AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
TCA 1
TCA
On order
WIKA · Temperature measurement · Family Process Compact Series

Compact temperature switch, flameproof enclosure Ex d

Compact temperature switch with flameproof enclosure Ex d for safety-critical applications in process industries, featuring vapour pressure measuring system and IP66 protection.
Range
-30…250 °C
Protection
IP66
CERoHSEACINMETROPESOKOSHAEx db I MbEx db IIC T6/T4 GbEx tb IIIC T85/T135 Db IP66Ex db IIC T6/T4Ex tb IIIC T85/T135 GbATEX II 2 GD
Price
On request
Depends on configuration
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

These high-quality temperature switches have been developed especially for safety-critical applications. The high quality of the products and manufacturing in accordance with ISO 9001 ensure reliable monitoring of your plant. In production, the switches are traced by quality assurance software at every step and subsequently are 100 % tested.

In order to ensure as flexible operation as possible, the temperature switches are fitted with micro switches, which enable the switching of an electrical load of up to AC 250 V, 15 A directly. For lower switching power ratings, such as for PLC applications, micro switches with gold-plated contacts can be selected as an option.

Special features

  • Case aluminium, epoxy resin coated
  • Ingress protection IP 65, NEMA 4
  • Ambient temperature -40 ... +85 °C
  • 1 switch point, SPDT or DPDT with a high contact rating of up to 15 A / AC 250 V
  • Capillary up to 10 m
Identification
ModelTCA
Product familyProcess Compact Series
SubcategoryTemperature switch
Measuring range
Temperature ranges−30…+10 °C · −15…+40 °C · +10…+70 °C · +40…+100 °C · +70…+120 °C · +90…+160 °C · +130…+190 °C · +160…+250 °C
Accuracy
Safety integrity level (SIL)SIL 2 per IEC 61508
Process conditions
Process connection
½ NPT male¾ NPT maleG ½ A maleG ¾ A male
Process temperature−40…+280 °C
Electrical characteristics
Power supplyNo power supply needed
Electrical connection½ NPT female (standard), ¾ NPT, M 20 x 1.5, G ½, G ¾ female, cable glands (armoured/non-armoured, Ex d, nickel-plated brass or stainless steel)
Materials
Probe materialstainless steel, Stainless steel 316L
Operating environment
Ambient temperature−40…+85 °C
IP ratingIP66
Weight1 kg
Certifications and approvals
Certifications
CERoHSEACINMETROPESOKOSHA
Explosion protection
Ex db I MbEx db IIC T6/T4 GbEx tb IIIC T85/T135 Db IP66Ex db IIC T6/T4Ex tb IIIC T85/T135 GbATEX II 2 GD
Additional details
Sensor typevapour pressure system
Probe length125 mm
Probe diameter9.5 mm
ThermowellNo
Datasheet languagesEN, DE
Tables from datasheet
Contact versionElectrical rating (resistive load)
ACDC
A1 x SPDT, silver250 V, 15 A24 V, 2 A, 125 V, 0.5 A, 220 V, 0.25 A
B1 x SPDT, silver, hermetically sealed, argon gas filling 2)250 V, 15 A24 V, 2 A, 220 V, 0.5 A
C1 x SPDT, gold-plated, hermetically sealed, argon gas filling 2)125 V, 1 A24 V, 0.5 A
G1 x DPDT, silver250 V, 5 A24 V, 0.5 A
H1 x DPDT, silver, hermetically sealed, air filling250 V, 5 A24 V, 0.5 A
SAMA class 1)Stem∅D inmmActive length X inmmInsertion length Y inmm
IIC9.565125 2)
IIA9.555125 2)
SAMA class 1)BulbInsertion length inmmCapillary length K inm
Stem ∅D inmmActive length X inmmY minY max
IIC9.5711003502
841309005
1241701,80010
IIA 3)9.5551003502
551009005
551001,80010
Setting rangeWorking rangeProof temperatureFixed dead band for contact versionSAMA class
in °Cin °Cin °C1 contact SPDT A, B, C in °C1 contact DPDT G in °C1 contact DPDT H in °C
-30 ...+10-40 ...+6090≤ 2≤ 2≤ 8IIC
-15 ...+40-40 ...+6090≤ 1.5≤ 2≤ 8IIC
10 ... 70-40 ...+7090≤ 1.5≤ 2≤ 8IIC
40 ... 100-40 ...+120140≤ 1.5≤ 2≤ 8IIC
70 ... 120-40 ...+170180≤ 1.5≤ 4≤ 16IIA 1)
90 ... 160-40 ...+170180≤ 2≤ 4≤ 16IIA
130 ... 190-40 ...+190210≤ 2.5≤ 4≤ 16IIA
160 ... 250-40 ...+280300≤ 2.5≤ 4≤ 16IIA
View all (6)
LogoDescriptionCountry
EU declaration of conformity ■ Low voltage directive ■ RoHS directive ■ ATEX 1) directive I M2 (only available with stainless steel 316L switch enclosure) II2GDEuropean Union
IECEx 1) Ex db I Mb (only available with stainless steel 316L switch enclosure) Ex db IIC T6/T4 2) Gb Ex tb IIIC T85/T135 2) Db IP66International
EAC (option) Hazardous areas 1)Eurasian Economic Com- munity
KOSHA (option) Hazardous areas 1)South Korea
INMETRO (option) ■ Metrology, measurement technology ■ Hazardous areas 1)Brazil
-PESO (option) Hazardous areas 1)India
LogoDescription
SIL 2 rating (option) , per IEC 61508 Functional safety The electrical rating for DC applications is limited to 30V / 100 mA.