General
Measuring ranges 4, switchable internally and externally; autoranging is also possible
Smallest possible measuring span (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature, smallest possible measuring span with heated version: 0.5% (< 65 °C); 0.5 … 1% (65 … 90 °C); 1 … 2% (90 … 130 °C) 0.5% v/v 2% v/v or 5% v/v O2
Largest possible measuring span 100% v/v O2 (for a pressure over 2000 hPa: 25% v/v O2)
Measuring ranges with suppressed zero Any zero point is possible between 0 and 100% v/v as long as a suitable reference gas is used (see Table 1 in „Function“)
Position of use Front panel vertical
Conformity CE identification EN 50081-1, EN 50082-2
Design, enclosure
Degree of protection IP65 according to EN 60529, gas restricted breathing according to EN 50021
Weight Approx. 28 kg
Electrical characteristics
Power supply 100 to 120 V AC (rated range 90 V to 132 V), 48 to 63 Hz or 200 to 240 V AC (rated range 180 V to 264 V), 48 to 63 Hz
Power consumption Approx. 35 VA, approx. 330 VA with heated version
EMC interference immunity (ElectroMagnetic Compatibility) According to standard requirements of NAMUR NE21 (08/98), EN 61326, EN 50270 (with gas warning unit)
Electrical safety According to EN 61010-1
- Heated units Overvoltage category II
- Unheated units Overvoltage category III
Fuses (units wihout heating)
- 100 ... 120 V F3: 1T/250; F4: 1T/250
- 200 ... 240 V F3: 0.63T/250; F4: 0.63T/250
Fuses (units wih heating)
- 100 ... 120 V F1: 1T/250; F2: 4T/250 F3: 4T/250; F4: 4T/250
- 200 ... 240 V F1: 0.63T/250; F2: 2.5T/250 F3: 2.5T/250; F4: 2.5T/250
Gas inlet conditions
Permissible sample gas pressure
- For analyzers with hoses 500 … 1500 hPa absolute
- For analyzers with piping 500 … 3000 hPa absolute
- For analyzers with piping, Ex version
- Leakage compensation 500 … 1160 hPa absolute
- Continuous purging 500 … 3000 hPa absolute
- Purging gas pressure
- Permanent < 165 hPa above environment
- Short time Max. 250 hPa above environment
Sample gas flow 18 … 60 l/h (0.3 … 1 l/min)
Sample gas temperature 0 … 50 ºC (unheated) 15 °C over temperature analyzer section (heated)
Sample gas humidity < 90 % relative humidity
Time response
Warm-up period With ambient temperature < 30 min (maximum accuracy achieved after 2 hours)
Response time T90 <1.5 s
Damping (electric time constant) 0 … 100 s, programmable
Dead time (purging time of gas path in analyzer at 1 l/min) Approx. 0.5 s
Time for internal signal processing < 1 s
Pressure correction range
Pressure sensor
- Internal 500 … 2000 hPa absolute
- External 500 … 3000 hPa absolute
Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Output signal fluctuation < 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (corresponds to ± 0.25% with 2σ)
Zero drift < 0.5%/month of smallest possible measuring span specified on rating plate
Measured-value drift < 0.5%/month of respective measuring span
Repeatability < 1%/ month of respective measuring span
Minimum detectable quantity 1% of current measuring range
Linearity error < 1%/ month of respective measuring span
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature < 0.5%/10 K referred to the smallest possible measuring span according to rating plate, with measuring span 0.5%: 1%/10 K
Sample gas pressure (with air (100 hPa) as reference gas, a correction of the atmospheric pressure fluctuations is only possible when the sample gas is vented to ambient air) Without pressure compensation: < 2% of measuring span/1% change in pressure with pressure compensation: < 0.2% of measuring span/1% change in pressure
Residual gases Deviation in zero point corresponding to paramagnetic or diamagnetic deviation of residual gas
Sample gas flow < 1% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range; heated version up to double error
Power supply < 0.1% of output signal span with rated voltage ± 10%
Electric inputs and outputs
Analog output 0/2/4 … 20 mA, floating; max. load 750 Ω
Relay outputs 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC/ 1 A, floating
Analog inputs 2, designed for 0/2/4 … 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference)
Binary inputs 6, designed for 24 V, floating, freely-selectable, e.g. for range switching
Serial interface RS 485
Options Autocal function with 8 binary inputs and 8 relay outputs; also with PROFIBUS PA or PROFIBUS DP
Ambient conditions
Perm. ambient temperature -30 … +70 °C during storage and transport, +5 to +45 ºC during operation
Permissible humidity < 90% relative humidity as annual average, during storage and transport (dew point must not be fallen below)

