# Resistance thermometer — TR10-B

> Resistance thermometer for additional thermowell with Pt100/Pt1000 sensors and various connection heads

- Catalog code: tr10_b
- Model: TR10-B
- Category: Temperature measurement › Resistance thermometers
- Product page: https://avigan.com.ua/wika/en/products/tr10_b
- Price: 206.37 EUR
- Supplier: AVIGAN (TOV VIKA Prylad), official WIKA partner in Ukraine

## Overview

Model TR10-B Resistance thermometer ✓For additional thermowell ✓High quality ►To the information!

## Applications

- Machine building, plant and vessel construction
- Energy and power plant technology
- Chemical industry
- Food and beverage industry
- Sanitary, heating and air-conditioning technology

## Special features

- Sensor ranges from -196 ... +600 °C [-320 ... +1,112 °F]
- For mounting in all standard thermowell designs
- Spring-loaded measuring insert (replaceable)
- Pt100 or Pt1000 sensors
- Explosion-protected versions are available for many approval types

## Specifications

| Parameter | Value |
|---|---|
| model | TR10-B |
| product family | TR10 |
| short description | Resistance thermometer for additional thermowell with Pt100/Pt1000 sensors and various connection heads |
| subcategory | Resistance thermometer |
| certifications | CE; RoHS; REACH; EAC; NAMUR NE 024; DNV GL |
| explosion protection | Ex i (gas) Zone 0, 1, 2; Ex i (dust) Zone 20, 21, 22; Ex e (gas) Zone 1, 2; Ex t (dust) Zone 21, 22; Ex nA (gas) Zone 2; ATEX; IECEx; EACEx; CCC; INMETRO; NEPSI; KCs; PESO; Ex Ukraine; ATEX: II 1/2G Ex ia IIC T3...T6 Ga/Gb, II 1/2D Ex ia IIIC T65°C Da/Db; IECEx: Ex ia IIC T3...T6 Ga/Gb, Ex ia IIIC T65°C Da/Db; EACEx: Ex ia IIC T6...T1 Ga X, Ex ia IIC T6...T1 Gb X, Ex ia IIC T80...T440°C Da X, Ex ia IIC T80...T440°C Db X, Ex eb IIC T6...T1 Gb X, Ex ec IIC T6...T1 Gc X, Ex nA IIC T6...T1 Gc X, Ex tb IIC T85°C Db X, Ex tc IIC T85°C Dc; CCC: Ex ia IIC T3...T6 Ga, Ex ia IIC T3...T6 Ga/Gb, Ex ia IIC T125...T65°C Da, Ex ia IIC T125...T65°C Da/Db; NEPSI: Ex ia IIC T1...T6 Ga, Ex ia IIC T1...T6 Gb, Ex ec IIC T1...T6 Gc, Ex tb IIC T135°C Db, Ex tc IIC T135°C Dc; KCs: Ex ia IIC T1~T6 Ga, Ex ia IIC T1~T6 Gb, Ex ia IIC T1~T6 Ga/Gb, Ex ia D20 T65/T95/T125°C, Ex ia D21 T65/T95/T125°C; INMETRO: Ex ia IIC T3...T6 Ga, Ex ia IIC T3...T6 Ga/Gb, Ex ia IIC T125...T65°C Da, Ex ia IIC T125...T65°C Da/Db; Ex Ukraine: Ex ia IIC T3...T6 Ga, Ex ia IIC T3...T6 Ga/Gb, Ex ia IIC T125...T65°C Da, Ex ia IIC T125...T65°C Da/Db; PESO: Ex ia IIC T3...T6 Ga, Ex ia IIC T3...T6 Ga/Gb, Ex ia IIC T125...T65°C Da, Ex ia IIC T125...T65°C Da/Db |
| safety integrity level | SIL 2 |
| ambient temp | -40 … 80 °C |
| ip rating | IP65 |
| languages in datasheet | en; de |
| temperature ranges | -196 … 600 °C; -196 … 450 °C; -50 … 500 °C; -50 … 250 °C; -30 … 300 °C; 0 … 150 °C; -100 … 450 °C |
| sensor type | Pt100 |
| sensor class | B |
| probe diameter mm | 3 |
| probe material | 1.4571; 316L |
| has thermowell | true |
| process connections | M24 x 1.5; ½ NPT; M20 x 1.5; ½NPT; ¾NPT; ¾ NPT |
| process temp | -196 … 600 °C |
| output signal | 4-20mA |
| electrical connection | Spring-loaded terminal block |
| accuracy class | ± (0.30 + 0.0050 \|t\|) °C |

