. air or nitrogen)
Material
Piping in basePneumatic low pressurePlastic tubing from polyeurethane, 4 x 0.75mm
Pneumatic high pressure/vacuum1.4571 stainless steel,3x1mm
Weight
Pneumatic base■ Low-pressure version ■ High-pressure version18.0 kg [39.8 lb]
■ Low-pressure version, incl. ConTect quick- release connector ■ High-pressure version, incl. ConTect quick- release connector19.0 kg [42.0 lb]
Dimensions (W x D x H)
Base401 x 375 x 265 mm[15.8 x 14.8 x 10.4 in] → For details, see technical drawings
Operating conditions
Operating temperature18 ... 28 °C [64 ... 82 °F]
Storage temperature range-10 ...+50 °C [14 ... 122 °F]
Relative humidity, condensation (place of storage)■ 35 ... 85 %relative humidity for instrument base and mass set (non-condensing) ■ 35 ... 65 %relative humidity for piston-cylinder system (non-condensing)
LogoDescriptionRegion
EACEurasian Economic Community
EMCdirective
Machinery Directive
-MChS Permission for commissioningKazakhstan
Certificate
Calibration
Set of fine increment mass CPM5000■ Without ■ 3.1 inspection certificate per DIN EN 10204 (factory calibration) ■ UKAS calibration certificate (pressure calibration with a piston-cylinder system) ■ UKAS calibration certificate (mass calibration)
Piston-cylinder system CPS5000■ 3.1 inspection certificate per DIN EN 10204 (factory calibration) ■ UKAS calibration certificate (pressure calibration with a mass set) ■ DAkkS calibration certificate (traceable and accredited in accordance with ISO/IEC 17025)
Recommended calibration interval2 to 5 years (dependent on conditions of use)
Patent numberDescription
EP 1475624ConTect quick-release system
Version in barWeight
netgross
0.015 ... 1 bar14 kg[30.9 lb]35 kg[77.2 lb]
0.015 ... 2 bar17.6 kg[38.8 lb]38.6 kg[85.1 lb]
0.1 …7bar34 kg[75.0 lb]55 kg[121.3 lb]
0.2 ... 25 bar32 kg[70.6 lb]53 kg[116.9 lb]
1 ... 70 bar36 kg[79.4 lb]57 kg[125.7 lb]
1 ... 120 bar62 kg[136.7 lb]83 kg[183.0 lb]
Version in lb/in²Weight
netgross
0.2 ... 1514 kg[30.9 lb]35 kg[77.2 lb]
0.2 ... 3017.6 kg[38.8 lb]38.6 kg[85.1 lb]
1 ... 100 lb/in²34 kg[75.0 lb]55 kg[121.3 lb]
3 ... 400 lb/in²34 kg[75.0 lb]55 kg[121.3 lb]
15 ... 1,000 lb/in²36 kg[79.4 lb]57 kg[125.7 lb]
10 ... 1,600 lb/in²58 kg[127.9 lb]79 kg[174.2 lb]
Version in kPaWeight
netgross
1.5 ... 100 kPa14 kg[30.9 lb]35 kg[77.2 lb]
1.5 ... 200 kPa17.6 kg[38.8 lb]38.6 kg[85.1 lb]
10 ... 700 kPa34 kg[75.0 lb]55 kg[121.3 lb]
20 ... 2,500 kPa32 kg[70.6 lb]53 kg[116.9 lb]
100 ... 7,000 kPa36 kg[79.4 lb]57 kg[125.7 lb]
100 ... 12,000 kPa62 kg[136.7 lb]83 kg[183.0 lb]
Pressure balance, hydraulic version, model CPB5800
HydraulicSingle-piston measuring ranges: 1 ... 120 to 2 ... 300 bar [10 ... 1,600 to 30 ... 4,000 lb/in²]
Dual-piston measuring ranges: 1 ... 60 bar / 10 ... 700 bar to 1 ... 60 bar / 20 ... 1,400 bar [10 ... 800 lb/in²/ 100 ... 10,000 lb/in² to 10 ... 800 lb/in² / 200 ... 20,000 lb/in²]
Accuracy■ 0.015 %of reading ■ up to 0.006 %of reading
→ For specifications, see data sheet CT 31.11
Pressure balance, differential pressure version, model CPB5600DP
Measuring ranges (= static pressure + differential pressure):
Pneumatic0.03 ... 2 to 0.4 …100 bar [0.435 ... 30 to 5.8 …1,500 lb/in²]
Hydraulic0.2 ... 60 to 25 …1,600 bar [2.9 ... 1,000 to 350 …23,200 lb/in²]
Accuracy■ 0.015 %of reading ■ 0.008 %of reading
→ For specifications, see data sheet CT 31.56
Adapter set
Adapter set with■ G¼■ ¼NPT
■ G⅜■ ½NPT
■ M20 x 1.5
Adapterset NPT mit■ ⅛NPT■ ⅜NPT
■ ¼NPT■ ½NPT
Set of fine increment masses M1 and F1
Fine increment masses■ 1 x 50 g ■ 2 x 20 g ■ 1 x 10 g ■ 1 x 5 g ■ 2 x 2 g ■ 1 x 1 g ■ 1 x 500 mg ■ 2 x 200 mg■ 1 x 100 mg ■ 1 x 50 mg ■ 2 x 20 mg ■ 1 x 10 mg ■ 1 x 5 mg ■ 2 x 2 mg ■ 1 x 1 mg
Description 1)Order code
CPB-A-AA-
Set of fine increment masses model CPM3500 In bar, accuracy class M1-2-
In lb/in², accuracy class M1-3-
In kPa, accuracy class M1-4-
Adapter set 'BSP' for test connection G½B,male thread on G⅛, G¼, G⅜andG½,female thread-5-
Adapter set 'NPT' for test connection G½B,male thread on ⅛NPT, ¼NPT, ⅜NPTand ½NPT, female thread-6-
Adapter set 'metric' for test connection G½B,male thread on M12 x 1.5 and M20 x 1.5, female thread-7-
Vacuum adapter For measuring ranges 1 bar and 100 kPa only (Model 24)-8-
-Storage case For model CPB3500 instrument base and piston-cylinder system-A-
Two storage cases For mass set-B-
-Test connection G1B, male thread to G½, female thread, rotating-C-
Dual test connection G1B, male thread to G½, female thread, rotating (model 27)-D-
Separator 0 ... 35 bar (Model 35)-E-
Hydraulic fluid based on VG22 mineral oil In plastic bottle, content 0.5 litre-F-
Sealing and maintenance set For instrument base model CPB3500-G-
Description 1)Order code
CPB-A-AA-
Tool set Consisting of: ■ Open-ended spanners ■ BSP adapter ■ Replacement seals ■ Pointer remover ■ Pointer punch-H-
Ordering information for your enquiry
⇓ [ ]