Чертеж

OXYMAT 6, field unit, dimensions in mm

Схема подключения

Pin assignment (electrical and gas connections)

OXYMAT 6, field unit, connector and terminal assignment

OXYMAT 6, field unit, connector and terminal assignment of the Autolcal board and PROFIBUS connectors

OXYMAT 6, field unit, gas and electrical connections

Заказные данные

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB2011-.....-.... OXYMAT 6F АНАЛИЗАТОР КИСЛОРОДА ДЛЯ ПОЛЕВОГО МОНТАЖА 28 по запросу (D) 1

Аксессуары

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB8000-1AA EX-НАДДУВНОЕ УСТРОЙСТВО MINIPURGE FM 3.385 по запросу (D) 1
7MB8000-1BA BARTEC ZONE 2 КОНТРОЛЬНЫЙ МОДУЛЬ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 60 МБАР 3.385 по запросу (D) 1
7MB8000-2BA BARTEC EEX P МОДУЛЬ ЭЛЕКТРОНИКИ НАДДУВА ATEX 100 A, 230 V РЕДУКТОР ДАВЛЕНИЯ ИЗ НЕРЖ.СТАЛИ 1.5 по запросу (D) 1
7MB8000-2BB BARTEC EEX P МОДУЛЬ ЭЛЕКТРОНИКИ НАДДУВА ATEX 100 A,115 V РЕДУКТОР ДАВЛЕНИЯ ИЗ НЕРЖ.СТАЛИ 1.5 по запросу (D) 1
7MB8000-2CA BARTEC ZONE 1 КОНТРОЛЬНЫЙ МОДУЛЬ НИЗКОГО ДАВЛЕНИЯ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 25 МБАР 115VAC 3.385 по запросу (D) 1
7MB8000-2CB BARTEC ZONE 1 КОНТРОЛЬНЫЙ МОДУЛЬ НИЗКОГО ДАВЛЕНИЯ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 25 МБАР 115VAC 3.385 по запросу (D) 1
7MB8000-4AA РЕЛЕЙНЫЙ ВЫХОД (8), EX 0.6 по запросу (D) 1
7MB8000-4AB EXPLOSION PROTECTED ISOLATING RELAY, 110 V 0.6 по запросу (D) 1
7MB8000-5AA РЕЛЕ ДИФДАВЛЕНИЯ ДЛЯ АГРЕССИВНЫХ ГАЗОВ 0.25 по запросу (D) 1
7MB8000-6BA ОГРАНИЧИТЕЛЬ ПЛАМЕНИ ИЗ НЕРЖ.СТАЛИ В КОМПЛЕКТЕ С МОНТАЖНЫМ УГОЛКОМ 0.5 по запросу (D) 1
7MB8000-6BB ОГРАНИЧИТЕЛЬ ПЛАМЕНИ ИЗ ХАСТЕЛЛОЯ В КОМПЛЕКТЕ С МОНТАЖНЫМ УГОЛКОМ 0.5 по запросу (D) 1
A5E00057315 ADDITIONAL ELECTRONICS PROFIBUS-PA (FOR FIELD UNITS) 0.5 6 - 8 недель (C) 1
A5E00057317 ADDITIONAL ELECTRONICS PROFIBUS PA EX I (FOR FIELD UNITS) 0.5 6 - 8 недель (C) 1
C79000-G5200-C143 РУКОВОДСТВО ULTRAMAT 6 / OXYMAT 6 0.42 4 - 6 недель (B) 1
C79000-G5272-C143 РУКОВОДСТВО (ITALIAN) ULTRAMAT 6 / OXYMAT 6 0.42 4 - 6 недель (B) 1
C79000-G5276-C143 РУКОВОДСТВО (ENGLISH) ULTRAMAT 6 / OXYMAT 6 0.8 4 - 6 недель (B) 1
C79000-G5277-C143 РУКОВОДСТВО (FRENCH) ULTRAMAT6 / OXYMAT6 0.335 4 - 6 недель (B) 1
C79000-G5278-C143 РУКОВОДСТВО (SPANISH) ULTRAMAT6 / OXYMAT6 0.41 4 - 6 недель (B) 1
C79451-Z1589-U1 INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) 1 6 - 8 недель (C) 1