19"" - корпус

Технические данные

ULTRAMAT 6, 19“ unit
General
Measuring ranges 4, switchable internally and externally; autoranging is also possible
Smallest possible measuring range Depending on application, e.g., CO: 0 … 10 vpm, CO2: 0 … 5 vpm
Largest possible measuring span Depending on application
Measuring range with suppressed zero Every zero possible within 0 to 100 Vol.%, smallest possible measuring span 20%
Characteristic Linearized
Position of use Front panel vertical
Conformity CE identification EN 50081-1, EN 50082-2
Design, enclosure
Weight Approx. 15 kg (with one IR channel), approx. 21 kg (with two IR channels)
Degree of protection IP20 according to EN 60529
Electrical characteristics
Electromagnetic compatibility (EMC) According to standard requirements of NAMUR NE21 (08/98)
Electrical safety According to EN 61010-1, overvoltage category III
Power supply 100 to 120 V AC (rated range 90 to 132 V), 48 to 63 Hz or 200 to 240 V AC (rated range 180 to 264 V), 48 to 63 Hz
Power consumption 1-channel unit: approx. 40 VA 2-channel unit: approx. 70 VA
Fuses
- 100... 120 V 1T/250 (7MB2121), 1.6T/250 (7MB2123)
- 200... 240 V 0.63T/250 (7MB2121), 1T/250 (7MB2123)
Gas inlet conditions
Perm. sample gas pressure
- for analyzers with hoses
- without pressure switch 600 to 1500 hPa (absolute)
- with pressure switch 600 to 1300 hPa (absolute)
- for analyzers with pipes (without pressure switch) 600 to 1500 hPa (absolute)
Sample gas flow 18 to 90 l/h (0.3 to 1.5 l/min)
Sample gas temperature 0 to 50 °C
Sample gas humidity < 90 % RH (relative humidity) or depending on application, non condensing
Time response
Warm-up period With amb. temperature < 30 min (maximum accuracy achieved after 2 hours)
Response time (T90 time) Dependent on length of analyzer cell, sample gas line and damping
Damping (electric time constant) 0 to 100 s, programmable
Dead time (purging time of gas path in analyzer at 1 l/min) Approx. 0.5 to 5 s, depending on version
Time for internal signal processing < 1 s
Pressure correction range
Pressure sensor
- internal 600 to 1200 hPa absolute
- external 600 to 1500 hPa absolute
Measuring response (maximum accuracy achieved after 2 hours)
Output signal fluctuation ± 0.1 % to ± 1 % of smallest possible measuring range specified on rating plate depending on application with the unit specific electronic time constant (corresponds to ± 0.33% at 2σ)
Zero drift < 1% of measuring range/week
Measured-value drift < 1% of measuring range/week
Repeatability ≤ 1% of respective measuring range
Minimum detectable quantity 1% of smallest measuring range
Linearity error < 0.5% of full-scale value
Influencing variables (referred to 1000 hPa sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature < 1% of measuring range/10 K (for a constant temperature of the reception cell)
Sample gas pressure With pressure compensation: < 0.15% of span/1% change in atmospheric pressure without pressure compensation: < 1.5% of span/1% change in atmospheric pressure
Sample gas flow Negligible
Power supply < 0.1% of output signal span with rated voltage ± 10%
Ambient conditions Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases
Electric inputs and outputs
Analog output 0/2/4 to 20 mA. floating; load ≤ 750 Ω
Relay outputs 6, with changeover contacts, freely parameterizable, e.g. for range identification; loading capacity: 24 V AC/DC / 1 A floating, non sparking
Analog inputs 2, designed for 0/2/4 to 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference)
Binary inputs 6, designed for 24 V, floating, freely parameterizable, e.g. for range switching
Serial interface RS 485
Options Autocal function with 8 additional binary inputs and 8 relay outputs, also with PROFIBUS PA and PROFIBUS DP
Ambient conditions
Permissible ambient temperature -30 to +70 °C during storage and transport, +5 to +45 °C during operation
Permissible humidity < 90 % RH (relative humidity) as annual average, during storage and transport (dew point must not be fallen below)

Чертеж

ULTRAMAT 6, 19“ unit, dimensions in mm

Схема подключения

Pin assignment (electrical and gas connections)

ULTRAMAT 6, 19“ unit, pin assignment

ULTRAMAT 6, 19“ unit, pin assignment of Autocal board and PROFIBUS connectors

ULTRAMAT 6, 19“ unit, gas and electrical connections (example: 2-channel version)

Заказные данные

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB2121-.....-.... ULTRAMAT 6E ГАЗОВЫЙ АНАЛИЗАТОР ОДНОКАНАЛЬНЫЙ, 19" ДЛЯ УСТАНОВКИ В ШКАФ 15 по запросу (D) 1
7MB2123-.....-.... ULTRAMAT 6 ГАЗОВЫЙ АНАЛИЗАТОР ДВУХКАНАЛЬНЫЙ, 19" ДЛЯ УСТАНОВКИ В ШКАФ 21 по запросу (D) 1
7MB2124-.....-.... ULTRAMAT 6 ГАЗОВЫЙ АНАЛИЗАТОР, 19" ДЛЯ УСТАНОВКИ В ШКАФ, ДЛЯ ИЗМЕРЕНИЯ 2-3 ГАЗОВ 30 по запросу (D) 1

Аксессуары

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
A5E00057307 AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA 0.5 по запросу (D) 1
A5E00057312 AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP 0.5 по запросу (D) 1
C79000-G5200-C143 РУКОВОДСТВО ULTRAMAT 6 / OXYMAT 6 0.42 4 - 6 недель (B) 1
C79000-G5272-C143 РУКОВОДСТВО (ITALIAN) ULTRAMAT 6 / OXYMAT 6 0.42 4 - 6 недель (B) 1
C79000-G5276-C143 РУКОВОДСТВО (ENGLISH) ULTRAMAT 6 / OXYMAT 6 0.8 4 - 6 недель (B) 1
C79000-G5277-C143 РУКОВОДСТВО (FRENCH) ULTRAMAT6 / OXYMAT6 0.335 4 - 6 недель (B) 1
C79000-G5278-C143 РУКОВОДСТВО (SPANISH) ULTRAMAT6 / OXYMAT6 0.41 4 - 6 недель (B) 1
C79165-A3044-B176 TEMPERATURE SENSOR 0.1 6 - 8 недель (C) 10
C79451-A3364-D61 ELAN GATEWAY SIPROM GA UP-DATE COMMUNICATION BY TCP/IP AND ETHERNET 0.5 4 - 6 недель (B) 1
C79451-A3480-D33 RETROFIT KIT SERIAL INTERFACE FOR AUTOMOTIV (AK)FOR ULTRAMAT 6E 2P AND ULTRAMAT/OXYMAT 6E 0.25 по запросу (D) 1
C79451-A3480-D512 ADDITIONAL ELECTRONICS FOR THE AUTOMOTIVE INDUSTRIE (AK) 0 6 - 8 недель (C) 1
C79451-Z1589-U1 INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) 1 6 - 8 недель (C) 1