WIKA · Calibration technology · Family CTD9100
Temperature dry-well calibrator
Portable temperature dry-well calibrator with high accuracy and stability at high temperatures, suitable for on-site use and calibration of temperature measuring instruments.
Range
200…1100 °C
CERoHS
Price
On request
Depends on configuration
Special features
- High accuracy and stability at high temperatures
- Temperature range 200 ... 1,100 °C [392 ... 2,012 °F]
- Possibility to check temperature switches
- Low weight and compact design
- Simple operation
Identification
| Model | CTD9100-1100 |
| Product family | CTD9100 |
| Subcategory | dry-well calibrator |
Measuring range
| Temperature ranges | +200…+1100 °C |
Accuracy
| Uncertainty | ±3 K |
Process conditions
| Calibration type | dry_block |
Operating environment
| Weight | 12 kg |
Certifications and approvals
| Certifications | CERoHS |
| Traceability | DKD · DAkkS |
Additional details
| Portable | Yes |
| Datasheet languages | EN |
Tables from datasheet
| Model CTD9100-1100 | |
|---|---|
| Indication | |
| Display | 2-line LC display 20 ch x lines (3.2 x 5.5 in) with backlighting |
| Temperature range | 200 …1,100 °C [392 ... 2,012 °F] |
| Accuracy 1) | ±3 K |
| Stability 2) | ±0,4 K at 1,000 °C [2,012 °F] |
| Display resolution | 0.01 °C up to 999 °C, then 0.1 °C [0.01 up to 1,830 °F, then 0.1] |
| Temperature distribution | |
| Axial homogeneity 3) | dependent on temperature, temperature probes and their quantity |
| Radial homogeneity 4) | dependent on temperature, temperature probes and their quantity |
| Temperature control | |
| Heating time | approx. 50 min (approx. 18 °C/min) from 20 °C to 900 °C [from 68 °F to 1,652 °F] |
| Cooling time | approx. 150 min from 1,100 °C to 200 °C [from 2,012 °F to 392 °F] |
| Stabilisation time 5) | dependent on temperature and temperature probe |
| Temperature block | |
| Dimensions (D x H) | Ø44 x 300 mm[Ø1.73 x 11.81 in] |
| Block depth | 175 mm[6.89 in] |
| Insert | |
| Immersion depth | 155 mm[6.10 in] |
| Insert dimensions | Ø42.5 x 175 mm[Ø1.67 x 6.89 in] |
| Insulator dimensions | Ø42.5 x 65 mm[Ø1.67 x 2.56 in] |
| Insert material | Ceramic |
| Voltage supply | |
| Power supply | AC 230 V, 50/60 Hz or AC 110 V, 50/60 Hz |
| Power consumption | 950VA |
| Fuse | 6.3 A slow blow fuse |
| Power cord | AC 230 V; for Europe |
| Communication | |
| Interface | RS-232 |
| Case | |
| Dimensions (W x D x H) | 170 x 390 x 330 mm[6.69 x 15.35 x 12.99 in] |
| Weight | 12 kg [26.46 lbs] |
| Logo | Description | Country |
|---|---|---|
| EU declaration of conformity ■ EMCdirective EN 61000-6-2, interference immunity for industrial environments and EN 61000-6-3, emitted interference for residential, commercial and light-industrial environments ■ Low voltage directive EN 61010, EN 61010-1 and EN 61010-2-10, safety requirements for electrical equipment for measurement, control and laboratory use ■ RoHS directive | European Union | |
| GOST (option) Metrology, measurement technology | Russia | |
| KazInMetr (option) Metrology, measurement technology | Kazakhstan | |
| - | MTSCHS (option) Permission for commissioning | Kazakhstan |
| BelGIM (option) Metrology, measurement technology | Belarus |
Certificate
| Calibration | Standard: 3.1 calibration certificate per DIN EN 10204 Option:DKD/DAkkS calibration certificate |
|---|---|
| Recommended recalibration interval | 1 year (dependent on conditions of use) |
| Accessories | Order code | |
|---|---|---|
| Description | CTX-A-K9 | |
| Transport case | -TB- | |
| Power cord for EU | -EU- | |
| for Switzerland | -CH- | |
| for UK | -UK- | |
| for USA/Canada | -US- | |
| Insert replacement tool | -RT- | |
| Ordering information for your enquiry: | ||
| 1. Order code: CTX-A-K9 | ⇓ [ ] |
Application areas not described yet
Describe your operating conditions — our engineer will suggest a model and use case.
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →