Measuring insert
Description
The measuring inserts per DIN 43735 for resistance thermometers described here are designed for installation in a protective fitting. Operation without thermowell is only recommended in certain applications. The measuring insert is made of a bendable mineral-insulated sheathed cable. The thermocouple is located at the tip of the measuring insert. The measuring inserts are delivered with pressure springs to ensure that the measuring inserts are pressed down to the
thermowell bottom.
Apart from the DIN versions, customer-specific versions are available, for example:
- ■ other measuring insert lengths (also intermediate lengths) - ■ with mounted sleeve to suit inner diameter of the thermowell - ■ without terminal block - ■ with transmitter
for further approvals see page 2
Measuring insert, model TC10-A
Type and number of sensors, and accuracy can each be selected to suit the respective application.
A large number of different explosion protection approvals are availabe for the TC10-A.
The range of applications is completed by designs without terminal block for direct transmitter installation. Optionally, transmitters from the WIKA range can be installed.
Special features
- Sensor ranges from -40 ... +1,200 °C [-40 ... +2,192 °F]
- Made of mineral-insulated sheathed measuring cable
- Functional safety (SIL) with model T32 temperature transmitter
- Spring-loaded design
- Explosion-protected versions are available for many approval types (see data sheet page 2)
Build a TC10-A for your task
The article number is assembled automatically. If unsure — ask an engineer; we'll match the line's parameters.
- Electrical approvalWithout
- Ignition protection typeWithout
- ZoneWithout
- Thermocouple typeType K
- Number of thermocouplesSingle thermocouple
- Accuracy class of the sensorClass 2, IEC 60584
- Validity limits of class accuracy-40…+1200 °C
- Hot junctionUngrounded
- Transmitter—
- Transmitter details—
- Measuring insert length [l5]—
- Insert diameter—
- Sheath material measuring insertAlloy 600
- CertificatesWithout
Finish your configuration
Select every option to submit the request.
| Model | TC10-A |
| Product family | TC10 |
| Subcategory | Measuring insert |
| Temperature ranges | −40…+1200 °C |
| Safety integrity level (SIL) | SIL 2 |
| Process temperature | −40…+1200 °C |
| Output signal | 4-20mA · HART |
| Probe material | Ni alloy: alloy 600 |
| Ambient temperature | −60…+80 °C |
| IP rating | IP00 |
| Certifications | RoHSNAMUR NE 024GOSTKazInMetrMTSCHSBelGIMUkrSEPROUzstandardCE |
| Explosion protection | II 1G Ex ia IIC T1 ... T6 GaIECExEACCCCPESOII 2G Ex ia IIC T1 ... T6 GbII 2G Ex eb IIC T1 ... T6 GbII 3G Ex ec IIC T1 ... T6 Gc XII 3G Ex nA IIC T1 ... T6 Gc XEx ia IIC T1 ... T6 GaEx ia IIC T1 ... T6 GbEx eb IIC T1 ... T6 GbEx ec IIC T1 ... T6 GcEx nA IIC T1 ... T6 Gc0Ex ia IIC T6…T1 Ga X2Ex nA IIC T6 ... T1 Gc XEx ia IIC T3 ... T6 GaEx ia IIC T1 ~ T6 GaEx e IIC T1 ~ T6 GbEx nA IIC T1 ~ T6 GcEx ia IIC T4 ... T6Ex ib IIC T4 ... T6ATEXEx UkraineINMETROKCs - KOSHAGOSTKazInMetrMTSCHSBelGIMUkrSEPROUzstandard |
| Sensor type | Type K/J/E/N/T |
| Sensor class | Class 1, Class 2 |
| Probe diameter | 6 mm |
| Thermowell | No |
