AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
612.20 1
612.20
On order
WIKA · Pressure measurement · Family 612

Capsule pressure gauge, copper alloy

Stainless steel case, NS 63 [2 ½"], 100 [4"], 160 [6"]
Range
-240…240 inH2O
Accuracy
1.6
Protection
IP54
CERoHSUKCA
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

The model 612.20 capsule pressure gauge is based upon the proven capsule measuring system. The capsule element measurement principle is suitable for very low pressures. On pressurisation, the expansion of the capsule element, proportional to the incident pressure, is transmitted to the movement and indicated.

The modular design enables a multitude of combinations of case materials, process connections, nominal sizes and scale ranges. Due to this high variance, the instrument is suitable for use in a wide range of applications within industry.

for further approvals, see page 7

Capsule pressure gauge, model 612.20

The case and the bayonet bezel are made from stainless steel. The material of the process connection is a copper alloy.

For mounting in control panels, the capsule pressure gauges can, depending on the process connection, be fitted with a mounting flange or with a triangular profile ring and mounting bracket.

The scale ranges of 0 ... 6 mbar to 0 ... 600 mbar or 0 ... 2.4 inH2O to 0 ... 240 inH2O and the vacuum and +/- scale ranges ensure the measuring ranges required for a wide variety of applications.

Special features

  • Zero point correction in front
  • Case from stainless steel
  • Robust design and ingress protection IP54
  • Low scale ranges from 0 ... 6 mbar to 0 ... 600 mbar or 0 ... 2.4 inH 2 O to 0 ... 240 inH 2 O
Identification
Model612.20
Product family612
SubcategoryCapsule pressure gauge
Measuring range
Pressure ranges0…6 mbar · 0…600 mbar · 0…2.4 inH2O · 0…240 inH2O · 0…0.006 kg/cm2 · 0…0.6 kg/cm2 · 0…0.6 kPa · 0…60 kPa · 0…600 Pa · 0…60000 Pa · 0…0.1 psi · 0…10 psi · 0…60 mmH2O · 0…6000 mmH2O · 0…1.5 oz/in2 · 0…150 oz/in2 · -600…0 mbar · -240…0 inH2O · -0.6…0 kg/cm2 · -6000…0 mmH2O · -150…0 oz/in2 · -6…0 kPa · -6000…0 Pa · -6…0 psi · -3…3 mbar · -120…120 inH2O · -0.013…0.013 kg/cm2 · -1.25…1.25 kPa · -1250…1250 Pa · -1.8…1.8 psi · -3000…3000 mmH2O · -75…75 oz/in2 · 0…10 mbar · 0…16 mbar · 0…25 mbar · 0…40 mbar · 0…60 mbar · 0…100 mbar · 0…160 mbar · 0…250 mbar · 0…400 mbar · 0…0.01 kg/cm2 · 0…0.016 kg/cm2 · 0…0.025 kg/cm2 · 0…0.04 kg/cm2 · 0…0.06 kg/cm2 · 0…0.1 kg/cm2 · 0…0.16 kg/cm2 · 0…0.25 kg/cm2 · 0…0.4 kg/cm2 · 0…1 kPa · 0…1.6 kPa · 0…2.5 kPa · 0…4 kPa · 0…6 kPa · 0…10 kPa · 0…16 kPa · 0…25 kPa · 0…40 kPa · -60…0 kPa · 0…1000 Pa · 0…1600 Pa · 0…2500 Pa · 0…4000 Pa · 0…6000 Pa · 0…10000 Pa · 0…16000 Pa · 0…25000 Pa · 0…40000 Pa · -60000…0 Pa · 0…0.15 psi · 0…0.25 psi · 0…0.36 psi · 0…0.6 psi · 0…1 psi · 0…1.5 psi · 0…2.5 psi · 0…3.6 psi · 0…6 psi · -10…0 psi · 0…1.5 inH2O · 0…4 inH2O · 0…6 inH2O · 0…10 inH2O · 0…16 inH2O · 0…24 inH2O · 0…40 inH2O · 0…60 inH2O · 0…100 inH2O · 0…160 inH2O · 0…100 mmH2O · 0…160 mmH2O · 0…250 mmH2O · 0…400 mmH2O · 0…600 mmH2O · 0…1000 mmH2O · 0…1600 mmH2O · 0…2500 mmH2O · 0…4000 mmH2O · 0…2.5 oz/in2 · 0…4 oz/in2 · 0…6 oz/in2 · 0…10 oz/in2 · 0…15 oz/in2 · 0…25 oz/in2 · 0…40 oz/in2 · 0…60 oz/in2 · 0…100 oz/in2 · -150…150 oz/in2
Accuracy
Accuracy class1.6
Accuracy standardEN 837-3
Process conditions
Process connection
G ⅛ BG ¼ BG ½ BM20 x 1.5R ¼R ½¼ NPT½ NPTG 1/8 B, male threadG 1/4 B, male threadG 1/2 B, male threadM20 x 1.5, male threadR 1/4, male threadR 1/2, male thread1/4 NPT, male thread1/2 NPT, male thread
Process temperature−20…+100 °C
Materials
Case materialStainless steel
Wetted parts materialCopper alloy, NBR
Sealing materialNBR
Operating environment
Ambient temperature−20…+60 °C
IP ratingIP54
Weight0.2 kg
Certifications and approvals
Certifications
CERoHSUKCA
Additional details
Dial size63 mm
Datasheet languagesEN
Tables from datasheet
Accuracy specifications
Accuracy class
EN 837-3■ Class 1.6 ■ Class 1.0 1)
ASME B40.100■ ±2 %│±1%│±2%ofmeasuring span (grade A) ■ ±1 %of measuring span (grade 1A) 1)
Zero point setting with adjustment screw■ In front, after opening the bayonet bezel 2) ■ In front, through the opening in the window 3)
Temperature errorOn deviation from the reference conditions at the measuring system: ≤ ±0.6 %per 10 °C [≤ ±0.6 %per 18 °F] of full scale value
Reference conditions
Ambient temperature+20 °C [+68 °F]
mbar
0 ... 6 1)0 ... 100
0 ... 10 2)0 ... 160
0 ... 16 2)0 ... 250
0 ... 250 ... 400
0 ... 400 ... 600
0 ... 60
kPa
0 ... 0.6 1)0 ... 10
0 ... 1 2)0 ... 16
0 ... 1.6 2)0 ... 25
0 ... 2.50 ... 40
0 ... 40 ... 60
0 ... 6
psi
0 ... 0.1 1)0 ... 1.5
0 ... 0.15 2)0 ... 2.5
0 ... 0.25 2)0 ... 3.6
0 ... 0.360 ... 6.0
0 ... 0.60 ... 10
0 ... 1.0
View all (20)
kg/cm 2
0 ... 0.006 1)0 ... 0.1
0 ... 0.016 2)0 ... 0.25
0 ... 0.0250 ... 0.4
0 ... 0.040 ... 0.6
0 ... 0.06
Pa
0 ... 600 1)0 ... 10,000
0 ... 1,000 2)0 ... 16,000
0 ... 1,600 2)0 ... 25,000
0 ... 2,5000 ... 40,000
0 ... 4,0000 ... 60,000
0 ... 6,000
mmH 2 O
0 ... 60 1)0 ... 1,000
0 ... 160 2)0 ... 2,500
0 ... 2500 ... 4,000
0 ... 4000 ... 6,000
0 ... 600
inH 2 O
0 ... 2.4 1)0 ... 40
0 ... 4 2)0 ... 60
0 ... 6 2)0 ... 100
0 ...100 ... 160
0 ... 160 ... 240
0 ... 24
mbar
-6…0 1)-3 ...+3 1)
-10…0 2)-5 ...+5 2)
-16…0 2)-8 ...+8 2)
-25…0-12.5 ... +12.5
-40…0-20 ...+20
-60…0-30 ...+30
-100…0-50 ...+50
-160…0-80 ...+80
-250…0-125 ...+125
-400…0-200 ...+200
-600 ... 0-300 ...+300
kPa
-0.6…0 1)-0.3 ...+ 0.3 1)
-1…0 2)-0.5 ... +0.5 2)
-1.6…0 2)-0.8 ... +0.8 2)
-2.5…0-1.25 ... +1.25
-4…0-2 ...+2
-6…0-3 ...+3
-10…0-5 ...+5
-16…0-8 ...+8
-25…0-12.5 ... +12.5
-40…0-20 ...+20
