Reference thermometer
Description
The CTP5000 provides a full range platinum resistance thermometers (PRTs) for use in every application from standards calibration to site temperature measurement. If the 'off the shelf' range will not suit the needs the thermometers can be supplied custom manufactured to almost any specification.
WIKA also provide a range of standards resistors for use when calibrating platinum resistance thermometers.
Using calibrated probes with a precise thermometer there is a choice between storing the calibration data into the memory of the instrument or if using ASL SMART probes for ease of convenience, the calibration is stored in an electronic memory chip located in the connector. Therefore moving the SMART probe between channels or instruments is easy as the calibration data is stored in the SMART-probe connector, no need to enter the calibration data into the instrument channel now being used.
Reference thermometer model CTP5000
For calibration, the test items and the standard thermometer are brought to the same temperature in a temperature conditioning unit.
As soon as a stable temperature is reached, the test items are read or their output signals are measured (resistance, thermoelectric voltage, standard signal) and compared with the standard thermometer.
Using this comparison method, the measuring uncertainty can be considerably reduced because not only the display of the temperature conditioning unit is taken into consideration.
Special features
- Temperature range: -196 ... +660 °C [-321 ... +1,220 °F]
- High stability
- Low drift, long service life
- Bare wires, DIN connector or SMART plug
| Model | CTP5000 |
| Product family | CTP |
| Subcategory | Reference thermometer |
| Temperature ranges | −196…+660 °C |
| Uncertainty | ±1 mK reproducibility, annual drift ±1 mK (model CTP5000-T25) |
| Calibration type | temperature_bath |
| Output signal | resistance |
| Certifications | DKDDAkkSUKAS |
| Traceability | DKD · DAkkS · UKAS |
| Datasheet languages | EN |
| Specifications | Model CTP5000-200 |
|---|---|
| Specific probe data 1) | |
| Temperature range | -50 …+200 °C [-58 ...+392 °F] |
| Resistance at 0 °C [32 °F] | 100Ω |
| Temperature coefficient | 0.00385 |
| R(Ga)/R(TPW) | Ratio less than 1.11807 |
| Annual drift 2) | ±10 ...±20mK |
| Recommended measurement current | 0.5 mAor1mA |
| Self heating error in water at 0 °C [32 °F] | 2 ...5mK |
| Sheath material | Stainless steel |
| Dimensions | |
| Probe diameter | d = 3 mm[0.12 in] |
| Probe length | l = 30 mm[1.18 in], fully immersible |
| Cable | |
| Length | 3 m[9.84 ft] |
| Connection | Bare wire, DIN plug or SMART connector |
| Specifications | Model CTP5000-250 |
|---|---|
| Specific probe data 1) | |
| Temperature range | -50 …+250 °C [-58 ...+482 °F] |
| Resistance at 0 °C [32 °F] | 100Ω |
| Temperature coefficient | 0.00385 |
| R(Ga)/R(TPW) | Ratio less than 1.11807 |
| Annual drift 2) | ±10 ...±20mK |
| Typical stability | ±5mK |
| Recommended measurement current | 0.5 mAor1mA |
| Self heating error in water at 0 °C [32 °F] | typically 2 ... 5 mKat1mA |
| Sheath material | Stainless steel |
| Dimensions | |
| Probe diameter | d = 6 mm[0.24 in] |
| Probe length | l = 450 mm(immersion depth max. 400 mm, min. 100 mm) l = 17.72 in (immersion depth max. 15.75 in, min. 3.94 in) |
| Cable | |
| Length | 2 m[6.56 ft] |
| Connection | Bare wire, DIN plug or SMART connector |
| Specifications | Model CTP5000-450 |
|---|---|
| Specific probe data 1) | |
| Temperature range | -80 …+450 °C [-112 ...+842 °F] |
| Resistance at 0 °C [32 °F] | 100Ω |
| Temperature coefficient | 0.00385 |
| R(Ga)/R(TPW) | Ratio less than 1.11807 |
| Annual drift 2) | ±20 ...±30mK |
| Typical stability | ±5mK |
| Recommended measurement current | 0.5 mAor1mA |
| Self heating error in water at 0 °C [32 °F] | typically 5 ... 10 mKat1mA |
| Sheath material | Stainless steel |
| Dimensions | |
| Probe diameter | d = 6 mm[0.24 in] |
| Probe length | l = 450 mm(immersion depth max. 400 mm, min. 100 mm) l = 17.72 in (immersion depth max. 15.75 in, min. 3.94 in) |
