Технические данные

General
Position of use Front panel vertical
Conformity CE identification EN 50081-1 and EN 50082-2
Design, enclosure
Weight Approx. 21 kg
Degree of protection IP20 according to EN 60529
Electrical characteristics
EMC interference immunity(ElectroMagnetic Compatibility) According to standard requirements of NAMUR NE21 (08/98)
Electrical safety According to EN 61010-1, overvoltage category III
Power supply 100 … 120 V AC (rated range 90 V … 132 V), 48 … 63 Hz or 200 … 240 V AC (rated range 180 V … 264 V), 48 … 63 Hz
Power consumption approx. 70 VA
Fuses 120 ... 120 V: F1/F2 = T 1.6 A; 200 ... 240 V: F1/F2 = T 1 A
Electrical inputs and outputs (per channel)
Analog output 0/2/4 … 20 mA, floating, max. load 750 Ώ
Relay outputs 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC/1 A, floating, non sparking
Analog inputs 2, designed for 0/2/4 … 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference)
Binary inputs 6, designed for 24 V, floating, freely selectable, e.g. for range switching
Serial interface RS 485
Options Autocal function with 8 additional binary inputs and 8 relay outputs; also with PROFIBUS PA and PROFIBUS DP
Ambient conditions
Perm. ambient temperature -30 … +70 °C during storage and transport, +5 … +45 ºC during operation
Permissible humidity < 90% relative humidity, during storage and transport (dew point must not be fallen below)
Technical data, ULTRAMAT channel
Measuring ranges 4, switchable internally and externally; autoranging is also possible
Smallest possible measuring range Depending on application, e.g.: CO: 0 … 10 vpm CO2: 0 … 5 vpm
Largest possible measuring range Depending on application
Measuring ranges with suppressed zero Any zero point is possible between 0 and 100%; smallest possible span 20%
Characteristic Linearized
Gas inlet conditions
Permissible sample gas pressure
- Without pressure switch 600 … 1500 hPa (absolute)
- With pressure switch 600 … 1300 hPa (absolute)
Sample gas flow 18 … 90 l/h (0.3 … 1.5 l/min)
Sample gas temperature 0 … 50 ºC
Sample gas humidity < 90 % relative humidity or depending on application, without condensation
Time response
Warm-up period With amb. temperature < 30 min (maximum accuracy achieved after 2 hours)
Response time (T90 time) Dependent on length of analyzer cell, sample gas line and parameterizable damping
Damping (electric time constant) 0 to 100 s, programmable
Dead time (purging time of gas path in analyzer at 1 l/min) Approx. 0.5 to 5 s, depending on version
Time for internal signal processing < 1 s
Pressure correction range
Pressure sensor
- internal 600 … 1200 hPa absolute
- external 600 … 1500 hPa absolute
Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Output signal fluctuation ± 0,1% ... ± 1% of smallest possible measuring range specified on rating plate depending on the unit electronic time constant (corresponds to ± 0.33% with 2σ)
Zero drift < 1% of measuring range/week
Measured-value drift < 1% of measuring range/week
Repeatability ≤ 1% of respective measuring range
Linearity error < 0.5% of full-scale value
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature < 1% of measuring range/10 K (only with stable EK temperature)
Sample gas pressure With pressure compensation: < 0.15% of span/1% change in atmospheric pressure without pressure compensation: < 1.5% of span/1% change in atmospheric pressure
Sample gas flow Negligible
Power supply < 0.1% of output signal span with rated voltage ± 10%
Ambient conditions Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases
Technical data, OXYMAT channel
Measuring ranges 4, switchable internally and externally; autoranging is also possible
Smallest possible measuring span(referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) 0.5% v/v, 2% v/v or 5% v/v O2
Largest possible measuring span 100 % v/v O2
Measuring ranges with suppressed zero Any zero point is possible between 0 and 100% v/v as long as a suitable reference gas is used
Gas inlet conditions
Permissible sample gas pressure
- for analyzers with pipes 500 … 3000 hPa absolute
- for analyzers with hoses
- without pressure switch 500 … 1500 hPa absolute
- with pressure switch 500 … 1300 hPa absolute
Sample gas flow 18 … 60 l/h (0.3 … 1 l/min)
Sample gas temperature 0 … 50 ºC
Sample gas humidity < 90% RH (relative humidity)
Time response
Warm-up period With ambient temperature < 30 min (maximum accuracy achieved after 2 hours)
Reading delay time min. 1.5 … 3.5 s, depending on version
Damping (electric time constant) 0 to 100 s, programmable
Dead time (purging time of gas path in analyzer at 1 l/min) Approx. 0.5 … 2.5 s depending on version
Time for internal signal processing < 1 s
Pressure correction range
Pressure sensor
- internal 500 … 2000 hPa absolute
- external 500 … 3000 hPa absolute
Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Output signal fluctuation < 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (corresponds to ± 0.25% with 2σ)
Zero drift < 0.5%/month of smallest possible meas. span specified on rating plate
Measured-value drift < 0.5%/month of respective measuring span
Repeatability < 1%/month of respective measuring span
Linearity error < 1%/month of respective measuring span
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature < 0.5%/10 K referred to the smallest possible measuring span according to rating plate with span 0.5%: 1%/10 K
Sample gas pressure (with air (100 hPa) as reference gas, a correction of the atmospheric pressure fluctuations is only possible when the sample gas is vented to ambient air) Without pressure compensation: < 2% of measuring span/1% change in pressure; with pressure compensation: < 0.2% of measuring span/1% change in pressure
Residual gases Deviation in zero point corresponding to paramagnetic or diamagnetic deviation of residual gas
Sample gas flow < 1% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range
Power supply < 0.1% of output signal span with rated voltage ± 10%

Чертеж

ULTRAMAT/OXYMAT 6, 19“ unit, dimensions in mm

Схема подключения

Pin assignment (electrical and gas connections)

ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment

ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment of Autocal board and PROFIBUS connectors

ULTRAMAT/OXYMAT 6, 19“ unit, gas and electrical connections

Заказные данные

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB2023-.....-.... АНАЛИЗАТОР ULTRAMAT/OXYMAT 6 ИСПОЛНЕНИЕ 19" ДЛЯ МОНТАЖА В ШКАФ 21 по запросу (D) 1
7MB2024-.....-.... АНАЛИЗАТОР ULTRAMAT 6 ДЛЯ ИЗМЕРЕНИЯ О2 И IR АКТИВНОГО ГАЗА 21 по запросу (D) 1

Аксессуары

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
A5E00057307 AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA 0.5 по запросу (D) 1
A5E00057312 AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP 0.5 по запросу (D) 1
C79000-G5200-C143 РУКОВОДСТВО ULTRAMAT 6 / OXYMAT 6 0.42 4 - 6 недель (B) 1
C79000-G5272-C143 РУКОВОДСТВО (ITALIAN) ULTRAMAT 6 / OXYMAT 6 0.42 4 - 6 недель (B) 1
C79000-G5276-C143 РУКОВОДСТВО (ENGLISH) ULTRAMAT 6 / OXYMAT 6 0.8 4 - 6 недель (B) 1
C79000-G5277-C143 РУКОВОДСТВО (FRENCH) ULTRAMAT6 / OXYMAT6 0.335 4 - 6 недель (B) 1
C79000-G5278-C143 РУКОВОДСТВО (SPANISH) ULTRAMAT6 / OXYMAT6 0.41 4 - 6 недель (B) 1
C79451-A3364-D61 ELAN GATEWAY SIPROM GA UP-DATE COMMUNICATION BY TCP/IP AND ETHERNET 0.5 4 - 6 недель (B) 1
C79451-A3480-D33 RETROFIT KIT SERIAL INTERFACE FOR AUTOMOTIV (AK)FOR ULTRAMAT 6E 2P AND ULTRAMAT/OXYMAT 6E 0.25 по запросу (D) 1
C79451-Z1589-U1 INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) 1 6 - 8 недель (C) 1