## Datasheet tables

| 1 | 2 | 3 | 4 | 5 | 6 |
|---|---|---|---|---|---|
| Approval | Explosion protection |  |  |  |  |
| Ex i (gas) Zone 0, 1, 2 | Ex i (dust) Zone 20, 21, 22 | Ex e (gas) Zone 1, 2 | Ex t (dust) Zone 21, 22 | Ex nA (gas) Zone 2 |  |
| ATEX | x | x | x | x | x |
| IECEx | x | x | x | x | x |
| ECASEx | - | - | x | x | x |
| EACEx | x | x | x | x | x |
| Ex Ukraine | x | x | - | - | - |
| INMETRO | x | x | - | - | - |
| CCC | x | x | x | x | - |
| NEPSI | x | x | - | - | - |
| KCs | x | - | - | - | - |
| PESO | x | - | - | - | - |

| Model | 2 | Material | Thread size of cable inlet | Ingress pro- tection (max.) 1) 2) IEC/EN 60529 | Cap | Surface | Connection to neck tube |
|---|---|---|---|---|---|---|---|
| BSZ, BSZ-K | BSZ | Aluminium | ■ M20 x 1.5 ■ ½NPT | IP65 | Spherical hinged cov- er with cylinder head screw | Blue, painted (RAL 5022) | ■ M24 x 1.5 ■ ½NPT |
| BSZ-H, BSZ-HK BSZ-K | BSS Plastic | BSS-H ■ M20 x 1.5 ■ ½NPT | BVS IP65 | BVS (NuG) Spherical hinged cov- er with cylinder head screw | 1/4000 Black | 5/6000 M24 x 1.5 |  |
| BSZ-H, BSZ-HK | BSZ-H | Aluminium | ■ M20 x 1.5 ■ ½NPT | IP65 | Raised hinged cover with cylinder head screw | Blue, painted (RAL 5022) | ■ M24 x 1.5 ■ ½NPT |
| BSS BSZ-H (2 x cable outlet) | BSS-H Aluminium | BVS ■ 2 x M20 x 1.5 ■ 2x½NPT | BVS (NuG) IP65 | 5/6000 1/4000 Raised hinged cover with cylinder head screw | 7/8000 Blue, painted (RAL 5022) | 7/8000 DIH50 M24 x 1.5 |  |
| BSZ-H / DIH10 3) | Aluminium | ■ M20 x 1.5 ■ ½NPT | IP65 | Raised hinged cover with cylinder head screw | Blue, painted (RAL 5022) | ■ M24 x 1.5 ■ ½NPT |  |
| BSZ-H /TND 4) | Aluminium | ■ M20 x 1.5 ■ ½NPT | IP65 | Raised hinged cover with cylinder head screw | Blue, painted (RAL 5022) | ■ M24 x 1.5 ■ ½NPT |  |
| BSZ-HK | Plastic | ■ M20 x 1.5 ■ ½NPT | IP65 | Raised hinged cover with cylinder head screw | Black | M24 x 1.5 |  |
|  | BS | Aluminium | ■ M20 x 1.5 ■ ½NPT | IP65 | Flat cover with 2 screws | Blue, painted (RAL 5022) | ■ M24 x 1.5 ■ ½NPT |
| BS | BSZ, BSZ-K BSS | BSZ-H, BSZ-HK Aluminium | BSS ■ M20 x 1.5 ■ ½NPT | BSS-H IP65 | BVS Spherical hinged cover with clamping lever | BVS (NuG) Blue, painted (RAL 5022) | 1/4000 ■ M24 x 1.5 ■ ½NPT |
| BSS | BSS-H BSS-H | BVS Aluminium | BVS (NuG) ■ M20 x 1.5 ■ ½NPT | 5/6000 1/4000 IP65 | 7/8000 Raised hinged cover with clamping lever | 7/8000 DIH50 andere Blue, painted (RAL 5022) | Anschlussgehäuse ■ M24 x 1.5 ■ ½NPT |
| BSS-H | BVS BVS | BVS (NuG) Stainless steel | 1/4000 M20 x 1.5 | 5/6000 IP65 | 7/8000 DIH50 7/8000 Precision-cast screw- on lid | andere Anschlussgehäuse Natural finish, electropo- lished | KN4-P M24 x 1.5 |

| 1 | 2 | 3 | 4 | 5 | 6 | 7 |
|---|---|---|---|---|---|---|
| Model | Explosion protection |  |  |  |  |  |
|  | Without | Ex i (gas) Zone 0, 1, 2 | Ex i (dust) Zone 20, 21, 22 | Ex e (gas) Zone 1, 2 | Ex t (dust) Zone 21, 22 | Ex nA (gas) Zone 2 |
| BSZ | x | x | x | x 1) | x 1) | x 2) |
| BSZ-K | x | x | - | - | - | - |
| BSZ-H | x | x | x | x 1) | x 1) | x 2) |
| BSZ-H (2 x cable outlet) | x | x | x | x 1) | x 1) | x 2) |
| BSZ-H / DIH10 3) | x | x | - | - | - | - |
| BSZ-H /TND 4) | x | x | - | - | - | - |
| BSZ-HK | x | x | - | - | - | - |
| BS | x | x | x | - | - | - |
| BSS | x | x | - | - | - | - |
| BSS-H | x | x | - | - | - | - |
| BVS | x | x | - | - | - | - |

| Model | 2 | Material | Thread size of cable inlet | Ingress protection (max.) 1) 2) IEC/EN 60529 | Cap | Surface | Connection to neck tube |
|---|---|---|---|---|---|---|---|
| KN4-P | KN4-A | Aluminium | ■ ½NPT ■ M20 x 1.5 | IP65 | Screw-on lid | Blue, painted (RAL 5022) | ■ M24 x 1.5 ■ ½NPT |
| KN4-P 3) | Polypropylene | ½NPT | IP65 | Screw-on lid | White | ½NPT |  |
| 1/4000 | BVC 1/4000 | Aluminium | ■ ½NPT ■ ¾NPT ■ M20 x 1.5 | IP66 | Screw-on lid | Blue, painted (RAL 5022) | ½NPT |
| 5/6000 1/4000 | 7/8000 Stainless steel | 7/8000 DIH50 andere ■ ½NPT ■ ¾NPT ■ M20 x 1.5 | KN4-P Anschlussgehäuse IP66 | BVC Screw-on lid | Natural finish | ½NPT |  |
| 7/8000 | 7/8000 | Aluminium | ■ ½NPT ■ ¾NPT ■ M20 x 1.5 | IP66 | Screw-on lid | Blue, painted (RAL 5022) | ½NPT |
| 7/8000 DIH50 7/8000 | andere Anschlussgehäuse Stainless steel | KN4-P BVC ■ ½NPT ■ ¾NPT ■ M20 x 1.5 | IP66 | Screw-on lid | Natural finish | ½NPT |  |
| 7/8000 DIH50 | 7/8000 / DIH50 4) | Aluminium | ■ ½NPT ■ ¾NPT ■ M20 x 1.5 | IP66 | Screw-on lid | Blue, painted (RAL 5022) | ½NPT |
| andere Anschlussgehäuse 7/8000 / DIH50 4) | KN4-P Stainless steel | BVC ■ ½NPT ■ ¾NPT ■ M20 x 1.5 | IP66 | Screw-on lid | Natural finish | ½NPT |  |
|  | PIH-L | Aluminium | ■ ½NPT/ closed ■ M20 x 1.5 / closed ■ 2x½NPT ■ 2 x M20 x 1.5 | IP66 | Screw-on lid, flat | Blue upper body, painted (RAL 5022) | ■ ½NPT ■ M20 x 1.5 |
|  | Grey lower body, painted (RAL 7032) |  |  |  |  |  |  |
|  | PIH-H | Aluminium | ■ ½NPT ■ M20 x 1.5 ■ 2x½NPT ■ 2 x M20 x 1.5 | IP66 | Screw-on lid, high | Blue upper body, painted (RAL 5022) | ■ ½NPT ■ M20 x 1.5 |
|  | Grey lower body, painted (RAL 7032) |  |  |  |  |  |  |
|  | PIH-W / TND 5) | Aluminium | ■ ½NPT ■ M20 x 1.5 ■ 2x½NPT ■ 2 x M20 x 1.5 | IP66 | Screw-on lid, high | Blue upper body, painted (RAL 5022) Grey lower painted | ■ ½NPT ■ M20 x 1.5 |
|  | body, (RAL 7032) |  |  |  |  |  |  |