-60 ... 0-30 ...+30
oz/in 2
0 ... 1.5 1)0 ... 25
0 ... 2.5 2)0 ... 40
0 ... 4 2)0 ... 60
0 ... 60 ... 100
0 ... 100 ... 150
0 ... 15
kg/cm 2
-0.006…0 1)-0.003 ... +0.003 1)
-0.01…0 2)-0.005 ... +0.005 2)
-0.016…0 2)-0.008 ... +0.008 2)
-0.025…0-0.0125 ... +0.0125
-0.04…0-0.02 ... +0.02
-0.06…0-0.03 ... +0.03
-0.1…0-0.05 ... +0.05
-0.16…0-0.08 ... +0.08
-0.25…0-0.125 ... +0.125
-0.4…0-0.2 ... +0.2
-0.6 ... 0-0.3 ... +0.3
Pa
-600…0 1)-300 ...+300 1)
-1,000…0 2)-500 ...+500 2)
-1,600…0 2)-800 ...+800 2)
-2,500…0-1,250 ... +1,250
-4,000…0-2,000 ... +2,000
-6,000…0-3,000 ... +3,000
-10,000…0-5,000 ... +5,000
-16,000…0-8,000 ... +8,000
-25,000…0-12,500 ... +12,500
-40,000…0-20,000 ... +20,000
-60,000 ... 0-30,000 ... +30,000
psi
-0.1 ... 0 1)-0.05 ... +0.05 1)
-0.15 ... 0 2)-0.075 ... +0.075 2)
-0.25 ... 0 2)-0.125 ... +0.125 2)
-0.36 ... 0-0.18 ... +0.18
-0.6 ... 0-0.3 ... +0.3
-1 ... 0-0.5 ... +0.5
-1.5 ... 0-0.75 ... +0.75
-2.5 ... 0-1.25 ... +1.25
-3.6 ... 0-1.8 ... +1.8
-6 ... 0-3 ...+3
-10 ... 0-5 ...+5
inH 2 O
-2.4 ... 0 1)-1.2 ... +1.2 1)
-4 ... 0 2)-2 ...+2 2)
-6 ... 0 2)-3 ...+3 2)
-10 ... 0-5 ...+5
-16 ... 0-8 ...+8
-24 ... 0-12 …+12
-40 ... 0-20 ...+20
-60 ... 0-30 ...+30
-100 ... 0-50 ...+50
-160 ... 0-80 ...+80
-240 ... 0-120 ...+120
Further details on: Scale ranges
Unit■ mbar ■ kg/cm² ■ kPa ■ Pa■ psi ■ mmH 2 O ■ inH 2 O ■ oz/in 2
Overpressure safety
Scale range < 0 ... 40 mbar [0 ... 16 inH 2 O]■ Without ■ 3 x full scale value
Scale range ≥ 0 ... 40 mbar [0 ... 16 inH 2 O]■ Without ■ 10 x full scale value
Vacuum safety
Scale range < 0 ... 40 mbar [0 ... 16 inH 2 O]■ Without ■ 3 x full scale value
Scale range ≥ 0 ... 40 mbar [0 ... 16 inH 2 O]■ Without ■ 10 x full scale value
mmH 2 O
-60 ... 0 1)-30 ...+30 1)
-100 ... 0 2)-50 ...+50 2)
-160 ... 0 2)-80 ...+80 2)
-250...0-125 ...+125
-400 ... 0-200 ...+200
-600 ... 0-300 ...+300
-1,000 ... 0-500 ...+500
-1,600 ... 0-800 ...+800
-2,500 ... 0-1,250 ... +1,250
-4,000 ... 0-2,000 ... +2,000
-6,000 ... 0-3,000 ...+3000
oz/in 2
-1.5…0 1)-0.75 ... +0.75 1)
-2.5…0 2)-1.25 ... +1.25 2)
-4…0 2)-2 ...+2 2)
-6…0-3 ...+3
-10…0-5 ...+5
-15…0-7.5 ... +7.5
-25…0-12.5 ... +12.5
-40…0-20 ...+20
-60…0-30 ...+30
-100…0-50 ...+50
-150…0-75 ...+75
Operating conditions
Medium temperature-20 ...+100 °C [-4 ...+212 °F] ≤ 80 °C [≤ 176 °F] with NS 160 [6"]
Ambient temperature-20 ...+60 °C [-4 ...+140 °F]
Pressure limitation
SteadyFull scale value
Fluctuating0.9 x full scale value
Short time1.3 x full scale value
Ingress protection per IEC/EN 60529■ IP54 ■ IP55 ■ IP65 1)
DescriptionRegion
EU declaration of conformity UnionEuropean
Pressure Equipment Directive PS > 200 bar, module A, pressure accessory RoHS directive UKCAUnited Kingdom
Pressure equipment (safety) regulations Restriction of hazardous substances (RoHS) regulations