AI helper
Describe the task — I'll match 3 WIKA models in 30 seconds
F1201 1
F1201
In stock at WIKA
WIKA · Force measurement · Family F1201

Compression force transducer with strain gauge

Compression force transducer with strain gauge up to 30 t [66,139 lb], hermetically sealed, IP68/IP69K, stainless steel measuring element.
Range
0…30 t
Output
2.0 ±0.1 % mV/V
Accuracy
C3
Protection
IP68
RoHS
Price
From €231
Confirm at order time
Request a quote
+38 (067) 285-77-70·info@avigan.com.uaViberTelegram

Description

Compression force transducers are used for the determination of compression forces in a wide variety of applications and are suitable for static and dynamic measuring requirements.

Force transducers of the F1201 series, due to their compactness, are very frequently used in industrial applications and in the lifting platforms. The spherical cap (load introduction button) allows for a very simple force introduction. The standard mounting position of the force transducer is horizontal or vertical.

The force transducer is hermetically sealed and welded, correspondingly splash-proof and, thanks to an ingress protection of IP68, works reliably, even under difficult operating conditions.

Compression force transducer, model F1201

Special features

  • Measuring ranges 0 ... 5 t [0 ... 11,023 lb] to 0 ... 30 t [0 ... 66,139 lb]
  • Measuring element from stainless steel
  • Hermetically sealed and welded, ingress protection IP68
Identification
ModelF1201
SubcategoryCompression force transducer
Measuring range
Force ranges0…5 t · 0…10 t · 0…30 t · 0…11023 lbf · 0…66139 lbf
Accuracy
Accuracy classC3
Process conditions
Process temperature−10…+40 °C
Mounting typeCompression
Electrical characteristics
Output signal2.0 ±0.1 % mV/V
Materials
Cell materialStainless steel
Operating environment
Ambient temperature−50…+70 °C
IP ratingIP68
Weight1.1 kg
Certifications and approvals
CertificationsRoHS
Additional details
Overload protection150 %
Cable length10 m
Datasheet languagesEN
Tables from datasheet
Model F1201
Rated load F nom t5, 10, 30
Combined error (Non-linearity and hysteresis)< ±0.025%F nom
Relative creep, 30 min.< ±0.024%F nom
Limit force F L150%F nom
Breaking force F B200%F nom
Material of the measuring bodyStainless steel
Temperature error
Zero point< ±0.02%F nom /10 K
Characteristic value< ±0.018%F nom /10 K
Rated temperature range B T, nom-10 …+40 °C
Operating temperature range B T, G-50 …+70 °C
Input resistance R e800 ±30 Ω
Output resistance R a700 ±5 Ω
Insulation resistance R is> 5,000MΩ
Rated displacement (at F nom )< 0.6mm
Output signal (rated characteristic value ) C nom2.0 ±0.1 %mV/V
Electrical connectionCable Ø4 × 10m
Material of the electrical connectionPVC
Supply voltage U BDC 10V ...15V
Ingress protection (per IEC/EN 60529)IP68 and IP69K
Weight
5 t, 10 t1.1 kg
30 t2.8 kg
LogoDescriptionCountry
EU declaration of conformityEuropean Union
RoHS-directive
Rated load in tDimensions inmm
∅ABC∅DER
5, 108244322212130
3012654403514200
Electrical connection
Excitation voltage (Exc+)Green
Excitation voltage (Exc-)Black
Signal (Sig+)Red
Signal (Sig-)White
ShieldShield