| Datasheet languages | EN |
| Logo | Description | Country | ||
|---|---|---|---|---|
| EU declaration of conformity ■ EMCdirective 1) | European Union | |||
| ■ RoHS directive ■ ATEX directive (option) | International | |||
| IECEx (option) - in conjunction with ATEX Hazardous areas - Ex i Zone 0 gas Ex ia IIC T1 ...T6 Ga Zone 1 gas Ex ia IIC T1 ...T6 Gb - Ex e 4) Zone 1 gas Ex eb IIC T1 ...T6 Gb 3) 5) Zone 2 gas Ex ec IIC T1 ...T6 Gc - Ex n 4) Zone 2 gas Ex nA IIC T1 ...T6 Gc EAC (option) Hazardous areas | ||||
| - Ex i Zone 0 gas 0Ex ia IICT6…T1 Ga X | Eurasian Economic | |||
| - Ex n Zone 2 gas 2Ex nA IIC T6 ...T1 Gc X Ex Ukraine (option) Hazardous areas | Community | |||
| - Ex i Zone 0 gas II 1G Ex ia IIC T1 ...T6 Ga INMETRO (option) | Ukraine | |||
| Hazardous - Ex i | areas Zone 0 gas | Ex ia IIC T3 ...T6 Ga | China | |
| CCC (option) Hazardous areas | ||||
| - Ex i - Ex e 4) | Zone 0 gas | Ex ia IIC T1 ~ T6 Ga | ||
| - Ex n 4) Zone 2 | Zone 1 gas gas | Ex e IIC T1 ~ T6 Gb 3) 5) Ex nA IIC T1 ~ T6 Gc |
| Logo | Description | Country |
|---|---|---|
| KCs - KOSHA (option) Hazardous areas - Ex i Zone 0 gas Ex ia IIC T4 ...T6 Zone 1 gas Ex ib IIC T4 ...T6 | South Korea | |
| - | PESO (option) Hazardous areas - Ex i Zone 0 gas Ex ia IIC T1 ...T6 Ga Zone 1 gas Ex ia IIC T1 ...T6 Gb | India |
| GOST (option) Metrology, measurement technology | Russia | |
| KazInMetr (option) Metrology, measurement technology | Kazakhstan | |
| - | MTSCHS (option) Permission for commissioning | Kazakhstan |
| BelGIM (option) Metrology, measurement technology | Belarus | |
| UkrSEPRO (option) Metrology, measurement technology | Ukraine | |
| Uzstandard (option) Metrology, measurement technology | Uzbekistan |
| Type | Validity limits of class accuracy | ||
|---|---|---|---|
| IEC 60584-1 | ASTM E230 | ||
| Class 2 | Class 1 | Standard Special | |
| K | -40 ... +1,200 °C | -40 ... +1,000 °C | 0 ... 1,260 °C |
| J | -40 ...+750 °C | -40 ...+750 °C | 0 ... 760 °C |
| E | -40 ...+900 °C | -40 ...+800 °C | 0 ... 870 °C |
| N | -40 ... +1,200 °C | -40 ... +1,000 °C | 0 ... 1,260 °C |
| T | -40 ...+350 °C | 0 ... 370 °C |
| Transmitter (selectable versions) | Model T16 | Model T32 |
|---|---|---|
| Data sheet | TE 16.01 | TE 32.04 |
| Output | ||
| 4 ...20mA | x | x |
| HART ® protocol | - | x |
| Input | ||
| Thermocouple IEC 60584-1 | K, J, E, N,T | K, J, E, N,T |
| Galvanic isolation | Yes | Yes |
View all (7)
| Measuring insert length l 5 inmm | Tolerance inmm |
|---|---|
| 75 ... 825 | +2 0 |
| > 825 | +3 0 |
| Measuring insert diameter Ø d inmm | Index per DIN 43735 | Tolerance inmm | |
|---|---|---|---|
| 3 1) | Standard | 30 | 3 ±0.05 |
| 6 | Standard | 60 | 6 0 -0.1 |
| 8 (6 mmwith sleeve) | Standard | - | 8 0 -0.1 |
| 8 | Standard | 80 | 8 0 -0.1 |
| 1/8 in [3.17 mm] 1/4 in [6.35 mm] 3/8 in [9.53 mm] | On request | - | - |
| Certification type | Measurement accuracy | Material certificate |
|---|---|---|
| 2.2 test report | x | x |
| 3.1 inspection certificate | x | x |
| DKD/DAkkS calibration certificate | x | - |
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →88 standard articles of family TC10
If you know the code — pick directly. If not — use the configurator in mode “01”.