| Cable | |
| Length | 2 m[6.56 ft] |
| Connection | Bare wire, DIN plug or SMART connector |
| Specifications | Model CTP5000-651 |
|---|---|
| Specific probe data 1) | |
| Temperature range | -189 …+650 °C [-308 ... +1,202 °F] |
| Resistance at 0 °C [32 °F] | 100Ω |
| Temperature coefficient | 0.003925 |
| R(Ga)/R(TPW) | Ratio not less than 1.11807 |
| Annual drift 2) | ±10mK |
| Typical stability | ±5mK |
| Recommended measurement current | 0.5 mAor1mA |
| Self heating error in water at 0 °C [32 °F] | typically 13 mKat 1 mAor 3 mKat 0.5mA |
| Sheath material | Fused silica |
| Dimensions | |
| Probe diameter | d = 7.5 mm[0.30 in] |
| Probe length | l = 450 mm(immersion depth max. 350 mm, min. 200 mm) l = 17.72 in (immersion depth max. 13.78 in, min. 7.87 in) |
| Cable | |
| Length | 2 m[6.56 ft] |
| Connection | Bare wire, DIN plug or SMART connector |
View all (18)
| Specifications | Model CTP5000-652 |
|---|---|
| Specific probe data 1) | |
| Temperature range | -80 …+650 °C [-112 ... +1,202 °F] |
| Resistance at 0 °C [32 °F] | 100Ω |
| Temperature coefficient | 0.00385 |
| R(Ga)/R(TPW) | Ratio less than 1.11807 |
| Annual drift 2) | ±15mK |
| Typical stability | ±5mK |
| Recommended measurement current | 1mA |
| Self heating error in water at 0 °C [32 °F] | 10 ... 15 mKat1mA |
| Sheath material | Metal alloy |
| Dimensions | |
| Probe diameter | d = 6 mm[0.24 in] |
| Probe length | l = 450 mm(immersion depth max. 400 mm, min. 300 mm) l = 17.72 in (immersion depth max. 15.75 in, min. 11.81 in) |
| Cable | |
| Length | 2 m[6.56 ft] |
| Connection | Bare wire, DIN plug or SMART connector |
| Specifications | Model CTP5000-T25 |
|---|---|
| Specific probe data 1) | |
| Temperature range | -200 ...+670 °C [-328 ... +1.238 °F] |
| Resistance at 0 °C [32 °F] | 25 ±0.5Ω |
| Calibration | Suitable for calibration per ITS-90 up to a maximum temperature of 660.323 °C [1,220.581 °F] (freezing point aluminium) |
| R(Ga)/R(TPW) | Ratio not less than 1.11807 |
| R(Me)/R(TPW) | Ratio not greater than 0.844235 |
| Reproducibility | ±1mK |
| Annual drift 2) | ±1 mK, depending on use |
| Typical stability | ±1mK |
| Recommended measurement current | 1mA |
| Gas filling | Dry argon / oxygen mixture |
| Sheath material | Fused quartz |
| Dimensions | |
| Sheath | d = 7.5 mm[0.30 in] l = 480 mm(immersion depth max. 400 mm, min. 300 mm) l = 18.90 in (immersion depth max. 15.75 in, min. 11.81 in) |
| Head | l = 90 mm[3.54 in] |
| Cable | |
| Length | 2,5 m[8,20 ft] with low thermal EMF;for high temperatures |
| Connection | gold-plated copper spade lugs |
| Logo | Description | Country |
|---|---|---|
| - | MTSCHS Permission for commissioning | Kazakhstan |
| Calibration | ■ without ■ 3.1 calibration certificate per DIN EN 10204 ■ DKD/DAkkS/UKAS calibration certificate |
|---|---|
| Recommended recalibration interval | 1 year (dependent on conditions of use) |
| Model | Dimensions | Temperature range | Detector length |
|---|---|---|---|
| CTP5000-250 | Pt100, d = 6 mm, l = 450mm Pt100, d = 0.24 in, l = 17.72 in | -50 ...+250 °C [-58 ...+482 °F] | 25mm [0.98 in] |
| CTP5000-450 | Pt100, d = 6 mm, l = 450mm Pt100, d = 0.24 in, l = 17.72 in | -80 ...+450 °C [-112 ...+842 °F] | 25mm [0.98 in] |
| CTP5000-652 | Pt100, d = 6 mm, l = 450 mm(without spring strain relief, 100 mmhandle) Pt100, d = 0.24 in, l = 17.72 in (without spring strain relief, 3.94 in handle) | -80 ...+650 °C [-112 ... +1,202 °F] | 30mm [1.18 in] |
| CTP5000-651 | Pt100, d = 7.5 mm, l = 450 mm(125 mmhandle) Pt100, d = 0.30 in, l = 17.72 in (4.92 in handle) | -189 ...+650 °C [-308 ... +1,202 °F] | 50mm [1.97 in] |
| Model | Dimensions | Temperature range | Detector length |
|---|---|---|---|
| CTP5000-200 | Pt100, d = 3 mm, l=30mm Pt100, d = 0.12 in, l = 1.18 in | -50 ...+200 °C [-58 ...+392 °F] | 6mm [0.24 in] |
| Order code | ||
|---|---|---|
| Description | CTX-A-R2 | |
| SMART DIN plug 5 pin | -G- | |
| DIN plug 5 pin | -C- | |
| Ordering information for your enquiry: | ||
| 1. Order code: CTX-A-R2 | ⇓ | |
