Технические данные
| General | |
|---|---|
| Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
| Smallest possible measuring range | Depending on application, e.g. CO: 0 to 10 vpm, CO2: 0 to 5 vpm |
| Largest possible measuring range | Depending on application |
| Measuring range with suppressed zero | Any zero point within 0 to 100 % can be achieved; smallest possible span 20 % |
| Heated version | 65 °C |
| Characteristic | Linearized |
| Position of use | Front panel vertical |
| Conformity | CE identification EN 50081-1, EN 50082-2 |
| Design, enclosure | |
| Weight | Approx. 32 kg |
| Degree of protection | IP65 according to EN 60529, restricted breathing to EN 50021 |
| Electrical characteristics | |
| Power supply | 100 to 120 V AC (rated range 90 to 132 V), 48 to 63 Hz or200 to 240 V AC (rated range 180 to 264 V), 48 to 63 Hz |
| Power consumption | Approx. 35 VA; approx. 330 VA with heated version |
| Electromagnetic compatibility (EMC) | According to standard requirements of NAMUR NE21 (08/98) |
| Electrical safety | According to EN 61010-1 |
| - heated units | overvoltage category II |
| - unheated units | overvoltage category III |
| Fuses (unit without heater) | |
| - 100... 120 V | F3: 1T/250; F4: 1T/250 |
| - 200... 240 V | F3: 0.63T/250; F4: 0.63T/250 |
| Fuses (unit with heater) | |
| - 100... 120 V | F1: 1T/250; F2: 4T/250 F3: 4T/250; F4: 4T/250 |
| - 200... 240 V | F1: 0.63T/250; F2: 2.5T/250 F3: 2.5T/250; F4: 2.5T/250 |
| Gas inlet conditions | |
| Perm. sample gas pressure | |
| - for analyzers with hoses (without pressure switch) | 600 to 1500 hPa (absolute) |
| - for analyzers with pipes (without pressure switch) | 600 to 1500 hPa (absolute) |
| - Ex (leakage compensation) | 600 to 1160 hPa (absolute) |
| - Ex (continuous purging) | 600 to 1500 hPa (absolute) |
| Purging gas pressure | |
| - Permanent | < 165 hPa above ambient |
| - For short periods | 250 hPa above ambient |
| Sample gas flow | 18 to 90 l/h (0.3 to 1.5 l/min) |
| Sample gas temperature | 0 to 50 °C; with heated version: 0 to 80 °C |
| Sample gas humidity | < 90% RH (relative humidity) or depending on application |
| Time response | |
| Warm-up period | With amb. temperature < 30 min (maximum accuracy achieved after 2 hours); heated version: approx. 90 min |
| Response time (T90 time) | Dependent on length of analyzer cell, sample gas line and programmed damping |
| Damping (electric time constant) | 0 to 100 s, programmable |
| Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 to 5 s, depending on version |
| Time for internal signal processing | < 1 s |
| Pressure correction range | |
| Pressure sensor | |
| - internal | 600 to 1200 hPa absolute |
| - external | 600 to 1500 hPa absolute |
| Measuring response (maximum accuracy achieved after 2 hours) | |
| Output signal fluctuation | ± 0.1% to ± 1% of smallest possible measuring range specified on rating plate, depending on application with the unit specific electronic time constant (corresponds to ±0.33% at 2σ) |
| Zero drift | < 1% of measuring range/week |
| Measured-value drift | < 1% of measuring range/week |
| Repeatability | Between 0.1% and 1% of respective measuring range |
| Minimum detectable quantity | 1% of smallest measuring range |
| Linearity error | < 0.5% of full-scale value |
| Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
| Ambient temperature | < 1% of measuring range/10 K (for a constant temperature of the reception cell) |
| Sample gas pressure | With pressure compensation: < 0.15% of setpoint/1% change in atmospheric pressure |
| Sample gas flow | Negligible |
| Power supply | < 0.1% of output signal span at rated voltage ± 10% |
| Ambient conditions | Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases |
| Electric inputs and outputs | |
| Analog output | 0/2/4 to 20 mA. floating; max. load 750 Ω |
| Relay outputs | 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC / 1 A floating, non sparking |
| Analog inputs | 2, designed for 0/2/4 to 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross-interference) |
| Binary inputs | 6, designed for 24 V, floating, freely selectable, e.g. for range switching |
| Serial interface | RS 485 |
| Options | Autocal function with 8 additional binary inputs and 8 relay outputs, also with PROFIBUS PA and PROFIBUS DP |
| Ambient conditions | |
| Permissible ambient temperature | -30 to +70 °C during storage and transport, +5 to +45 °C during operation |
| Permissible humidity | < 90 % RH (relative humidity) as annual average, during storage and transport (dew point must not be fallen below) |
Чертеж
ULTRAMAT 6, field unit, dimensions in mm
Схема подключения
Pin assignment (electrical and gas connections)
ULTRAMAT 6, field unit, connector and terminal assignment
ULTRAMAT 6, field unit, connector and terminal assignment of the Autocal board and PROFIBUS connectors
ULTRAMAT 6, field unit, gas and electrical connections
Заказные данные
| Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
|---|---|---|---|---|
| 7MB2111-.....-.... | ULTRAMAT 6F ГАЗОВЫЙ АНАЛИЗАТОР ДЛЯ ПОЛЕВОГО МОНТАЖА | 30 | по запросу (D) | 1 |
| 7MB2112-.....-.... | ULTRAMAT 6F-2R ГАЗОВЫЙ АНАЛИЗАТОР ДЛЯ ПОЛЕВОГО МОНТАЖА | 30 | по запросу (D) | 1 |
Аксессуары
| Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
|---|---|---|---|---|
| 7MB8000-1AA | EX-НАДДУВНОЕ УСТРОЙСТВО MINIPURGE FM | 3.385 | по запросу (D) | 1 |
| 7MB8000-1BA | BARTEC ZONE 2 КОНТРОЛЬНЫЙ МОДУЛЬ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 60 МБАР | 3.385 | по запросу (D) | 1 |
| 7MB8000-2BA | BARTEC EEX P МОДУЛЬ ЭЛЕКТРОНИКИ НАДДУВА ATEX 100 A, 230 V РЕДУКТОР ДАВЛЕНИЯ ИЗ НЕРЖ.СТАЛИ | 1.5 | по запросу (D) | 1 |
| 7MB8000-2BB | BARTEC EEX P МОДУЛЬ ЭЛЕКТРОНИКИ НАДДУВА ATEX 100 A,115 V РЕДУКТОР ДАВЛЕНИЯ ИЗ НЕРЖ.СТАЛИ | 1.5 | по запросу (D) | 1 |
| 7MB8000-2CA | BARTEC ZONE 1 КОНТРОЛЬНЫЙ МОДУЛЬ НИЗКОГО ДАВЛЕНИЯ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 25 МБАР 115VAC | 3.385 | по запросу (D) | 1 |
| 7MB8000-2CB | BARTEC ZONE 1 КОНТРОЛЬНЫЙ МОДУЛЬ НИЗКОГО ДАВЛЕНИЯ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 25 МБАР 115VAC | 3.385 | по запросу (D) | 1 |
| 7MB8000-4AA | РЕЛЕЙНЫЙ ВЫХОД (8), EX | 0.6 | по запросу (D) | 1 |
| 7MB8000-4AB | EXPLOSION PROTECTED ISOLATING RELAY, 110 V | 0.6 | по запросу (D) | 1 |
| 7MB8000-5AA | РЕЛЕ ДИФДАВЛЕНИЯ ДЛЯ АГРЕССИВНЫХ ГАЗОВ | 0.25 | по запросу (D) | 1 |
| 7MB8000-6BA | ОГРАНИЧИТЕЛЬ ПЛАМЕНИ ИЗ НЕРЖ.СТАЛИ В КОМПЛЕКТЕ С МОНТАЖНЫМ УГОЛКОМ | 0.5 | по запросу (D) | 1 |
| 7MB8000-6BB | ОГРАНИЧИТЕЛЬ ПЛАМЕНИ ИЗ ХАСТЕЛЛОЯ В КОМПЛЕКТЕ С МОНТАЖНЫМ УГОЛКОМ | 0.5 | по запросу (D) | 1 |
| A5E00057307 | AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA | 0.5 | по запросу (D) | 1 |
| A5E00057312 | AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP | 0.5 | по запросу (D) | 1 |
| A5E00057315 | ADDITIONAL ELECTRONICS PROFIBUS-PA (FOR FIELD UNITS) | 0.5 | 6 - 8 недель (C) | 1 |
| A5E00057317 | ADDITIONAL ELECTRONICS PROFIBUS PA EX I (FOR FIELD UNITS) | 0.5 | 6 - 8 недель (C) | 1 |
| C79000-G5200-C143 | РУКОВОДСТВО ULTRAMAT 6 / OXYMAT 6 | 0.42 | 4 - 6 недель (B) | 1 |
| C79000-G5272-C143 | РУКОВОДСТВО (ITALIAN) ULTRAMAT 6 / OXYMAT 6 | 0.42 | 4 - 6 недель (B) | 1 |
| C79000-G5276-C143 | РУКОВОДСТВО (ENGLISH) ULTRAMAT 6 / OXYMAT 6 | 0.8 | 4 - 6 недель (B) | 1 |
| C79000-G5277-C143 | РУКОВОДСТВО (FRENCH) ULTRAMAT6 / OXYMAT6 | 0.335 | 4 - 6 недель (B) | 1 |
| C79000-G5278-C143 | РУКОВОДСТВО (SPANISH) ULTRAMAT6 / OXYMAT6 | 0.41 | 4 - 6 недель (B) | 1 |
| C79451-A3364-D61 | ELAN GATEWAY SIPROM GA UP-DATE COMMUNICATION BY TCP/IP AND ETHERNET | 0.5 | 4 - 6 недель (B) | 1 |
| C79451-A3480-D33 | RETROFIT KIT SERIAL INTERFACE FOR AUTOMOTIV (AK)FOR ULTRAMAT 6E 2P AND ULTRAMAT/OXYMAT 6E | 0.25 | по запросу (D) | 1 |
| C79451-Z1589-U1 | INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) | 1 | 6 - 8 недель (C) | 1 |