Precision thermometer
Description
The precision thermometer model CTR2000 accepts both 3- and 4-wire resistance thermometers. This powerful digital 8-channel precision thermometer is suitable for use with calibrated and uncalibrated Pt100 probes. The model is characterised by its advanced features for high user-friendliness and low life cycle costs.
When using calibrated probes with the CTR2000, the calibration data can be stored either in the instrument or in the probe's protected SMART connector. Calibration data stays permanently with the probe in this connector, which the CTR2000 series instantly recognises, ensuring best measurement accuracy.
A built-in calibration programme can be used to recalibrate the CTR2000 instrument series using a calibrated, traceable reference resistor. This feature is password-protected.
Precision thermometer, model CTR2000
Further functions have been integrated into the CTR2000 series for use with SMART probes, such as the temperature monitor and calibration status monitoring.
The temperature monitor permanently monitors the temperature of the SMART probes and updates the probe log if they are used outside their specified operating range.
This allows you to see at a glance whether the probes have ever been used outside their calibrated range, as this may affect their calibration validity. The calibration status monitor alerts with an on-screen message when a SMART probe is due for recalibration.
Special features
- Temperature range -200 ... +962 °C [-328 ... +1,764 °F]
- Accuracy: 0.01 °C
- 8 channels available
- 3-wire and 4-wire measurement possible
- Revision and processing of SMART probes
| Model | CTR2000 |
| Product family | CTR |
| Subcategory | Multi-channel precision thermometer |
| Temperature ranges | −200…+962 °C |
| Uncertainty | ±0.01 K |
| Calibration type | temperature_bath |
| Ambient temperature | +0…+40 °C |
| Weight | 1 kg |
| Certifications | RoHSEMC Directive EN 61326CE |
| Traceability | NIST · DAkkS |
| Portable | No |
| Datasheet languages | EN |
| Interface | RS-232 |
|---|---|
| Baud rate | 9600 |
| Command sets | SCPI |
| Operating voltage | AC 90 ...264V / 47 ... 63 Hz |
|---|---|
| Power consumption | Max.30VA |
| Power cord | ■ Europe ■ USA/Canada ■ UK ■ China |
| Operating altitude | ≤ 2,000 m[≤ 6,561 ft] above sea level |
|---|---|
| Place of use | Indoor Not for wet locations |
| Operating temperature | 0 ... 40 °C [32 ... 104 °F] |
| Maximum achievable accuracy within 15 ... 25 °C [59 ... 77 °F] | |
| Storage temperature range | -20 ...+50 °C [-4 ...+122 °F] |
| Humidity | 0 ... 80 %relative humidity |
| Condensation | Non-condensing |
| EMC (HF field) | EN 61326-1 emission (group 1, class B) and immunity (industrial application) |
| Logo | Description | Country |
|---|---|---|
| EU declaration of conformity EMCDirective EN 61326 emission (group 1, class B) and immunity (industrial environments) | European Union |
View all (7)
| Key symbol | Description | Function |
|---|---|---|
| Selecting input channels | ||
| Ch | Select input channel 1 to 8 | Selects and displays measuring channel 1 to 8 |
| A-B | Select differential measurement ChA - ChB | Relative measurement function displays the difference between selected ChA and ChB inputs. |
| Setting up measuring options | ||
| Zero | Zero measurement display | Deactivates zero function display |
| Hold | Stop measurement display | Deactivates 'Stop measurement' display (measurement continues) |
| Units | Select units | Cyclical switching of the units (Ω, °C, °F or K) |
| Menu functions | ||
| Menu | Select menu options | Cycles through the main menu functions of the instrument |
| Clear | Clear data entry | Clears any data entry errors or min./max. statistics |
| Enter | Save entry | Saves data entry and returns to previous menu |
| Description | Order code | |
|---|---|---|
| CTX-A-A6 | ||
| - Case | Transport case, robust Suitable for one CTR2000 and two temperature probes model CTPxxxx of your choice | -C5- |
| enquiry: | ||
| your | ||
| Ordering information for | 1. Order code: CTX-A-A6 | ⇓ [ ] |
| 2. Option: |
| Description 1) | Order code |
|---|---|
| Temperature probe model CTP5000 Immersion probe | CTP5000 |
| Temperature probe model CTP5000-R25 Immersion probe | CTP5000-R25 |
| Temperature probe model CTP6000 Immersion probe | CTP6000 |