Pressure controller
Description
The model CPC3050 high-speed pressure controller is designed with a primary focus on improving throughput in a manufacturing or calibration process. Offering pressure ranges as low as 0 ... 350 mbar [0 ... 5 psi] up to 210 bar [3,045 psi], the CPC3050 can be the solution for multiple applications. This controller can be configured as a desktop or as a 19" rack-mounting version.
Capable of two internal reference sensors and a barometric reference, a single CPC3050 can provide a wide range of pressure control and also emulate both absolute or gauge pressure types. The removable sensors provide more flexibility for customer needs, whether it be spare sensors or various ranges.
Special features
- Pressure ranges: -1 ... 210 bar [-15 ... 3,045 psi]
- Control speed < 4 seconds for 25% FS steps
- Accuracy: to 0.02 % FS
- PACE emulation
Build a CPC3050 for your task
The article number is assembled automatically. If unsure — ask an engineer; we'll match the line's parameters.
- Case typedesk top model
- Pressure range basic instrument—
- Reference pressure sensor 1 (spec. following)Model CPR3050 (0.02 %)
- Reference pressure sensor 2 (spec. following)—
- Barometric reference—
- Type of certificate for barometric reference—
- Pressure port adapters—
- Power cordfor Europe
- Carrying caseWithout
- Further approvalswithout
- Additional ordering informationWithout
Finish your configuration
Select every option to submit the request.
| Model | CPC3050 |
| Product family | CPC |
| Subcategory | pressure controller |
| Pressure ranges | -1…210 bar · 0…0.35 bar · 0…100 bar · 0…210 bar · -0.17…0.17 bar · -1…100 bar · 0…1 bar · 0…101 bar · 0…211 bar |
| Temperature ranges | +10…+50 °C · −20…+70 °C |
| Uncertainty | 0.02 % FS |
| Medium | air |
| Calibration type | electronic |
| Ambient temperature | −20…+70 °C |
| Weight | 12.7 kg |
| Certifications | CERoHS |
| Traceability | A2LA · DAkkS |
| Portable | No |
| Datasheet languages | EN, DE, ES, FR, IT, NL, PL, PT, RO, SV, RU, EL, JA, ZH |
| Pressure range | Low pressure (LP) | High pressure (HP) |
|---|---|---|
| Accuracy 1) | 0.02%FS 2) | |
| Gauge pressure 3) | 0 ... 0.35 up to 0 ... 100 bar [0 ... 5 up to 0 ... 1,500 psi] | 0 ... 100 up to 0 ... 210 bar [0 ... 1,500 up to 0 ... 3,045 psi] |
| Bidirectional 3) 4) | -0.17 ... 0.17 up to -1 ... 100 bar [-2.5 ... 2.5 up to -15 ... 1,500 psi] | -1 ... 100 up to -1 ... 210 bar [-15 ... 1,500 up to -15 ... 3,045 psi] |
| Absolute pressure 5) | 0 ... 1 up to 0 ... 101 bar abs. [0 ... 15 up to 0 ... 1,515 psi abs.] | 0 ... 101 up to 0 ... 211 bar abs. [0 ... 1,515 up to 0 ... 3,065 psi abs.] |
| Precision 6) | 0.008%FS | |
| Calibration interval | 365 days | |
| Optional barometric reference | ||
| Function | The barometric reference can be used to switch pressure types 7) , absolute <=> gauge.With gauge pressure sensors, the measuring range of the sensors must begin with -1 bar [-15 psi] in order to carry out a complete absolute pressure emulation. | |
| Measuring range | 552 ... 1,172 mbar abs. [8 ... 17 psi abs.] | |
| Accuracy 1) | 0.02 %of reading | |
| Pressure units | 39 and two freely programmable |
| Instrument | |
|---|---|
| Instrument version | ■ Desktop case ■ 19" rack-mounting kit |
| Dimensions | See technical drawings |
| Weight | Approx. 12.7 kg [28 lbs] incl. all internal options |
| Display | |
| Screen | 7.0" colour LC display with resistive touchscreen |
| Warm-up time | Approx. 15 min |
| Resolution | 4 ... 6 digits depending on range and units |
