# Resistance thermometer — TR22-B

> For sanitary applications, for orbital welding

- Catalog code: tr22_b
- Model: TR22-B
- Category: Temperature measurement › Resistance thermometers
- Product page: https://avigan.com.ua/wika/en/products/tr22_b
- Price: on request
- Supplier: AVIGAN (TOV VIKA Prylad), official WIKA partner in Ukraine

## Overview

The model TR22-B resistance thermometer is used for temperature measurement in sanitary applications. To integrate it into the process, the patented ...

## Applications

- Sanitary applications
- Food and beverage industry
- Bio and pharmaceutical industry, production of active ingredients

## Special features

- Simplified calibration through removable measuring inserts
- Stainless steel head in optimised hygienic design, easily cleanable in all mounting positions (patent, property right: GM 000984349)
- Pt100, 4 … 20 mA or HART ® protocol
- Self-draining and dead-space minimised

## Specifications

| Parameter | Value |
|---|---|
| model | TR22-B |
| product family | TR |
| short description | Resistance thermometer for sanitary applications with orbital welding, suitable for food, beverage, and pharmaceutical industries. |
| subcategory | Sanitary resistance thermometer with orbital welding |
| certifications | CE; RoHS; 3-A; EHEDG; EAC |
| explosion protection | Ex ia IIC T1 ... T6 Ga; II 3D Ex tc IIIC TX °C Dc X; Ex ia IIC T1 ... T6 Gb; Ex ia IIIC T125 ... T65 °C Da; Ex ib IIC T3 ... T6 Ga/Gb; II 1G Ex ia IIC T1 ... T6 Ga; II 1/2G Ex ia IIC T1 ... T6 Ga/Gb; II 2G Ex ia IIC T1 ... T6 Gb; II 1D Ex ia IIIC T125 ... T65 °C Da; II 1/2D Ex ia IIIC T125 ... T65 °C Da/Db; II 2D Ex ia IIIC T125 ... T65 °C Db; II 2G Ex eb IIC T1 ... T6 Gb; II 3G Ex ec IIC T1 ... T6 Gc X; II 2D Ex tb IIIC TX °C Db; II 3G Ex nA IIC T1 ... T6 Gc X; Ex ia IIC T1 ... T6 Ga/Gb; Ex ia IIIC T125 ... T65 °C Da/Db; Ex ia IIIC T125 ... T65 °C Db; 0Ex ia IIC T6 ... T1 Ga X; 1Ex ia IIC T6 ... T1 Gb X; Ex ia IIIC T80 ... T440 Da X; Ex ia IIIC T80 ... T440 Db X; 2Ex nA IIC T6 ... T1 Gc X; Ex ia IIIC T65°C Da; Ex ia IIIC T65°C Db; Ex ia IIC T3 ... T6 Ga; Ex ib IIIC T125 ... T65 °C Da/Db; Ex ia IIC T1~T6 Ga; Ex ia IIC T1~T6 Gb; Ex ic IIC T1~T6 Gc; Ex iaD 20 T65/T95/T125; Ex iaD 21 T65/T95/T125; Ex eb IIC T1~T6 Gb; Ex ec IIC T1~T6 Gc; Ex nA IIC T1~T6 Gc; Ex ia IIC T4 ... T6; Ex ib IIC T4 ... T6; Ex ib IIC T3 ... T6 Gb; II 1G Ex ia IIC T1...T6 Ga; II 1/2G Ex ia IIC T1...T6 Ga/Gb; II 2G Ex ia IIC T1...T6 Gb; II 1D Ex ia IIC T125...T65 °C Da; II 1/2D Ex ia IIC T125...T65 °C Da/Db; II 2D Ex ia IIC T125...T65 °C Db; II 2G Ex eb IIC T1...T6 Gb; II 3G Ex ec IIC T1...T6 Gc X; II 2D Ex tb IIC Tx °C Db; II 3D Ex tc IIC Tx °C Dc X; II 3G Ex nA IIC T1...T6 Gc X; 0Ex ia IIC T6...T1 Ga X; 1Ex ia IIC T6...T1 Gb X; Ex ia IIC T80...T440 Da X; Ex ia IIC T80...T440 Db X; 2Ex nA IIC T6...T1 Gc X; Exia IIC T3...T6 Ga; Exib IIC T3...T6 Ga/Gb; Exia IIC T125...T65°C Da; Exib IIC T125...T65°C Da/Db; Exia IIC T1~T6 Ga; Exia IIC T1~T6 Gb; Exic IIC T1~T6 Gc; ExiaD 20 T65/T95/T125; ExiaD 21 T65/T95/T125; Exeb IIC T1~T6 Gb; Exia IIC T4...T6 |
| ambient temp | -40 … 85 °C |
| ip rating | IP65 |
| languages in datasheet | en; de |
| temperature ranges | -50 … 150 °C; 0 … 150 °C; -30 … 250 °C; -50 … 250 °C |
| sensor type | Pt100 |
| sensor class | AA |
| probe diameter mm | 3 |
| probe material | Stainless steel 1.4435; Stainless steel 316L |
| has thermowell | true |
| thermowell material | Stainless steel 1.4435; Stainless steel 316L |
| thermowell form | fabricated |
| process connections | Flow-through housing; Angular housing |
| process temp | -50 … 150 °C |
| output signal | HART; 4-20mA; 4-20 mA; HART® |
| electrical connection | M20 x 1.5, M12 x 1 coupler connector (4-pin) |

