19"" - корпус и переносная версия

19

Технические данные

General
Measured components Max. 4, of which up to 3 infrared-sensitive gases plus oxygen
Measuring ranges 2 per component
Characteristics Linearized
Display LCD with LED backlighting and contrast control, function keys 80 characters (4 lines/20 characters)
Position of use Front panel vertical
Design, enclosure
Weight Approx. 10 kg
Degree of protection, 19" unit and desktop unit IP20 to EN 60529
Electrical characteristics
EMC interference immunity (ElectroMagnetic Compatibility), (safety extra-low voltage (SELV) with safe isolation) According to standard requirements of NAMUR NE21 (08/98) or EN 50081-1, EN 50082-2
Power supply 100 V AC, +10%/-15%, 50 Hz, 120 V AC, +10%/-15%, 50 Hz, 200 V AC, +10%/-15%, 50 Hz, 230 V AC, +10%/-15%, 50 Hz, 100 V AC, +10%/-15%, 60 Hz, 120 V AC, +10%/-15%, 60 Hz, 230 V AC, +10%/-15%, 60 Hz
Power consumption Approx. 60 VA
Electric inputs and outputs
Analog output per component, 0/2/4 to 20 mA, NAMUR, floating, max. load 750 Ω
Relay outputs 8, with changeover contacts, freely selectable, e.g. for range identification, loading capacity, 24 V AC/DC /1 A, floating, non sparking
Binary inputs 3, designed for 24 V, floating
- Pump
- AUTOCAL
- Synchronization
Serial interface RS 485
AUTOCAL function Automatic analyzer calibration with ambient air (depending on measured component), cycle time adjustable from 0 (1) to 24 hours
Options Supplementary electronics with 8 additional binary inputs and relay outputs, e.g. for external automatic calibration and for PROFIBUS PA or PROFIBUS DP
Ambient conditions
Permissible ambient temperature
- Operation +5 … +45 °C
- Storage and transport -20 … +60 °C
Permissible ambient humidity < 90% RH (relative humidity) for storage and transport
Permissible pressure variations 700 … 1200 hPa
Sample gas pressure
Sample gas pressure
- Without pump unpressurized
- With pump unpressurized suction mode, factory-adjusted with 2 m hose at sample gas outlet; an upper range value calibration is required in case of other venting conditions
Sample gas flow 72 … 120 l/h (1.1 … 2 l/min)
Sample gas temperature 0 … 50 °C
Sample gas humidity < 90% RH (relative humidity) no condensation
Technical data, infrared channel
Measuring ranges See Ordering Data
Chopper compartment purge Inlet pressure ca 3000 hPa, purging gas consumption ca 100 ml/min
Time response
Warm-up period Approx. 30 min (at ambient temperature), (maximum accuracy is achieved after approx. 2 hours)
Response time (T90 time) Dependent on the length of the sample gas cell, the sample line and the programmable damping
Damping (electronic time constant) Selectable from 0 to 99.9 s
Measuring response
Noise of output signal < 1 % of smallest measuring range (see rating plate)
Display resolution dependent on selected measuring range; the number of digits after the decimal point can be selected
Minimum detectable quantity 1% of current measuring range
Resolution of output signal < 0.1% of output signal span, max. 4.5 digits
Characteristic Linearized
Linearity error In largest measuring range: < 1% of full-scale value; in smallest measuring range: < 2% of full-scale value
Reproducibility ≤ 1% of smallest measuring range
Influencing variables
Drift
- With AUTOCAL Negligible
- Without AUTOCAL < 2% of smallest measuring range/week
Temperature Max. 2% of smallest possible measuring range according to rating plate per 10 K with an AUTOCAL cycle time of 6 h
Atmospheric pressure < 0.2% of measuring range per 1% change in pressure, corrected by internal pressure sensor
Power supply < 0.1% of output signal span with a variation of ± 10%
Power frequency ± 2% of full-scale value with a frequency variation of ± 5%
Technical data, oxygen channel
Measuring ranges 0 … 5% or 0 … 25% O2, parameter can be set
Typical combustion exhaust gas Influence: < 0.05% O2
Lifetime Approx. 2 years with 21% O2; continuous operation < 0.5% O2 will damage the measuring cell
Time response
Response time (T90 time) Dependent on dead time and selectable damping, not > 30 s with a sample gas flow of approx. 1.2 l/min
Measuring response
Noise of output signal < 0.5% of full-scale value
Display resolution < 0.2% of full-scale value
Resolution of output signal < 0.2% of output signal span
Reproducibility ≤ 0.05% O2
Influencing variables
Oxygen content operation at times < 0.5 % O2 distorts the measured value
Auxiliary gas the oxygen sensor must not be used when the auxiliary gas contains the following components: H2S, chlorine or fluorine compounds, heavy metals, aerosols, mercaptans, alkaline components (as e.g. NH3 in % range)
Humidity H2O dew point ≥ 2 °C; the oxygen sensor must not be used with dry sample gases (no condensation)
Drift
- with AUTOCAL Negligible
- without AUTOCAL 1 % O2/year in air, typical
Temperature < 0.5% O2 per 20 K, referred to a measured value at 20 °C
Atmospheric pressure < 0.2% of measured value per 1% pressure variation