| 1 | 2 | 3 | 4 | 5 | 6 | 7 |
|---|---|---|---|---|---|---|
| Model | Explosion protection |  |  |  |  |  |
|  | Without | Ex i (gas) Zone 0, 1, 2 | Ex i (dust) Zone 20, 21, 22 | Ex e (gas) Zone 1, 2 | Ex t (dust) Zone 21, 22 | Ex nA (gas) Zone 2 |
| KN4-A | x | x | - | - | - | - |
| KN4-P 1) | x | - | - | - | - | - |
| 1/4000 | x | x | x | x | x | x |
| 7/8000 | x | x | x | x | x | x |
| 7/8000 / DIH50 2) | x | x | x | - | - | - |
| PIH-L / PIH-H | x | x | x | x | x | x |
| PIH-W /TND 3) | x | x | x | x | x | x |

| Cable inlet | Colour | Ingress protection | Thread size of cable inlet | Min./Max. ambient temperature |
|---|---|---|---|---|
| Standard cable inlet 2) | Natural finish | IP65 | M20 x 1.5 | -40 ...+80 °C [-40 ...+176 °F] |
| Plastic cable gland (cable Ø6...10 mm) 2) | ■ Black ■ Grey | IP66 | ■ M20 x 1.5 ■ ½NPT | -40 ...+80 °C [-40 ...+176 °F] |
| Plastic cable gland (cable Ø 6 ... 10 mm), Ex e 2) | ■ Light blue ■ Black | IP66 | ■ M20 x 1.5 ■ ½NPT | ■ -20 ...+80 °C [-4 ...+176 °F] ■ -40 ...+70 °C [-40 ...+158 °F] |
| Nickel-plated brass cable gland (cable Ø 6 ... 12 mm) | Natural finish | IP66 | ■ M20 x 1.5 ■ ½NPT | -60 3) / -40 ...+80 °C [-76 / -40 ...+176 °F] |
| Nickel-plated brass cable gland (cable Ø6...12 mm), Ex e | Natural finish | IP66 | ■ M20 x 1.5 ■ ½NPT | -60 3) / -40 ...+80 °C [-76 / -40 ...+176 °F] |
| Stainless steel cable gland (cable Ø 7 ... 12 mm) | Natural finish | IP66 | ■ M20 x 1.5 ■ ½NPT | -60 3) / -40 ...+80 °C [-76 / -40 ...+176 °F] |
| Stainless steel cable gland (cable Ø 7 ... 12 mm), Ex e | Natural finish | IP66 | ■ M20 x 1.5 ■ ½NPT | -60 3) / -40 ...+80 °C [-76 / -40 ...+176 °F] |
| Plain threaded | - | IP00 | ■ M20 x 1.5 ■ ½NPT | - |
| 2 x plain threaded 5) | - | IP00 | ■ 2 x M20 x 1.5 ■ 2x½NPT | - |
| Built-in connector (male) M12 x 1 (4-pin) Built-in connector (female) M12 x 1 (4-pin) | - | IP65 | M20 x 1.5 | -40 ...+80 °C [-40 ...+176 °F] |
| Sealing plugs for shipping | Transparent | - | ■ M20 x 1.5 ■ ½NPT | -40 ...+80 °C [-40 ...+176 °F] |

| 1 | 2 | 3 | 4 | 5 | 6 | 7 |
|---|---|---|---|---|---|---|
| Cable inlet | Explosion protection |  |  |  |  |  |
|  | With- out | Ex i (gas) Zone 0, 1, 2 | Ex i (dust) Zone 20, 21, 22 | Ex e (gas) Zone 1, 2 | Ex t (dust) Zone 21, 22 | Ex nA (gas) Zone 2 |
| Standard cable inlet 1) 3) | x | x | - | - | - | - |
| Plastic cable gland 1) | x | x | - | - | - | - |
| Plastic cable gland (light blue), Ex e 1) | x | x | x | - | - | - |
| Plastic cable gland (black), Ex e 1) | x | x | x | x | x | x |
| Brass cable gland, nickel-plated | x | x | x | - | - | - |
| Brass cable gland, nickel-plated, Ex e | x | x | x | x | x | x |
| Stainless steel cable gland | x | x | x | - | - | - |
| Stainless steel cable gland, Ex e | x | x | x | x | x | x |
| Plain threaded | x | x | x 5) | x 5) | x 5) | x 5) |
| 2 x plain threaded 2) | x | x | x 5) | x 5) | x 5) | x 5) |
| Junction box M12 x 1 (4-pin) 3) | x | x 4) | x 4) | - | - | - |
| Sealing plugs for shipping | Not applicable, transport protection 5) |  |  |  |  |  |

### Measuring insert

| 1 | 2 | 3 |
|---|---|---|
| Version | Vibration-resistant mineral-insulated metal-sheathed cable (MIMS cable) |  |
| Optimal heat transfer | Requirement | ■ Correct measuring insert length ■ Correct measuring insert diameter |
|  | Bore diameter of the thermowell / protection tube | Max. 1 mm[0.039 in] larger than the measuring insert diameter |
|  | Gap width | With gap widths > 0.5 mm[> 0.020 in] between thermowell / protection tube and measuring insert: →Negative impact on heat transfer →Unfavourable response behaviour of the thermometer |
| Insertion length | When mounting the measuring insert into a thermowell, it is very important to determine the correct insertion length (= thermowell length for tip thicknesses of ≤ 5.5 mm[≤ 0.217 in]). In order to ensure that the meas - uring insert is firmly pressed down onto the bottom of the thermowell / protection tube, the insert must be spring-loaded (spring travel: max. 10 mm[0.394 in]). |  |
| Spring travel | Max. 10 mm[0.394 in] |  |

| 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|
| Measuring insert diameter Ø d in mm[in] | Code number per DIN 43735 | Tolerance inmm | Sheath material |  |
| 3 [0.118] | Standard | 30 | 3 +0.05 -0.05 | ■ 1.4571 ■ 316L |
| 6 [0.236] | Standard | 60 | 6 0 -0.1 |  |
| 8 [0.315] (6 [0.236] with sleeve) | Standard | - | 8 0 -0.1 | ■ 1.4571 |
| 8 [0.315] | Standard | 80 | 8 0 -0.1 | ■ 1.4571 ■ 316L |

| Transmitter models | Model T15 | Model T38 |
|---|---|---|
| Transmitter data sheet | TE 15.01 | TE 38.01 |
| Figure |  |  |
| Output signal |  |  |
| 4 ...20mA | x | x |
| HART ® protocol | - | x |
| WIKATrue Drift Detection | - | x |
| → See data sheet SP 05.26 |  |  |
| Connection method | ■ 1 x 2-wire ■ 1 x 3-wire ■ 1 x 4-wire | ■ 1 x 2-wire ■ 1 x 3-wire ■ 1 x 4-wire ■ 2 x 2-wire ■ 2 x 3-wire |
| Measuring current | < 0.2mA | < 0.33mA |
| Explosion protection | Ex version possible |  |
| Mounting types |  |  |
| Mounting onto the measuring insert | With mounting on the measuring insert, the transmitter replaces the terminal block and is fixed directly to the terminal plate of the measuring insert. |  |
| Mounted within the cover of the connection head | Mounting the transmitter in the cover of the connection head is preferable to mounting it on the measuring insert.With this mounting type, for one, a better thermal insulation is ensured, and in addition, exchange and mounting for servicing is simplified. |  |

| Possible transmitter mounting positions | Model T15 | Model T38 |
|---|---|---|
| BSZ | ○ | ○ |
| BSZ-K | ○ | ○ |
| BSZ-H | ● | ● |
| BSZ-H (2 x cable outlet) | ● | ● |
| BSZ-H / DIH10 | ○ | ○ |
| BSZ-H /TND | ○ | ○ |
| BSZ-HK | ● | ● |
| BS | ○ | - |
| BSS | ○ | ○ |
| BSS-H | ● | ● |
| BVS | ○ | ○ |
| KN4-A | ○ | ○ |
| KN4-P | ○ | ○ |
| 1/4000 | ○ | ○ |
| 7/8000 | ○ | ○ |
| 7/8000 / DIH50 | ○ | ○ |
| PIH-L / PIH-H | ○ | ○ |
| PIH-W | ○ | ○ |

### Thread sizes

| Neck tube design | Diameter | Thread to the thermowell / protection tube | Connection to head | Materials 1) |
|---|---|---|---|---|
| Neck tube per DIN 43772 | ■ 12 x 1.5mm [0.472 x 0.059 in] ■ 12 x 2.5mm [0.472 x 0.098 in] | ■ G½B ■ G¾B ■ G¼B ■ M20 x 1.5 ■ M18 x 1.5 ■ M14 x 1.5 ■ ½NPT ■ ¾NPT ■ G½Bcompression fitting (metal ferrule) ■ G¾Bcompression fitting (metal ferrule) ■ M18 x 1.5 compression fitting (metal ferrule) ■ M20 x 1.5 compression fitting (metal ferrule) ■ G½Bunion nut ■ G¾Bunion nut ■ M20 x 1.5 union nut ■ G½Bmale nut ■ G¾Bmalenut ■ M20 x 1.5 male nut ■ Without threaded connection, plain | M24 x 1.5 (swivel connection) | 1.4571 |
| Neck tube per DIN 43772 | 14 x 2.5mm [0.551 x 0.098 in] | ■ G½B ■ G¾B ■ G¼B ■ M20 x 1.5 ■ M18 x 1.5 ■ M14 x 1.5 ■ ½NPT ■ ¾NPT ■ G½Bunion nut ■ G¾Bunion nut ■ M20 x 1.5 union nut ■ G½Bmale nut ■ G¾Bmalenut |  |  |
| Neck tube with counter nut to head | 14 x 2.5mm [0.551 x 0.098 in] | ■ ½NPT ■ ¾NPT ■ G½B ■ G¾B ■ G¼B ■ M14 x 1.5 ■ M18 x 1.5 ■ M20 x 1.5 | M20 x 1.5 (with counter nut) | 1.4571 |
| Double threaded hex bushing (with hexago- nal spanner flats) | - | ■ G½B ■ G¾B ■ G¼B ■ ½NPT ■ ¾NPT ■ M14 x 1.5 ■ M18 x 1.5 ■ M20 x 1.5 | M24 x 1.5,½NPT | 1.4571 |
| &#x27;Nipple-union-nipple&#x27; neck tube 2) | ~ 22 mm[~ 0.9 in] | ½NPT | ½NPT | 316 |
| Double threaded hex bushing (tube section) | ~ 22 mm[~ 0.9 in] | ½NPT | ½NPT | 316 |

### Neck length

| Neck tube design | Neck length | Min./Max. neck length |
|---|---|---|
| Neck tube per DIN 43772 | 150 mm[~ 6 in] | ■ 30 mm[~ 1.2 in] ■ 500 mm[~ 20 in] |
| Neck tube per DIN 43772, plain | 150 mm[~ 6 in] | ■ 75 mm[~ 3 in] ■ 900 mm[~ 35 in] |
| Neck tube with counter nut to head | 150 mm[~ 6 in] | ■ 75 mm[~ 3 in] ■ 250 mm[~ 10 in] |
| Double threaded hex bushing (with hexagonal spanner flats) |  |  |
| M24 x 1.5 to connection head, parallel thread to thermowell / protection tube | 13 mm[0.512 in] | - |
| ½NPTto connection head, parallel thread to thermowell / protection tube | ~ 25 mm[1 in] | - |
| M24 x 1.5 to connection head, tapered thread to thermowell / protection tube | ~ 25 mm[1 in] | - |
| ½NPTto connection head, tapered thread to thermowell / protection tube | ~ 25 mm[1 in] | - |
| &#x27;Nipple-union-nipple&#x27; neck tube | ~ 150 mm[6 in] | ■ ~ 75 mm[3 in] ■ ~ 250 mm[10 in] |
| Double threaded hex bushing (tube section) | ~ 50 mm[2 in] | ■ ~ 50 mm[2 in] ■ ~ 250 mm[10 in] |

### Operating conditions

| Ambient and storage temperature | ■ -40 ...+80 °C [-40 ...+176 °F] ■ -60 1) ...+80 °C [-76 ...+176 °F] |
|---|---|
| Vibration resistance | The information on vibration resistance refers to the tip of the measuring insert. |

### Examples of vibration resistance

| 1 | Measuring insert ∅6mm[0.236 in] | Measuring insert ∅3mm[0.118 in] |
|---|---|---|
| Thermometer version |  |  |
| Sensor | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) |
| Connection method | ■ 3-wire ■ 4-wire | ■ 3-wire ■ 4-wire (only 1 x Pt100) |
| Diameter | ∅6mm[0.236 in] | ∅3mm[0.118 in] |
| Insertion length (A) + neck length (N) | 100 ... 1,100 mm[~ 4 …43in] | 100 ... 1,100 mm[~ 4 …43in] |
| Material | ■ Stainless steel 1.4571 ■ Stainless steel 316L | ■ Stainless steel 1.4571 ■ Stainless steel 316L |
| Vibration resistance |  |  |
| Standard probe tip, (max. 3g amplitude / 6g peak-to-peak) | x | x |
| Vibration-resistant probe tip (max. 10g amplitude / 20g peak-to-peak) | x | x |
| Highly vibration-resistant probe tip (max. 25g amplitude / 50g peak-to-peak) | x | x |
| Extremely vibration-resistant probe tip (max. 50g amplitude / 100g peak-to-peak) | x | - |

| First numeral | Degree of protection / Short de- scription | Test parameters |
|---|---|---|
| Degrees of protection against solid foreign bodies (defined by the 1st numeral) |  |  |
| 5 | Dust-protected | Per IEC/EN 60529 |
| 6 | Dust-tight | Per IEC/EN 60529 |
| Degrees of protection against water (defined by the 2nd numeral) |  |  |
| 4 | Protected against splash water | Per IEC/EN 60529 |
| 5 | Protected against water jets | Per IEC/EN 60529 |
| 6 | Protected against powerful water jets | Per IEC/EN 60529 |

### Examples of response time for measuring insert

| 1 | Measuring insert ∅ 6 mm[0.236 in] | Measuring insert ∅ 3 mm[0.118 in] |
|---|---|---|
| Thermometer version |  |  |
| Sensor | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) |
| Connection method | ■ 3-wire ■ 4-wire | ■ 3-wire ■ 4-wire (only 1 x Pt100) |
| Diameter | 6 mm[0.236 in] | ∅3mm[0.118 in] |
| Insertion length (A) + neck length (N) | 100 ... 1,100 mm[4 ... 43 in] | 100 ... 1,100 mm[4 ... 43 in] |
| Material | ■ Stainless steel 1.4571 ■ Stainless steel 316L | ■ Stainless steel 1.4571 ■ Stainless steel 316L |
| Response time in seconds (+/- 10 %) |  |  |
| t 0.5 | 3.8 | 2.8 |
| t 0.63 | 4.8 | 3.5 |
| t 0.9 | 8.6 | 6.6 |

| 1 | TW10 Flanged thermowell | TW25 Weld-in thermowell | TW55 Weld-in thermowell (DIN 43772) |
|---|---|---|---|
| Thermometer version |  |  |  |
| Sensor | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) | ■ 1 x Pt100 (thin film) ■ 2 x Pt100 (thin film) |
| Connection method | ■ 3-wire ■ 4-wire | ■ 3-wire ■ 4-wire | ■ 3-wire ■ 4-wire (only 1 x Pt100) |
| Thermowell / Protection tube | TW10, tapered, 12.7/20.0mm | TW25-B, straight, 27.0mm | TW55-6 form 4, 12.5/24.0mm |
| Bore / Tip thickness | 6.6 mm/ 6.4 mm[0.26 in / 0.25 in] | 6.6 mm/ 6.4 mm[0.26 in / 0.25 in] | 7.0 mm/ min.4mm |
| Diameter | ∅6mm[0.236 in] | ∅6mm[0.236 in] | ∅6mm[0.236 in] |
| Insertion length U | 175 mm[approx. 7 in] | 250 mm[approx. 10 in] | 65 mm(L = 200 mm) |
| Material | Stainless steel 316L | Stainless steel 316L | Stainless steel 316L |
| Response time in seconds (+/- 10 %) |  |  |  |
| t 0.5 | 28 | 26 | 33 |
| t 0.63 | 36 | 33 | 45 |
| t 0.9 | 80 | 68 | 114 |

### Thermowell / Protection tube selection

| Representation | Model | Data sheet |
|---|---|---|
|  | TW10 | ■ TW95.10 ■ TW95.11 ■ TW95.12 |
|  | TW15 | TW95.15 |
|  | TW20 | TW95.20 |
|  | TW25 | TW95.25 |
|  | TW30 | TW95.30 |
|  | TW45 | TW95.45 |
|  | TW50 | TW95.50 |
|  | TW55 | TW95.55 |

| Logo | Description | Country |
|---|---|---|
|  | EU declaration of conformity EMCDirective 1) EN 61326 emission (group 1, class B) and immunity | European Union |

## Documents

- [ds_te6002_en_co_129973](https://avigan.com.ua/wika/docs/wika/ds_te6002_en_co.pdf) — PDF, 2671 KB, en
- [ds_sp0525_en_co_ccaa3e](https://avigan.com.ua/wika/docs/wika/ds_sp0525_en_co.pdf) — PDF, 792 KB, en
- [ds_sp0526_en_co_23105f](https://avigan.com.ua/wika/docs/wika/ds_sp0526_en_co.pdf) — PDF, 569 KB, en

## Configurator

This product has an interactive configurator on its product page.

```
TR10-B- Z- - - - T- - - -Z-ZZ
```