## Datasheet tables

| Measuring element | 2 | 3 |
|---|---|---|
| Type of measuring element | ■ Pt100 (thin-film) ■ Face-sensitive Pt100 | (thin-film) 1) |
| Measuring current | IN 00.17 at |  |
| Transmitter version Pt100 version (without transmitter) | Model T15 < Model T32 < 0.1 ...1.0mA | 0.2mA 0.3mA |
| Connection method | ■ 1 x |  |
| Transmitter version | 3-wire ■ 1 x 4-wire |  |
| Pt100 version (without transmitter) | ■ 1 x 3-wire ■ 1 x 4-wire ■ 2 x 3-wire |  |
| Tolerance value of the measuring element 2) per IEC 60751 | ■ Class AA ■ Class A ■ Class B | 0 ... 150 °C -30 ...+250 °C -50 ...+250 °C |

### Measuring range

| Temperature range | -50 ...+150 °C [-58 ...+302 °F] 1) |
|---|---|

### Connection head

| 1 | 2 | 3 | 4 | 5 | 6 |
|---|---|---|---|---|---|
| Model | Material | Cable inlet thread size Ingress (max.) | Cap | Surface |  |
| BS | BS | Aluminium ■ M20 x ■ M12 x connector | 1.5 1 coupler (4-pin) IP65 3) | Flat cover with 2 screws | Blue, painted (RAL 5022) |
| BSZ, BSZ-K | BSZ, BSZ-K BSZ-H, BSZ-HK BSZ Aluminium | BSS ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | BSS-H IP65 3) | BVS (NuG) Spherical hinged cover with cylinder head screw | 1/4000 Blue, painted (RAL 5022) |
| BSZ-H, BSZ-HK BSZ-K | BSS BSS-H PAV antistatic PA12 ■ M20 x ■ M12 x connector | BVS 1.5 1 coupler (4-pin) IP65 | 1/4000 Spherical hinged cover with cylinder head screw | 5/6000 Black |  |
|  | BSZ-H Aluminium | ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | IP65 3) | Raised hinged cover with cylinder head screw | Blue, painted (RAL 5022) |
| BSZ-H, BSZ-HK BSS BSZ-HK | BSS-H PAV antistatic PA12 | BVS BVS (NuG) ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | IP65 | 7/8000 5/6000 Raised hinged cover with cylinder head screw | 7/8000 DIH50 Black |
| BSZ-H / DIH10 2) | Aluminium | ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | IP65 | Raised hinged cover with cylinder head screw | Blue, painted (RAL 5022) |