Чертеж

ULTRAMAT 23, 19“ unit, dimensions in mm

ULTRAMAT 23, desktop unit, dimensions in mm

Схема подключения

Pin assignment

ULTRAMAT 23, pin assignment (standard)

ULTRAMAT 23, pin assignment of the optional PROFIBUS interface board

19“ unit

ULTRAMAT 23, 19“ unit, e.g. one infrared component with oxygen measurement

Desktop unit

ULTRAMAT 23, portable unit, in sheet-steel housing, gas and electrical connections

ULTRAMAT 23, designation of the different labels

Заказные данные

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB2335-.....-.... ULTRAMAT23 ГАЗОВЫЙ АНАЛИЗАТОР ДЛЯ ИЗМЕРЕНИЯ 1 ГАЗА И КИСЛОРОДА (ОПЦИЯ) 12 по запросу (D) 1
7MB2337-.....-.... ГАЗОВЫЙ АНАЛИЗАТОР ULTRAMAT 23 ДЛЯ ИЗМЕРЕНИЯ 2 ГАЗОВ И КИСЛОРОДА 12 по запросу (D) 1
7MB2338-.....-.... ГАЗОВЫЙ АНАЛИЗАТОР ULTRAMAT 23 ДЛЯ ИЗМЕРЕНИЯ 3 ГАЗОВ И КИСЛОРОДА 12 по запросу (D) 1

Аксессуары

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
A5E00056834 AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA 0.5 6 - 8 недель (C) 1
A5E00057159 AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP 0.5 по запросу (D) 1
A5E00059149 SAFETY FILTER 0.005 6 - 8 недель (C) 1
C79000-G5200-C216 MANUAL (GERMAN) ULTRAMAT 23 0.1 4 - 6 недель (B) 1
C79000-G5272-C216 MANUAL (ITALIAN) ULTRAMAT 23 0.1 4 - 6 недель (B) 1
C79000-G5276-C216 MANUAL (ENGLISH) ULTRAMAT 23 0.1 4 - 6 недель (B) 1
C79000-G5277-C216 MANUAL, ULTRAMAT 23 (FRENCH) 0.1 4 - 6 недель (B) 1
C79000-G5278-C216 MANUAL (SPANISH) ULTRAMAT 23 0.1 4 - 6 недель (B) 1
C79451-A3364-D61 ELAN GATEWAY SIPROM GA UP-DATE COMMUNICATION BY TCP/IP AND ETHERNET 0.5 4 - 6 недель (B) 1
C79451-Z1589-U1 INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) 1 6 - 8 недель (C) 1