| Base instrument | ||
|---|---|---|
| Connections | ||
| Pressure connections | 5 ports with 7/16"- 20 F SAE and 1 port with 10-32 UNF female | |
| Filter elements | The instrument has a 40-micron filters on all pressure ports. | |
| Pressure port adapters | ■ Without ■ 6 mmtube fitting ■ 1/4" tube fitting ■ 1/4" female NPT fittings ■ 1/8" female NPT fittings 1/8" female BSP fittings | |
| Barometer port adapters | ■ Barb fitting ■ 6 mmtube fitting ■ 1/4" tube fitting | |
| Permissible pressure media | ■ Dry, clean air ■ Nitrogen (ISO 8573-1:2010 class 5.5.4 or better) | |
| Wetted parts | ■ Brass ■ Buna N ■ PEEK ■ RTV ■ Silicone ■ Urethane ■ Nickel ■ FKM/FPM ■ PTFE ■ Silicone grease ■ Cerazur ® ceramic ■ Tungsten carbide ■ Stainless 13-8 PH) ■ Stainless ■ Steel alloy | ■ Glass-filled epoxy steel (303 / 304 / 316 / 316L / steel (300 series CRES) 430FR ■ Aluminium (6061-T6 / 2024-T4) |
| Overpressure protection | Safety relief valve for regulator 250 bar [3,600 psi] | |
| Permissible pressure | ||
| Supply port | 110 ... 120 %FSor 0.69 bar [10 psi] → Whichever is greater | |
| Measure/Control port | Max. 105 %FSof the primary sensor range | |
| Plumbing | ≥ 2.8 mm[0.11 in] ID for optimal performance | |
| Voltage supply | ||
| Operating voltage | ■ AC 100 ... 120 V, 50/60 Hz ■ AC 220 ... 240 V, 50/60 Hz | |
| Power consumption | Max.150VA | |
| Permissible ambient conditions | ||
| Compensated temperature range | 10 ... 50 °C [50 ... 122 °F] | |
| Operating temperature range | -20 ...+70 °C [-4 ...+158 °F] | |
| Storage temperature range | -20 ...+70 °C [-4 ...+158 °F] | |
| Humidity | 5 …95%r.h.(non-condensing) | |
| Mounting position | Horizontal | |
| Control parameter | ||
| Control stability | High-speed control mode Industrial control mode < 0.01 | < 0.025 %FSprimary sensor %FSprimary sensor |
| Control mode | ■ Industrial ■ High speed (default) Customised | |
| Control time 1) 2) | 3 ... 4 s (regarding a 25 %FSpressure increase above atmosphere in high-speed control mode) | |
| Control range | 0 ...100%FS | |
| Minimum control pressure 3) | 0.0017 bar [0.025 psi] over exhaust pressure or 0.05%FS Whichever is greater | |
| Overshoots | High-speed control mode | <1%FS |
| Test volume | 0 ... 1,000 ccm |
| Communication | |
|---|---|
| Interface | ■ Ethernet ■ USB ■ IEEE-488 (GPIB) ■ RS-232 |
| Command sets | ■ Mensor ■ PACE ■ WIKA SCPI |
| Others command sets optional | |
| Response time | Approx. 100 ms |
| Internal program | Up to 24 sequences with up to 99 steps each |
View all (10)
| Logo | Description | Region |
|---|---|---|
| EU declaration of conformity | European Union |
| Calibration 1) | |
|---|---|
| CPC3050 | ■ Without ■ A2LA calibration certificate (standard on factory) (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate for barometric reference (traceable and accredited in accordance with ISO/IEC 17025) |
| CPR3050 | ■ A2LA calibration certificate (standard on factory) (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate - gauge pressure (traceable and accredited in accordance with ISO/IEC 17025) ■ DAkkS calibration certificate - absolute pressure (traceable and accredited in accordance with ISO/IEC 17025) |
| Recommended calibration interval | 1 year (dependent on conditions of use) |
| Operating conditions | |
|---|---|
| Maximum operating pressure | 211 bar [3,065 psi] |
| Maximum operating temperature | 80 °C [176 °F] |
| Voltage supply | |
| Power supply | DC 12V |
| Power consumption | 13VA |
| Pressure connection | |
