SiemensETIJohnson ControlsIFMSICKWIKABosch RexrothELKO EPHYUNDAIABBVEGAKSR KUEBLERAlfa Laval

Полевой корпус

Технические данные

CALOMAT 6, field unit
General (following DIN EN 61207 / IEC 1207. All data referred to binary gas mixture H2 in N2)
Measuring ranges 4, switchable internally and externally; autoranging is also possible
Largest possible measuring span 100% H2 (smallest span see „Function“)
Measuring ranges with suppressed zero Any zero point within 0 … 100% can be achieved; smallest possible measuring span 5% H2
Position of use Front panel vertical
Conformity CE identification EN 61326/A1, EN 61010/1
Design, enclosure
Degree of protection IP65 according to EN 60529
Weight Approx. 25 kg
Electrical characteristics
EMC (ElectroMagnetic Compatibility) interference immunity (all signal wires must be shielded; variations of up to 4% of the smallest range may appear in ranges with strong electromagnetic interferences) According to standard requirements of NAMUR NE21 (08/98)
Electrical safety According to EN 61010-1; overvoltage category II
Power supply (see rating plate) 100 -10 % … 120 V AC +10 %, 48 … 63 Hz or 200 -10 % … 240 V AC +10 %, 48 … 63 Hz
Power consumption (unit) Approx. 20 VA
Fuses 100 ... 120 V: 1.0T/250 200 ... 240 V: 0.63T/250
Gas inlet conditions
Sample gas pressure 800 … 1100 hPa (absolute)
Sample gas flow 30 … 90 l/h (0.5 … 1.5 l/min)
Sample gas temperature 0 … 50 ºC
Sample cell temperature Approx. 60 ºC
Sample gas humidity < 90% relative humidity
Purging gas pressure
- Permanent 165 hPa above environment
- for short periods max. 250 hPa above environment
Time response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Warm-up period < 30 min (maximum accuracy achieved after 2 hours)
Response time (T90) < 5 s
Damping (electric time constant) 0 … 100 s, programmable
Dead time (purging time of gas path in analyzer at 1 l/min) Approx. 0.5 s
Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Output signal fluctuation (maximum accuracy achieved after 2 hours) < ± 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (σ = 0.25%)
Zero drift < 1%/week of smallest possible measuring span specified on rating plate
Repeatability < 1% of respective measuring span
Minimum detectable quantity 1% of current measuring range
Linearity error < ± 1% of respective measuring span
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature < 1%/10 K referred to the smallest possible measuring span according to rating plate
Residual gases Deviation in zero point (interfering gas influence see “Influence of interfering gases“)
Sample gas flow < 0.2% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/h within the permissible flow range
Sample gas pressure < 1% for a pressure variation of 100 hPa
Power supply < 0.1% of output signal span with rated voltage ± 10%
Electric inputs and outputs
Analog output 0/2/4 … 20 mA, floating; max. load 750 Ω
Relay outputs 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC/ 1 A, floating
Analog inputs 2, designed for 0/2/4 … 20 mA, for external pressure sensor and correction of influence of residual gas
Binary inputs 6, designed for 24 V, floating, freely-selectable, e.g. for range switching
Serial interface RS 485
Options Autocal function with 8 binary inputs and 8 relay outputs; also with PROFIBUS PA or PROFIBUS DP
Ambient conditions
Perm. ambient temperature -30 to +70 °C during storage and transport, +5 to +45 ºC during operation
Permissible humidity (dew point must not be fallen below) < 90% relative humidity as annual average, during storage and transport

Чертеж

CALOMAT 6, field unit, dimensions in mm

Схема подключения

Pin assignment (electrical and gas connections)

CALOMAT 6, field unit, connector and terminal assignment

CALOMAT 6, field unit, connector and terminal assignment of the Autocal board and PROFIBUS connectors

CALOMAT 6, field unit, gas and electrical connections

Заказные данные

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB2511-.....-.... CALOMAT 6F АНАЛИЗАТОР ПОЛЕВОГО МОНТАЖА 14.5 по запросу (D) 1

Аксессуары

Заказной № Описание Вес (кг) Доставка Мин. зак. (шт)
7MB8000-1AA EX-НАДДУВНОЕ УСТРОЙСТВО MINIPURGE FM 3.385 по запросу (D) 1
7MB8000-1BA BARTEC ZONE 2 КОНТРОЛЬНЫЙ МОДУЛЬ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 60 МБАР 3.385 по запросу (D) 1
7MB8000-2BA BARTEC EEX P МОДУЛЬ ЭЛЕКТРОНИКИ НАДДУВА ATEX 100 A, 230 V РЕДУКТОР ДАВЛЕНИЯ ИЗ НЕРЖ.СТАЛИ 1.5 по запросу (D) 1
7MB8000-2BB BARTEC EEX P МОДУЛЬ ЭЛЕКТРОНИКИ НАДДУВА ATEX 100 A,115 V РЕДУКТОР ДАВЛЕНИЯ ИЗ НЕРЖ.СТАЛИ 1.5 по запросу (D) 1
7MB8000-2CA BARTEC ZONE 1 КОНТРОЛЬНЫЙ МОДУЛЬ НИЗКОГО ДАВЛЕНИЯ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 25 МБАР 115VAC 3.385 по запросу (D) 1
7MB8000-2CB BARTEC ZONE 1 КОНТРОЛЬНЫЙ МОДУЛЬ НИЗКОГО ДАВЛЕНИЯ ДЛЯ РАБОЧЕГО ДАВЛЕНИЯ ДО 25 МБАР 115VAC 3.385 по запросу (D) 1
7MB8000-4AA РЕЛЕЙНЫЙ ВЫХОД (8), EX 0.6 по запросу (D) 1
7MB8000-4AB EXPLOSION PROTECTED ISOLATING RELAY, 110 V 0.6 по запросу (D) 1
7MB8000-5AA РЕЛЕ ДИФДАВЛЕНИЯ ДЛЯ АГРЕССИВНЫХ ГАЗОВ 0.25 по запросу (D) 1
7MB8000-6BA ОГРАНИЧИТЕЛЬ ПЛАМЕНИ ИЗ НЕРЖ.СТАЛИ В КОМПЛЕКТЕ С МОНТАЖНЫМ УГОЛКОМ 0.5 по запросу (D) 1
7MB8000-6BB ОГРАНИЧИТЕЛЬ ПЛАМЕНИ ИЗ ХАСТЕЛЛОЯ В КОМПЛЕКТЕ С МОНТАЖНЫМ УГОЛКОМ 0.5 по запросу (D) 1
A5E00057315 ADDITIONAL ELECTRONICS PROFIBUS-PA (FOR FIELD UNITS) 0.5 6 - 8 недель (C) 1
A5E00057317 ADDITIONAL ELECTRONICS PROFIBUS PA EX I (FOR FIELD UNITS) 0.5 6 - 8 недель (C) 1
C79451-Z1589-U1 INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) 1 6 - 8 недель (C) 1