Технические данные
| Analytical performance | |
|---|---|
| Measuring range | internally adjustable |
| Minimum detection limits (20 °C, 1000 hPa, 1 m path length) | gas-depending: HF: 0.1 vpm, HCl: 0.2 vpm, NH3: 0.3 vpm, H2O: 1 vpm, O2: 200 vpm, CO:200 vpm, CO2: 200 vpm |
| Smallest recommended measuring range (20 °C, 1000 hPa, 1 m path length) | gas-depending: HF: 0 … 2 vpm, HCl: 0 … 10 vpm, NH3: 0 … 10 vpm, H2O: 0 … 50 vpm |
| in combination with HF, HCl, NH3 | 0 … 5 Vol%, O2: 0 … 1 Vol%, CO: 0 … 1 Vol%, CO2: 0 … 1 Vol% |
| Largest recommended measuring range (20°C, 1000 hPa, 1 m path length or smaller) | gas-depending: HF: 0 … 10 Vol%, HCl: 0 … 25 Vol%, NH3: 0 … 100 Vol%, H2O: 0 … 100 Vol%, %, O2: 0 … 100 Vol%, CO: 0 … 100 Vol%, CO2: 0 … 100 Vol% |
| The max. ranges possible with an analyser might be dependent on the measurement conditions and its individual configuration. In case, a required range is larger than 200 times the smallest range given above, please, contact Siemens to discuss your application. | |
| General | |
| Concentration units | vpm, Vol%, mg/Nm3 (EU/US) |
| Display | Digital concentration display (5 digits with floating decimal point) |
| Mounting orientation | Front plate vertical |
| Laser protection class | Class 1, safe to the eye |
| Laser power | Varies with application |
| Certificates | CE mark, ATEX |
| Linearity | better than 1% |
| Accuracy | better than 2% of reading above the minimum detection limit |
| Design, enclosure | |
| Protection class | IP20 according to EN 60529 |
| Dimensions | 177x440x380 mm |
| Weight | approx 13 kg |
| Electrical characteristics | |
| Power supply | 100 … 240 V AC 50-60 Hz, automatically adjusted to the system |
| Power consumption | 50 W |
| EMC-compatibility | According to EN 61326 and stan-dard classification of NAMUR NE21 |
| Electrical safety | According to EN 61010-1, overvoltage classification II |
| Fuse specifications | 100 ... 240 V: 2.5T/250 |
| Time response | |
| Warm-up temperature at 20 °C ambient | Approx. 15 min |
| Delay of display (T90) | < 1 s |
| Electrical time constant | 0.3 (adjustable), typically 1 … 3 s |
| Dead time | < 1 s |
| Response time | better than 3 s, application dependent |
| Electrical damping | 1 … 100 s, selectable |
| Measurement response | |
| Output signal fluctuation | 2% of measurement value |
| Precision | < 2% … < 5% of measurement value, application dependent |
| Zero point drift | negligible |
| Drift of measurement value | negligible |
| Deviation of linearity | < 1% of measurement value |
| Influencing variables | |
| Ambient temperature | < 1%/10 K of measurement value |
| Ambient pressure | < 1%/50 hPa |
| Measurement gas pressure | < 2% change of reading upon change in pressure of 50 hPa |
| Power supply | < 1% at change of output signal range of ± 10% |
| Tilting | < 1% for non-horizontal mounting of the sensors below 15° |
| Electrical inputs and outputs | |
| Number of measurement channels | 1 … 3, optional |
| Analog output | 2, 4 … 20 mA, potential-free, ohmic resistance max. 750 Ω |
| Analog inputs | 2, designed for 4 … 20 mA |
| Binary outputs | 6, with changeover contacts, configurable, AC/DC 24 V/1 A, potential-free |
| Binary inputs | 6, designed for 24 V, potential-free, configurable |
| Communication Interface | Ethernet 10BaseT (RJ-45) |
| Ambient conditions | |
| Temperature range | +5 … +45 °C during operation, -40 … +70 °C during transportation and storage |
| Humidity | < 85% RH, above dew-point |
| Ambient pressure | 700 … 1200 hPa |
Чертеж
LDS 6, 19“ central unit, dimensions in mm
Схема подключения
Pin assignment
LDS 6, 19“ central unit, pin assignment
Optical and electrical connections
LDS 6, 19“ central unit, optical and electrical connections
Заказные данные
| Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
|---|---|---|---|---|
| 7MB6121-.....-.... | LDS 6 DIODE LASER SPECTROMETER, CENTRAL UNIT, | 14 | по запросу (X) | 1 |
Аксессуары
| Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
|---|---|---|---|---|
| A5E00534675 | CALIBRATION AND VERIFICATION KIT VERSION 1 | 1.5 | по запросу (X) | 1 |
| A5E00823339003 | CALIBRATION KIT RC3009 TUEV/ MCERT, NH3, 2 CELLS, VERSION 1 SERVICE CALIBRATION SET FOR ANALYSERS CERTIFIED ACC. TO 17 TH BIMSCHV. THIS SET IS SUITABLE TO PERFORM REGULAR LINEARITY CHECKS AS REQUIRED FROM THE TEST AUTHORITY | 4.8 | по запросу (X) | 1 |
| A5E00823339004 | CALIBRATION KIT RC3009,TUEV/ MCERT, NH3/H2O, 3 CELLS, VERSION 1, SERVICE CALIBRATION SET FOR ANALYSERS CERTIFIED ACC. TO 17 TH BIMSCHV. THIS SET IS SUITABLE TO PERFORM REGULAR LINEARITY CHECKS AS REQUIRED FROM THE TEST AUTHORITY | 5.8 | по запросу (X) | 1 |
| A5E00823339005 | CALIBRATION KIT RC3009 TUEV/ MCERT, H2O, 2 CELLS VERSION 1 SERVICE CALIBRATION SET FOR ANALYSERS CERTIFIED ACC. TO 17 TH BIMSCHV. THIS SET IS SUITABLE TO PERFORM REGULAR LINEARITY CHECKS AS REQUIRED FROM THE TEST AUTHORITY | 4.8 | по запросу (X) | 1 |
| A5E00823339013 | TUEV/MCERT CALIBRATION KIT NH3, VERSION 2 | 4.8 | по запросу (X) | 1 |
| A5E00823339014 | TUEV/MCERT CALIBRATION KIT NH3/H2O VERSION 2 | 5.8 | по запросу (X) | 1 |
| A5E00823339015 | TUEV/MCERT CALIBRATION KIT H2O VERSION 2 | 4.8 | по запросу (X) | 1 |
| A5E00854188 | LDS6, 3 CHANNEL EXTERNAL POWER SUPPLY | 3 | по запросу (X) | 1 |
| A5E01075594 | CALIBRATION VERIFICATION KIT NH3, VERSION 2 | 3.8 | по запросу (X) | 1 |