### Connection head

| 1 | 2 | 3 | 4 | 5 | 6 |
|---|---|---|---|---|---|
| Model | Material | Cable inlet thread size Ingress (max.) | protection Cap | Surface |  |
| BVS | Stainless steel (1.4308) | ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | IP65 | Screw-on lid, hygienic design | Natural finish, electropolished |
| BVS BVS (NuG) BVC | 1/4000 Stainless steel (1.4571) | 7/8000 5/6000 ■ M16 x 1.5 ■ M12 x 1 coupler connector (4-pin) | 7/8000 IP68 4) | KN4-P Anschlussgehäuse Flat screw-on lid | BVC Natural finish |
| BVC KN4-A | Aluminium | ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | IP65 3) | Screw-on lid | Blue, painted (RAL 5022) |
| KN4-P BVC KN4-P | Polypropylene | ■ M20 x 1.5 ■ M12 x 1 coupler connector (4-pin) | IP65 3) | Screw-on lid | White |

| Transmitter | Model T15 | Model T32 |
|---|---|---|
| Transmitter data sheet | TE 15.01 | TE 32.04 |
| Figure |  |  |
| Analogue output | 4 ...20mA | ■ 4 ...20mA ■ HART ® protocol |
| Possible assembly combinations |  |  |
| BVC | ○ | ○ |
| BVS | ○ | ○ |
| BS | ○ | - |
| BSZ / BSZ-K | ○ | ○ |
| BSZ-H / BSZ-HK | ● | ● |
| KN4-P / KN4-A | ○ | ○ |

### Process connection

| 1 | 2 | 3 |
|---|---|---|
| Type of process connection | ■ Flow-through housing ■ Angular housing |  |
| Measuring insert diameter (Pt100 version) | 3 mm[0.12 in] |  |
| Protection tube |  |  |
| Protection tube model | TW61 |  |
| Protection tube diameter | → see tables on page 13 and 14 |  |
| Surface roughness | Per DIN 11866 row A, B | ■ R a < 0.8 μm ■ R a < 0.4 μm electropolished |
| Per DIN 11866 row C, ASME-BPE | ■ R a < 0.76 μm ■ R a < 0.38 μm electropolished |  |
| Others on request |  |  |
| Connection to thermometer | M24 x 1.5 |  |
| Insertion length U 1 , tube length TL and L 1 , protection tube insertion length U 1 | → see tables on page 13 and 14 |  |
| Neck tube lengthM | 125 mm[4.92 in] |  |
| The use of uniform measuring insert lengths, even for different nominal widths of pipes, reduces the inventory of the measuring inserts, particularly for larger plants. Additionally, the measuring insert length is optimised for an on-site calibration, for example with aWIKA temperature dry-well calibrator model CTD9x00. |  |  |
| Other lengths on request |  |  |
| Material (wetted) | Per DIN 11866 row A, B | Stainless steel 1.4435 |
| Per DIN 11866 row C, ASME-BPE | Stainless steel 316L |  |

| Process connection | 2 |
|---|---|
| Sealing combination (option) | Transition from the connection head to the protection tube is effected via an optional sealing combination (polyurethane) of flat gasket and wiper |
|  | ■ Permanently prevents the penetration and depositing of humidity and impurities in this area (IP68) ■ Simplifies cleaning considerably (in combination with the patented BVS head and the hygienic cable gland, an easy to clean and hygienic measuring location results, |
|  | even in areas which are not in contact with the product) |
| Patented hygienic design (for flow-through housing) | ■ Dead-space minimised, invasive temperature measurement ■ Flexible mounting position due to self-draining ■ With horizontal installation, make sure that the pipeline is slightly inclined for self- draining ■ The installation is carried out by means of orbital welding; thus, the weld seams are reproducible and controllable |

### Output signal (transmitter version)

| Analogue output | 2 | 3 |
|---|---|---|
| Transmitter models T15, T32 | 4 ...20mA |  |
| Transmitter model T32 | HART ® protocol |  |
| Factory configuration (transmitter) | → Customer-specific configuration available on request |  |
| Sensor | Pt100 |  |
| Connection method | 3-wire |  |
| Measuring range | Measuring range 0 ... 150 °C [32 ... 302 °F] |  |
| Other measuring ranges are adjustable |  |  |
| Time response |  |  |
| Response time per IEC 60751 1) | Transmitter version | t 50 < 3.2 s or t 90 < 7.3 s + response time of the respective transmitter → see data sheet for the respective transmitter |
| Pt100 version | t 50 < 3.2 s t 90 < 7.3 s |  |

### Operating conditions

| Ambient temperature range | -40 ...+85 °C [-40 ...+185 °F] |
|---|---|
| Storage temperature range | -40 ...+85 °C [-40 ...+185 °F] |

| Description | Region |
|---|---|
| EU declaration of conformity | European Union |
| EMCdirective 1) EN 61326 emission (group 1, class B) and immunity (industrial application) |  |
| Pressure equipment directive For thermowells/protection tubes > DN 25 [1"] and for the associated marking on the measuring instrument or thermowell/protection tube,WIKA confirms conformity with the pressure equipment directive in accordance with the conformity assessment procedure, module H. For thermowells/protection tubes with nominal widths of ≤ DN 25 [1"], a CE marking in |  |

| Logo | Description | Region |
|---|---|---|
|  | EU declaration of conformity ATEX directive Hazardous areas | European Union |
|  | - Ex i Zone Zone Zone Zone Zone Zone - Ex e 2) Zone Zone Zone Zone - Ex n 2) Zone Zone 22 dust II 3D Ex tc IIICTX °C Dc X IECEx - in combination with ATEX Hazardous areas | International |
|  | EAC EMCdirective 1) | Eurasian Economic |
|  | Hazardous areas - Ex i Zone 0 gas 0Ex ia IIC T6 ...T1 Ga X Zone 1 gas 1Ex ia IIC T6 ...T1 Gb X Zone 20 dust Ex ia IIIC T80 ...T440 Da X Zone 21 dust Ex ia IIIC T80 ...T440 Db X - Ex n Zone |  |

| Logo | Description | Region |
|---|---|---|
|  | INMETRO | Brazil |
|  | Hazardous areas Ex i Zone 0 gas Ex ia IIC T3 ...T6 Ga Zone CCC 3) | China |
|  | KCs Hazardous areas Ex i Zone 0 gas Ex ia IIC T4 ...T6 | Korea |
| - | Zone 1 gas Ex ib IIC T4 ...T6 PESO Hazardous areas - Ex i Zone 0 gas Ex ia IIC T1 ...T6 Ga Zone 1 mounting to zone 0 gas Ex ib IIC T3 ...T6 Ga/Gb | India |
|  | PAC Russia Metrology, measurement technology | Russia |
|  | PAC Kazakhstan Metrology, measurement technology | Kazakhstan |
| - | MChS Permission for commissioning | Kazakhstan |
|  | PAC Usbekistan Metrology, measurement technology 3-A 4) Sanitary Standard | Uzbekistan |
| ® | Flow-through housing:yes, for all dimensions Angular housing: yes, from DIN 11866 row A:DN 32 ... 100 DIN 11866 row B:DN 25 ... 80 DIN 11866 row C:DN 1 ½" ... 4" | USA |
| TYPE EL | EHEDG 4) Hygienic Equipment Design Flow-through housing:yes, for all dimensions Angular housing: yes, from DIN 11866 row A:DN 32 ... 100 DIN 11866 row B:DN 25 ... 80 | European Union |

### Certificates

| Certificates | ■ 2.2 test report ■ 3.1 inspection certificate ■ DAkkS calibration certificate, traceable and accredited in accordance with ISO/IEC 17025 ■ Manufacturer&#x27;s declaration regarding regulation (EC) 1935/2004 ■ Certificate of the surface roughness of wetted parts |
|---|---|
| Hygienic certificates | 3-A approval EHEDG approval |

| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
|---|---|---|---|---|---|---|---|
| Nominal width of pipe Workflow Freigabe-08 | Max. permissible pressure in bar | Outer diameter of pipe 11528266.02 E CN108275 02 | Pipe schedule material KLI-ZCHG TW61 DURCHGGEH + BVS F | Pipe length TW61 .STRD-PREISLISTE TW61 | Protection tube insertion length reference modified released UV/DR drawing number 08 18.11.21 THIESR 18.11.21 DILLMAA | Neck tube length 11528274.02 F sheets / . 11528274 02 1 1 |  |
| DN / OD REVISION | PS | Ø D first created 01 06.11.09 | s size nominal raw part Surface texture | TL Check Tol. ISO 2768 - mK | U 1 ALEXANDER SIZE ISO 14405 units comments GG | M WIEGAND SE & Co. KG |  |
| DIN 11866 row A or metric KLI-ZCHG TW61 F.STRD-PREISLISTE | 10 | 25 | 13 by WIKA 11.11.2021 DILLMAA | 1.5 µm z R in 1 4 16 63 - ISO 1302 x y z | 70 ±0.5 ±0.3 ±0.2 ...400 ...120 ...30 >120 >30 >6 ±0.1 ...6 0.5 | 6 ISO 8015 mm [inch] | 129 www.wika.com D-63911 Klingenberg |
| 15 TW61 | 25 drawing number | 19 | 1.5 | 70 | 9 | 126 DEKLI1_EN_A4M_T01 15.04.2020 |  |
| 20 TW61 | 25 reference modified 18.11.21 DILLMAA | 23 . 11528266 02 | 1.5 | 80 | 11 | 124 |  |
| 25 size nominal | 25 released UV/DR comments 08 18.11.21 THIESR | 29 F sheets / 1 1 | 1.5 | 100 | 18 | 117 |  |
| 32 Check language ...400 ...120 ...30 >120 >30 >6 ...6 0.5 Tol. ISO 2768 - mK z | 25 ALEXANDER ISO 8015 SIZE ISO 14405 mm units indicator GG | 35 D-63911 Klingenberg WIEGAND SE & Co. KG | 1.5 | 110 | 18 | 117 |  |
| 40 GER/ENG ±0.5 ±0.3 ±0.2 ±0.1 1 | 25 [inch] | 41 www.wika.com DEKLI1_EN_A4M_T01 15.04.2020 | 1.5 | 120 | 18 | 117 |  |
| 50 | 25 | 53 | 1.5 | 160 | 30 | 105 |  |
| 65 | 16 | 70 | 2.0 | 210 | 30 | 105 |  |
| 80 | 16 | 85 | 2.0 | 260 | 45 | 90 |  |
| 100 | 12.5 | 104 | 2.0 | 310 | 45 | 90 |  |
| DIN 11866 row B or ISO | 8 (13.5) | 25 | 13.5 | 1.6 | 64 | 6 | 129 |
| 10 (17.2) | 25 | 17.2 | 1.6 | 68 | 9 | 126 |  |
| 15 (21.3) | 25 | 21.3 | 1.6 | 72 | 11 | 124 |  |
| 20 (26.9) | 25 | 26.9 | 1.6 | 110 | 11 | 124 |  |
| 25 (33.7) | 25 | 33.7 | 2.0 | 120 | 18 | 117 |  |
| 32 (42.4) | 25 | 42.4 | 2.0 | 130 | 18 | 117 |  |
| 40 (48.3) | 25 | 48.3 | 2.0 | 130 | 18 | 117 |  |
| 50 (60.3) | 25 | 60.3 | 2.0 | 180 | 30 | 105 |  |
| 65 (76.1) | 16 | 76.1 | 2.0 | 220 | 30 | 105 |  |
| 80 (88.9) | 16 | 88.9 | 2.3 | 260 | 45 | 90 |  |
| DIN 11866 row C or ASME BPE | 1/2" | 13.8 | 12.7 | 1.65 | 95.2 | 6 | 129 |
| 3/4" | 13.8 | 19.05 | 1.65 | 101.6 | 9 | 126 |  |
| 1" |  | 25.4 | 1.65 | 108.0 |  |  |  |
| 1 1/2" | 13.8 |  | 1.65 |  | 11 | 124 |  |
|  | 13.8 | 38.1 50.8 | 1.65 | 120.6 | 18 | 117 117 |  |
| 2" 2 1/2" | 13.8 | 63.5 | 1.65 | 146.0 158.8 | 18 30 | 105 |  |
| 3" | 13.8 13.8 | 76.2 | 1.65 | 171.4 | 30 | 105 |  |
| 4" | 13.8 | 101.6 | 2.11 | 209.6 | 45 | 90 |  |

| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 |
|---|---|---|---|---|---|---|---|---|
| Nominal width of pipe KLI-ZCHG TW61 ECKGEHÄUSE F.STRD-PREISLISTE | TW61 | Max. permissible pressure in bar 11528452.02 drawing number 11528452 | Outer diameter of pipe . 02 | Pipe schedule by WIKA 1 first created 05.07.2012 BAUERK | Pipe length 2 | Protection tube insertion length 3 | Neck tube length 4 |  |
| DN / OD material scale TW61 1:1 | PS sheets reference modified released UV/DR 08 18.11.21 THIESR 18.11.21 DILLMAA | Ø D F / 1 1 | s | TL | L 1 | U 1 | M |  |
| DIN 11866 row A or metric by WIKA size nominal raw part µm z R in 1 4 16 63 - ISO 1302 x y z Surface texture first created 01 CN108275 02 09.11.09 05.07.2012 BAUERK | 10 Check indicator Tol. ISO 2768 - mK | 25 ALEXANDER WIEGAND SIZE ISO 14405 units comments GG | 13 SE & Co. KG | 1.5 | 35 | 55 | 14 | 121 |
| ±0.1 ...6 0.5 | 15 language GER/ENG ±0.5 ±0.3 ±0.2 ...400 ...120 ...30 >120 >30 >6 | 25 ISO 8015 mm [inch] | 19 www.wika.com D-63911 Klingenberg | 1.5 | 35 | 55 | 18 | 117 |
| 20 | 25 | 23 DEKLI1_EN_A4M_T01 15.04.2020 | 1.5 | 40 | 63 | 18 | 117 |  |
| 25 | 25 | 29 | 1.5 | 50 | 77 | 30 | 105 |  |
| DIN 11866 row A or metric | 32 | 25 | 35 | 1.5 | 55 | 87 | 30 | 105 |
| 40 | 25 | 41 | 1.5 | 60 | 97 | 30 | 105 |  |
| ® JANUARY 2019 TYPE EL CLASS I | 50 | 25 | 53 | 1.5 | 80 | 126 | 30 | 105 |
| 65 | 16 | 70 | 2.0 | 105 | 165 | 45 | 90 |  |
| 80 | 16 | 85 | 2.0 | 130 | 201 | 45 | 90 |  |
| 100 | 12.5 | 104 | 2.0 | 155 | 241 | 45 | 90 |  |
| DIN 11866 row B or ISO | 8 (13.5) | 25 | 13.5 | 1.6 | 32 | 55 | 14 | 121 |
| 10 (17.2) | 25 | 17.2 | 1.6 | 34 | 55 | 16 | 119 |  |
| 15 (21.3) | 25 | 21.3 | 1.6 | 36 | 58 | 18 | 117 |  |
| 20 (26.9) | 25 | 26.9 | 1.6 | 55 | 81 | 30 | 105 |  |
| DIN 11866 row B or ISO ® TYPE EL CLASS I | 25 (33.7) | 25 | 33.7 | 2.0 | 60 | 91 | 30 | 105 |
|  | 32 (42.4) | 25 | 42.4 | 2.0 | 65 | 102 | 30 | 105 |
|  | 40 (48.3) | 25 | 48.3 | 2.0 | 65 | 108 | 30 | 105 |
| JANUARY 2019 | 50 (60.3) | 25 | 60.3 | 2.0 | 90 | 145 | 45 | 90 |
|  | 65 (76.1) | 16 | 76.1 | 2.0 | 110 | 173 | 45 | 90 |
|  | 80 (88.9) | 16 | 88.9 | 2.3 | 130 | 203 | 45 | 90 |
| DIN 11866 row C or ASME BPE | 1/2" | 13.8 | 12.7 | 1.65 | 47.6 | 71 | 14 | 121 |
| 3/4" | 13.8 | 19.05 | 1.65 | 50.8 | 71 | 18 | 117 |  |
| 1" | 13.8 | 25.4 | 1.65 | 54.0 | 79 | 18 | 117 |  |
| DIN 11866 row C or ASME BPE | 1 1/2" | 13.8 | 38.1 | 1.65 | 60.3 | 94 | 30 | 105 |
| 2" | 13.8 | 50.8 | 1.65 | 73.0 | 118 | 30 | 105 |  |
| ® JANUARY 2019 TYPE EL CLASS I | 2 1/2" | 13.8 | 63.5 | 1.65 | 79.4 | 134 | 45 | 90 |
| 3" | 13.8 | 76.2 | 1.65 | 85.7 | 150 | 45 | 90 |  |
| 4" | 13.8 | 101.6 | 2.11 | 104.8 | 190 | 45 | 90 |  |

## Documents

- [ds_te6023_en_co_a2bd1e](https://avigan.com.ua/wika/docs/wika/ds_te6023_en_co.pdf) — PDF, 1754 KB, en

## Configurator

This product has an interactive configurator on its product page.

```
TR22-B- Z- - C - - -S-ZZ
```