Технические данные
| General | |
|---|---|
| Position of use | Front panel vertical |
| Conformity | CE identification EN 50081-1 and EN 50082-2 |
| Design, enclosure | |
| Weight | Approx. 21 kg |
| Degree of protection | IP20 according to EN 60529 |
| Electrical characteristics | |
| EMC interference immunity(ElectroMagnetic Compatibility) | According to standard requirements of NAMUR NE21 (08/98) |
| Electrical safety | According to EN 61010-1, overvoltage category III |
| Power supply | 100 … 120 V AC (rated range 90 V … 132 V), 48 … 63 Hz or 200 … 240 V AC (rated range 180 V … 264 V), 48 … 63 Hz |
| Power consumption | approx. 70 VA |
| Fuses | 120 ... 120 V: F1/F2 = T 1.6 A; 200 ... 240 V: F1/F2 = T 1 A |
| Electrical inputs and outputs (per channel) | |
| Analog output | 0/2/4 … 20 mA, floating, max. load 750 Ώ |
| Relay outputs | 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC/1 A, floating, non sparking |
| Analog inputs | 2, designed for 0/2/4 … 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference) |
| Binary inputs | 6, designed for 24 V, floating, freely selectable, e.g. for range switching |
| Serial interface | RS 485 |
| Options | Autocal function with 8 additional binary inputs and 8 relay outputs; also with PROFIBUS PA and PROFIBUS DP |
| Ambient conditions | |
| Perm. ambient temperature | -30 … +70 °C during storage and transport, +5 … +45 ºC during operation |
| Permissible humidity | < 90% relative humidity, during storage and transport (dew point must not be fallen below) |
| Technical data, ULTRAMAT channel | |
|---|---|
| Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
| Smallest possible measuring range | Depending on application, e.g.: CO: 0 … 10 vpm CO2: 0 … 5 vpm |
| Largest possible measuring range | Depending on application |
| Measuring ranges with suppressed zero | Any zero point is possible between 0 and 100%; smallest possible span 20% |
| Characteristic | Linearized |
| Gas inlet conditions | |
| Permissible sample gas pressure | |
| - Without pressure switch | 600 … 1500 hPa (absolute) |
| - With pressure switch | 600 … 1300 hPa (absolute) |
| Sample gas flow | 18 … 90 l/h (0.3 … 1.5 l/min) |
| Sample gas temperature | 0 … 50 ºC |
| Sample gas humidity | < 90 % relative humidity or depending on application, without condensation |
| Time response | |
| Warm-up period | With amb. temperature < 30 min (maximum accuracy achieved after 2 hours) |
| Response time (T90 time) | Dependent on length of analyzer cell, sample gas line and parameterizable damping |
| Damping (electric time constant) | 0 to 100 s, programmable |
| Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 to 5 s, depending on version |
| Time for internal signal processing | < 1 s |
| Pressure correction range | |
| Pressure sensor | |
| - internal | 600 … 1200 hPa absolute |
| - external | 600 … 1500 hPa absolute |
| Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
| Output signal fluctuation | ± 0,1% ... ± 1% of smallest possible measuring range specified on rating plate depending on the unit electronic time constant (corresponds to ± 0.33% with 2σ) |
| Zero drift | < 1% of measuring range/week |
| Measured-value drift | < 1% of measuring range/week |
| Repeatability | ≤ 1% of respective measuring range |
| Linearity error | < 0.5% of full-scale value |
| Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
| Ambient temperature | < 1% of measuring range/10 K (only with stable EK temperature) |
| Sample gas pressure | With pressure compensation: < 0.15% of span/1% change in atmospheric pressure without pressure compensation: < 1.5% of span/1% change in atmospheric pressure |
| Sample gas flow | Negligible |
| Power supply | < 0.1% of output signal span with rated voltage ± 10% |
| Ambient conditions | Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases |
| Technical data, OXYMAT channel | |
|---|---|
| Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
| Smallest possible measuring span(referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | 0.5% v/v, 2% v/v or 5% v/v O2 |
| Largest possible measuring span | 100 % v/v O2 |
| Measuring ranges with suppressed zero | Any zero point is possible between 0 and 100% v/v as long as a suitable reference gas is used |
| Gas inlet conditions | |
| Permissible sample gas pressure | |
| - for analyzers with pipes | 500 … 3000 hPa absolute |
| - for analyzers with hoses | |
| - without pressure switch | 500 … 1500 hPa absolute |
| - with pressure switch | 500 … 1300 hPa absolute |
| Sample gas flow | 18 … 60 l/h (0.3 … 1 l/min) |
| Sample gas temperature | 0 … 50 ºC |
| Sample gas humidity | < 90% RH (relative humidity) |
| Time response | |
| Warm-up period | With ambient temperature < 30 min (maximum accuracy achieved after 2 hours) |
| Reading delay time | min. 1.5 … 3.5 s, depending on version |
| Damping (electric time constant) | 0 to 100 s, programmable |
| Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 … 2.5 s depending on version |
| Time for internal signal processing | < 1 s |
| Pressure correction range | |
| Pressure sensor | |
| - internal | 500 … 2000 hPa absolute |
| - external | 500 … 3000 hPa absolute |
| Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
| Output signal fluctuation | < 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (corresponds to ± 0.25% with 2σ) |
| Zero drift | < 0.5%/month of smallest possible meas. span specified on rating plate |
| Measured-value drift | < 0.5%/month of respective measuring span |
| Repeatability | < 1%/month of respective measuring span |
| Linearity error | < 1%/month of respective measuring span |
| Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
| Ambient temperature | < 0.5%/10 K referred to the smallest possible measuring span according to rating plate with span 0.5%: 1%/10 K |
| Sample gas pressure (with air (100 hPa) as reference gas, a correction of the atmospheric pressure fluctuations is only possible when the sample gas is vented to ambient air) | Without pressure compensation: < 2% of measuring span/1% change in pressure; with pressure compensation: < 0.2% of measuring span/1% change in pressure |
| Residual gases | Deviation in zero point corresponding to paramagnetic or diamagnetic deviation of residual gas |
| Sample gas flow | < 1% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
| Power supply | < 0.1% of output signal span with rated voltage ± 10% |
Чертеж
ULTRAMAT/OXYMAT 6, 19“ unit, dimensions in mm
Схема подключения
Pin assignment (electrical and gas connections)
ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment
ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment of Autocal board and PROFIBUS connectors
ULTRAMAT/OXYMAT 6, 19“ unit, gas and electrical connections
Заказные данные
| Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
|---|---|---|---|---|
| 7MB2023-.....-.... | АНАЛИЗАТОР ULTRAMAT/OXYMAT 6 ИСПОЛНЕНИЕ 19" ДЛЯ МОНТАЖА В ШКАФ | 21 | по запросу (D) | 1 |
| 7MB2024-.....-.... | АНАЛИЗАТОР ULTRAMAT 6 ДЛЯ ИЗМЕРЕНИЯ О2 И IR АКТИВНОГО ГАЗА | 21 | по запросу (D) | 1 |
Аксессуары
| Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
|---|---|---|---|---|
| A5E00057307 | AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA | 0.5 | по запросу (D) | 1 |
| A5E00057312 | AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP | 0.5 | по запросу (D) | 1 |
| C79000-G5200-C143 | РУКОВОДСТВО ULTRAMAT 6 / OXYMAT 6 | 0.42 | 4 - 6 недель (B) | 1 |
| C79000-G5272-C143 | РУКОВОДСТВО (ITALIAN) ULTRAMAT 6 / OXYMAT 6 | 0.42 | 4 - 6 недель (B) | 1 |
| C79000-G5276-C143 | РУКОВОДСТВО (ENGLISH) ULTRAMAT 6 / OXYMAT 6 | 0.8 | 4 - 6 недель (B) | 1 |
| C79000-G5277-C143 | РУКОВОДСТВО (FRENCH) ULTRAMAT6 / OXYMAT6 | 0.335 | 4 - 6 недель (B) | 1 |
| C79000-G5278-C143 | РУКОВОДСТВО (SPANISH) ULTRAMAT6 / OXYMAT6 | 0.41 | 4 - 6 недель (B) | 1 |
| C79451-A3364-D61 | ELAN GATEWAY SIPROM GA UP-DATE COMMUNICATION BY TCP/IP AND ETHERNET | 0.5 | 4 - 6 недель (B) | 1 |
| C79451-A3480-D33 | RETROFIT KIT SERIAL INTERFACE FOR AUTOMOTIV (AK)FOR ULTRAMAT 6E 2P AND ULTRAMAT/OXYMAT 6E | 0.25 | по запросу (D) | 1 |
| C79451-Z1589-U1 | INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) | 1 | 6 - 8 недель (C) | 1 |