Cable temperature probe
Description
The TF-2000 cable temperature probe provides reliable temperature values for refrigeration technology, heating and heat pumps. Frequent changes between the dew point and the freezing point are not a problem - thanks to a measuring element that is completely overmoulded with plastic and is thus dust-tight and waterproof in accordance with IP68. Quick-mounting brackets enable time-saving installation. In addition, the thermometer can be mechanically stabilised using a stainless steel sleeve.
Special features
- Permanently protected against condensation
- Cost savings thanks to quick assembly
- Delivery reliability, even for large orders
Build a TF-2000 for your task
The article number is assembled automatically. If unsure — ask an engineer; we'll match the line's parameters.
- ModelTF-2000
- Approval-CertificateWithout
- Unit of measurementMetric (mm)
- Probe designPlastic moulded, diameter 5 mm [0.2 inch]
- Probe length20,0
- Output signalMeasuring element
- Measuring element—
- Wiring2-wire connection
- Connection lead—
- Lead length [W]—
- Electrical connection—
- Mounting accessoriesWithout
- Certificate EN 10204Without
- MarkingWIKA
- Operating instructionWithout printed operating instruction
- PackagingMulti Packaging
- Additional ordering informationWithout
Finish your configuration
Select every option to submit the request.
| Model | TF-2000 |
| Product family | TF |
| Subcategory | Cable temperature probe |
| Temperature ranges | −50…+105 °C · −40…+90 °C |
| Process temperature | −50…+105 °C |
| Electrical connection | Bare wires or end splices |
| Probe material | TPE, stainless steel, brass |
| Ambient temperature | −40…+90 °C |
| IP rating | IP68 |
| Certifications | RoHSEU declaration of conformityCE |
| Sensor type | Pt1000, Pt100, NTC |
| Sensor class | F 0.3 |
| Probe diameter | 5 mm |
| Thermowell | No |
| Datasheet languages | EN |
| Measuring element | Version |
|---|---|
| Type of measuring element | Pt1000, class F 0.3 per IEC/EN 60751 |
| Pt100, class F 0.3 per IEC/EN 60751 | |
| NTC, R 25 = 10 kΩ, B(25/85) = 3977 | |
| NTC, R 25 = 10 kΩ, B(25/85) = 3435 | |
| Other measuring elements on request | |
| Connection method | 2-wire connection |
| Accuracy specifications | |
|---|---|
| Influence of lead resistance | With the 2-wire connection, the lead resistance of the connection lead affects the measured value and must be taken into consideration. |
| 0.162 Ω/m (guideline value for copper cable with cross-section 0.22mm 2 ) Example Pt100: 0.42 °C/m | |
| Reference conditions | |
| Ambient temperature | 15 ... 25 °C [59 ... 77 °F] |
| Air pressure | 860 ... 1,060 mbar [12.47 ... 15.37 psi] |
| Air humidity | 50 ... 70 %r.h. |
| Dynamic behaviour per IEC/EN 60751 | |
|---|---|
| Response time | The response time is essentially influenced by the probe design, the heat transfer to the measuring element and the flow rate of the medium. |
| Due to the design of the model TF-2000, there is optimum heat transfer from the medium to the measuring element. | |
| t0,5 < 8 s | |
| t0,63 < 11 s | |
| t0,9 < 21 s |
| Probe tip | ■ Plastic moulded (TPE), ∅5mm[0.2 in] ■ Plastic moulded (TPE) with stainless steel probe sleeve, ∅6,1 mm[0.24 in] ■ Plastic moulded (TPE) with brass probe sleeve, square bar 6 x 6 mm[0.24 x 0.24 in] |
|---|
View all (8)
| Connection type | ■ Blank bare wires ■ End splices | |
|---|---|---|
| Insulation material of the connection lead | TPE | ■ Double-insulated round cable ■ Single-insulated ribbon cable |
| The connection lead and the measuring element are moulded together as one single unit.TheTPE variant is halogen-free and has been designed as a double-in- sulated round cable or single-insulated ribbon cable. | ||
| Length of connection lead | 500 ... 10,000 mm[19,7 ... 393.7 in], in 500 mm[19,7 in] increments | |
| → Others on request |
| Medium temperature limit | |
|---|---|
| Single insulated cable | -40 ...+90 °C [-40 ...+194 °F] (120 h at +105 °C [+221 °F], 12 h at +120 °C [+248 °F]) |
| Double insulated cable | -50 ...+105 °C [-58 ...+221 °F] (120 h at +120 °C [+248 °F], 12 h at +150 °C [+302 °F]) |
| Ambient temperature limit 1) | -40 ...+90 °C [-40 ...+194 °F] |
| Storage temperature limit | -25 ...+85 °C [-13 ...+185 °F] |
| Static operating pressure | Atmospheric pressure |
| Ingress protection (IP code) per IEC 60529 | IP68 |
| Mounting position | As required |
| Logo | Description | Region |
|---|---|---|
| EU declaration of conformity | European Union | |
| RoHS directive |
| Model | Order number |
|---|---|
| Quick-mounting bracket, galvanised steel | |
| for pipe diameters 12 …16mm[0.47 ... 0.63 in] | 14145991 |
| for pipe diameters 18 …22mm[0.71 ... 0.87 in] | 14100349 |
| for pipe diameters 25 …28mm[0.98 ... 1.1 in] | 14100347 |
| for pipe diameters 32 …35mm[1.26 ... 1.38 in] | 14149603 |
| for pipe diameters 39 …42mm[1.54 ... 1.65 in] | 14149604 |
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →168 standard articles of family TF
If you know the code — pick directly. If not — use the configurator in mode “01”.