| To the Measure/Control port of CPC3050 | 1 port with 1/4" tube adapted to 7/16"-20 F SAE |
| To the test item | 2 ports |
| ■ 7/16"-20 F SAE ■ 6 mmtube fitting ■ 1/4" tube fitting ■ 1/4" female NPT fittings ■ 1/8" female NPT fittings ■ 1/8" female BSP fittings | |
| Dimensions | |
| Dimensions (W x H x D) | 139.7 x 266.7 x 139.7 mm[5.5 x 10.5 x 5.5 in] |
| Weight | 3.99 kg [8.8 lbs] |
| Cal-Template (demo version) | Cal-Template (light version) | Cal-Template (full version) | Log-Template (full version) |
|---|---|---|---|
| Fully automatic calibration | Semi-automatic calibration | Fully automatic calibration | ■ Live measured value recording for a certain period of time with selectable interval, duration and start time ■ Creation of logger protocols with graphic and/or tabular representation of the measuring results in PDF format ■ Possibility of exporting |
| Limitation to two measuring points | No limitation of the measuring points approached | ||
| ■ Creation of 3.1 inspection certificates per DIN EN 10204 ■ Calibration data can be exported to Excel ® template or XML file ■ Calibration of pressure measuring instruments | measuring results as CSV file | ||
| Ordering information for your enquiry for a single licence: | |||
| Is available for a cost-free download | WIKA-CAL-LZ-Z-Z | WIKA-CAL-CZ-Z-Z | WIKA-CAL-ZZ-L-Z |
| Ordering information for your enquiry for a pair licence: | |||
| Cal-Template (light version) together with Log-Template (full version) | WIKA-CAL-LZ-L-Z | ||
| Cal-Template (full version) together with Log-Template (full version) | WIKA-CAL-CZ-L-Z |
| Accessories for CPC3050 | Order code | |
|---|---|---|
| Description | CPX-A-C4 | |
| - | 19" rack-mounting kit With side panels | -R- |
| Barometric reference Measuring range 8 ... 17 psi abs. Accuracy to 0.02 %of reading | -6- | |
| Measuring range 552 ... 1,172 mbar abs. Accuracy to 0.02 %of reading | -P- | |
| Measuring range 552 ... 1,172 hPa abs. Accuracy to 0.02 %of reading | -Q- | |
| Measuring range 8 ... 17 psi abs. Accuracy to 0.01 %of reading | -3- | |
| Measuring range 552 ... 1,172 mbar abs. Accuracy to 0.01 %of reading | -K- | |
| Measuring range 552 ... 1,172 hPa abs. Accuracy to 0.01 %of reading | -L- | |
| Calibration adapter For reference pressure sensor, voltage supply and software | -4- | |
| For barometric reference, voltage supply and software | -5- | |
| Transport case | -7- | |
| Adapter set 1/4" tube fitting (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -I- | |
| Adapter set 6 mmtube fitting (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -M- | |
| Adapter set 1/8" BSPG female thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -B- | |
| Adapter set 1/4" NPT female thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -N- | |
| Adapter set 1/8" NPT female thread (4 adapters) Max. 137 bar [2,000 psi] Material: brass | -S- | |
| Block-and-bleed valve Pressure range: ≤ 400 bar [≤ 6,000 psi] | -8- | |
| Coalescing filter Pressure range: ≤ 240 bar [≤ 3,600 psi] | -9- |
| Accessories for CPC3050 | Order code | |
|---|---|---|
| Description | CPX-A-C4 | |
| Automatic contamination protection system Pressure range: ≤ 211 bar [3,065 psi] | -A- | |
| Replacement filters for automatic CPS | -2- | |
| Ordering information for your enquiry: | ||
| ⇓ [ ] |
Couldn't find the answer in the documentation? An engineer will reply within a business day.
Ask an engineer →187 standard articles of family CPC
If you know the code — pick directly. If not — use the configurator in mode “01”.