| 2. Option: | [ ] |
| Temperature range | -200 …+670 °C [-328 ... +1,238 °F] |
|---|---|
| Max. temperature | |
| Handle | Max. 80 °C [176 °F] |
| Cable | Max. 80 °C [176 °F] |
| Type of measuring element | Pt25 |
| Resistance at 0 °C [32 °F] | 25 ±0.5Ω |
| Connection method | 4-wire connection |
| Material | Platinum resistance |
| Calibration | Suitable for calibration per ITS-90 up to a maximum temperature of 660.323 °C [1,220.581 °F] (freezing point aluminium) |
| Probe | |
| Gas filling | Dry argon / oxygen mixture |
| Sheath material | Fused quartz |
| Case | |
| Dimensions | 680 x 170 x 70 mm[26.77 x 6.69 x 2.76 in] |
| Weight | 2.4 kg [5.29 lb.] (including thermometer) |
| Tolerance class per EN 60751 | Class A |
|---|---|
| Temperature coefficient | α = 0.003926 |
| R(Ga)/R(TPW) | Ratio not less than 1.11807 |
| Reproducibility | ±1mK |
| Basic accuracy | ±1mK |
| Annual drift 2) | ±5mK |
| Typical stability | ±1mK |
| Recommended measurement current | 1mA |
| Self heating error in water at 0 °C [32 °F] | ±2 ...±3 mKwith 1mA in un-stirred water |
| Reference conditions | |
| Ambient temperature | 23 ±2 °C [73 ±2 °F] |
| Humidity | 40 %r.F.±25 %r.h. |
| Connection type | ■ Flying leads ■ 5-pin Din connector ■ SMART plug ■ Banana plugs ■ Gold-plated copper spade lugs |
|---|---|
| Cable | |
| Length | 2.5 m[8.20 ft] with low thermal EMF For high temperatures |
| Material | PTFE sheath |
| Description | |
|---|---|
| Certificates | ■ Without ■ Acceptance test certificate 3.1 with Wvalue verification ■ Acceptance test certificate 3.1 in accordance with DIN EN 10204 (factory calibration) ■ Acceptance test certificate 3.1 in conjunction with a readout instrument (6 test points) ■ Report for ITS-90 suitability |
| Calibration | ■ Without ■ DAkkS/UKAS calibration certificate (traceable and accredited in according with ISO/IEC 17025) ■ DAkkS/UKAS calibration certificate (traceable and accredited in according with ISO/IEC 17025 with a readout instrument (6 test points) |
| Test points for calibration certificate | ■ Without ■ -40 °C, -20 °C, 0 °C, +50 °C, +70 °C and +100 °C ■ Fixed point TPWater, MPGallium ■ Fixed pointTP Argon,TP Mercury, TPWater ■ Fixed pointTP Mercury, TPWater, MPGallium ■ Fixed pointTP Argon,TP Mercury, TPWater, FP Indium ■ Fixed pointTP Argon,TP Mercury, TPWater, FP Indium, FPTin ■ Fixed pointTP Argon,TP Mercury, TPWater, FPTin, FP Zinc ■ Fixed pointTP Argon,TP Mercury, TPWater, FPTin, FP Zinc, FP Aluminum ■ Fixed point low temperature:TP Mercury, TPWater, MPGallium, FP Indium and comparison measurement -196 °C / -80 °C ■ Fixed point medium temperature:TP Mercury, TPWater, MPGallium, FP Indium, FPTin, FP Zinc ■ Fixed point high temperature:TP Water, MPGallium, FP Indium, FPTin, FP Zinc, FP Aluminum ■ Customized calibration points |
| Recommended calibration interval | 1 year (dependent on conditions of use) |
| Fixed point | Temperature | Uncertainty in ±mK | ||||||
|---|---|---|---|---|---|---|---|---|
| °C | [°F] | Range 1 | Range 2 | Range 3 | Range 4 | Range 5 | Range 6 | |
| TP argon | -189.3442 | [-308.8196] | 0.5 | - | 0.5 | 0.5 | 0.5 | 0.5 |
| TP mercury | -38.8344 | [-37.9019] | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 |
| TP water | 0.01 | [32.02] | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 |
| MP gallium | 29.7646 | [85.5763] | - | 0.2 | - | - | - | - |
| FP indium | 156.5985 | [313.8773] | - | - | 1 | 1 | - | - |
| FP tin | 231.928 | [449.470] | - | - | - | 1 | 1 | 1 |
| FP zinc | 419.527 | [787.149] | - | - | - | - | 1.2 | 1.2 |
| FP aluminium | 660.323 | [1,220.581] | - | - | - | - | - | 2 |
| Probe tube length | l = 480 mm[18.90 in] Ø=7.5 mm[0.30 in] |
|---|---|
| Head / Handle | l = 90 mm[3.54 in] Ø=23mm[0.91 in] |
| Immersion depth | Min. 300 mm[11.81 in] Max. 400 mm[15.75 in] |
| Active sensor length | 35 mm[1.37 in] |
| Description | Order code |
|---|---|
| SMART DIN plug 5 pin | -G- |
| DIN plug 5 pin | -C- |
| Ordering information for your enquiry: | ⇓ |
| [ ] |
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →