SiemensETIJohnson ControlsIFMSICKWIKABosch RexrothELKO EPHYUNDAIABBVEGAKSR KUEBLERAlfa Laval

19"" - корпус

19

Технические данные

General
Measuring ranges 4, switchable between internal and external; manual and automatic measuring range switchover possible
Smallest possible measuring span 0 ... 10 ppm
Verification limit 0.1 ppm
Largest possible measuring span 99.999 ppm
Concentration units ppm, C1, C3, C6 or mgC/m3
Automatic measuring range switchover Hysteresis, selectable
Measured value display Digital concentration display (5 digits with floating point)
Resolution of the digital display 0.1% of the measured value
Position of use Front panel vertical
Conformity CE markEN 50081-1, EN 50082-2
Oven temperature Adjustable 100 ... 200 °C
Design, enclosure
Degree of protection IP20 in accordance with EN 60529
Weight approx. 23 kg
Electrical characteristics
Auxiliary power 100 ... 120 V AC (rated operational range 90 V ... 132 V), 48 ... 63 Hz or 200 ... 240 V AC (rated operational range 180 V ... 264 V), 48 ... 63 Hz
Power consumption approx. 150 VA in operation,approx. 350 VA in the warm-up phase
EMC (Electromagnetic Compatibility) in accordance with standard NAMUR NE21 requirements (08/98)
Electrical safety in accordance with EN 61010-1, overvoltage category II
Fuse values 100 ... 120 V: 4.0T/250200 ... 240 V: 2.5T/250
Gas inlet conditions
Permissible sample gas pressure
- without pump < 2000 hPa abs.
- with built-in pump 600 … 1100 hPa
Sample gas flow 18 ... 60 l/h (0.3 ... 1 l/min)
Sample gas temperature 0 ... 200 °C
Sample gas humidity < 90 % RH (RH: relative humidity.)
Time response
Warm-up period approx. 2-3 h at room temperature
Response time (T90) 2 ... 3 s
Attenuation (electrical time constant) 0 ... 100 s, parameterizable
Dead time (purging time of the gas line in the device at 1 l/min) with filter 2 ... 3 s
Time for signal processing in the device < 1 s
Measuring response
Output signal fluctuation < 0.75% of the smallest possible measuring range according to the rating plate at an electronic attenuation constant of 1 s (this corresponds to ± 0.25% at 2σ)
Zero drift < 0.5%/month of the smallest possible measuring span according to the rating plate
Measured value drift < 1%/week of the appropriate measuring span
Repetition precision < 1% of the appropriate measuring span
Minimum detectable quantity 0.1 ppm (version for ultra-pure gas measurement: 50 ppb)
Linearity error < 1% of the appropriate measuring span
Influencing variables
Ambient temperature < 1%/10 K related to the least possible measuring span according to the rating plate
Ambient pressure < 1%/50 hPa
Sample gas pressure < 2% of the measuring span/1% pressure change (within 600 … 1100 hPa)
Auxiliary power < 1% at a change in the output signal span of ± 10%
Frequency < 1%
Position influence < 1% at an inclination < 15°
Electrical inputs and outputs
Analog output 0/2/4 ... 20 mA, floating; max. load 750 Ω
Relay outputs 6, with freely parameterizable changeover contacts for identifying the measuring range, for example; load rating: 24 V AC/DC /1 A, floating
Analog inputs 2, designed for 0/2/4 to 20 mA for pressure pick-ups externally and for correcting the influence of residual gas (interfering gas)
Binary inputs 6, designed for 24 V, floating, freely parameterizable, for measuring range switchover, for example
Serial interface RS 485
Options Autocal function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP
Climatic conditions
Permissible ambient temperature 5 ... +45 °C in operation, -30 ... +70 °C for storage and transport
Permissible humidity < 90% RH (RH: relative humidity) mid-year, for storage and transport (dew point must not be fallen below)
FIDAMAT 6 with pump and heated oven, with combustion air connection
Gases Operating pressure
Input pressure Pump startup Flow through FID Flow through bypass
without with
hPa (abs.) hPa (abs.) hPa (abs.) ml/min ml/min
Combustion gas 3000 ... 5000 1000 ± 20 ~ 25
Combustion air 3000 ... 5000 1420 ± 20 1500 ~ 320 ~ 500
Sample gas ~ 1000 1500 ± 2 ~ 3 ~ 1000
Zero gas 3500 ... 4000 1500 ± 2 ~ 3 ~ 1000
Calibration gas 3500 ... 4000 1500 ± 2 ~ 3 ~ 1000
FIDAMAT 6 without pump and with heated oven, with combustion air connection
Gases Operating pressure
Input pressure Sample/calibration gas Flow through FID Flow through bypass
without with
hPa (abs.) hPa (abs.) hPa (abs.) ml/min ml/min
Combustion gas 3000 ... 5000 1000 ± 20 ~ 25
Combustion air 3000 ... 5000 1485 ± 2 1500 ~ 320 ~ 300
Sample gas ~ 2000 1500 ± 2 ~ 3 ~ 500
Zero gas ~ 2000 1500 ± 2 ~ 3 ~ 500
Calibration gas ~ 2000 1500 ± 2 ~ 3 ~ 500
FIDAMAT 6 with pump and with heated oven, ambient air = combustion air
Gases Input pressure Operating pressure Flow through FID Flow through bypass
Pump startup
without with
hPa (abs.) hPa (abs.) hPa (abs.) ml/min ml/min
Combustion gas 3000 ... 5000 1000 ± 20 ~ 25 xxxx
Sample gas ~ 1000 Xxxx 1500 ± 2 ~ 3 ~ 500
Zero gas 3500 ... 4000 Xxxx 1500 ± 2 ~ 3 ~ 500
Calibration gas 3500 ... 4000 xxxx 1500 ± 2 ~ 3 ~ 500

Чертеж

FIDAMAT 6, 19“ unit, dimensions in mm

Схема подключения

Pin assignment (electrical and gas connections)

FIDAMAT 6, 19“ unit, pin assignment

FIDAMAT 6, 19“ unit, pin assignment of the Autocal board and PROFIBUS connectors

FIDAMAT 6, gas and electrical connections, version with pump

Заказные данные

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB2421-.....-.... FIDAMAT 6 GAS ANALYZER 19 " INCH UNIT FOR INSTALLATION IN CABINETS 23 по запросу (X) 1

Заказные данные [PDF]

.

Для просмотра информации в формате PDF необходимо установить Adobe Acrobat Reader. Adobe Acrobat Reader

Аксессуары

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
A5E00057307 AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA 0.5 по запросу (D) 1
A5E00057312 AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP 0.5 по запросу (D) 1
C79451-Z1589-U1 INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) 1 6 - 8 недель (C) 1