# Measuring insert — TC10-K

> For thermocouple with flameproof enclosure (TC10-L)

- Catalog code: tc10_k
- Model: TC10-K
- Category: Temperature measurement › Thermocouples
- Product page: https://avigan.com.ua/wika/en/products/tc10_k
- Price: on request
- Supplier: AVIGAN (TOV VIKA Prylad), official WIKA partner in Ukraine

## Overview

The measuring inserts for thermocouples described here are designed for installation in a thermometer model TC10-L. Operation without thermowell is ...

## Applications

- Replacement measuring insert for servicing

## Special features

- Sensor ranges from -40 ... +1,200 °C [-40 ... +2,192 °F]
- Made of mineral-insulated sheathed measuring cable
- For all standard thermowell designs
- Spring-loaded design
- Explosion-protected versions

## Specifications

| Parameter | Value |
|---|---|
| model | TC10-K |
| product family | TC10 |
| short description | Measuring insert for thermocouple with flameproof enclosure, spring-loaded design, for use in hazardous areas. |
| subcategory | Measuring insert for thermocouple |
| certifications | RoHS; IECEx; EAC; INMETCO; NEPSI; KCs - KOSHA; PESO; GOST; KazInMetr; MTSCHS; BelGIM; UkrSEPRO; Uzstandard; EU Declaration of Conformity; EMC directive EN 61326; 2.2 test report; 3.1 inspection certificate; DKD/DAkkS calibration certificate; EMC EN 61326; CE; ATEX |
| explosion protection | Ex ia IIC T1...T6 Gb; Ex ib IIC T3...T6 Gb; Ex ib IIC T3 ~ T6; Ex ib IIC T4...T6; ATEX: II 2G Ex ia IIC T1 ... T6 Gb; IECEx: Ex ia; EAC: Ex ia IIC T1 ... T6 Gb; INMETRO: 1 Ex ib IIC T3/T4/T5/T6; NEPSI: Ex ib IIC T3 ... T6; KCs-KOSHA: Ex ib IIC T3 ~ T6; PESO: Ex ib IIC T3 ... T6 Gb; GOST; KazInMetr; MTSCHS; BelGIM; UkrSEPRO; Uzstandard; Ex d; 1 Ex ib IIC T3/T4/T5/T6 |
| safety integrity level | SIL 2 (in combination with model T32.1S transmitter) |
| ambient temp | -40 … 1200 °C |
| ip rating | IP65 |
| languages in datasheet | en; de |
| free form specs | sensor types: K, J, E, N, T (single or dual thermocouple); measuring point: Ungrounded (standard), Grounded (optional); transmitter options: T16, T32, T53, T91.10; output signals: 4...20 mA, HART, FOUNDATION™ Fieldbus, PROFIBUS PA; sheath material: Stainless steel (standard), other materials on request; measuring insert diameter mm: 6, 8; measuring insert length mm: 75 to 1100 (tolerance +2/0 or +3/0); spring travel mm: max 10; vibration resistance: 6 g peak-to-peak (in combination with TC10-L) |

## Datasheet tables

| Logo | Description | Country |
|---|---|---|
|  | EU declaration of conformity ■ EMCdirective 1) EN 61326 emission (group 1, class B) and immunity (industrial application) ■ RoHS directive | European Union |
|  |  | International |
|  | IECEx (option) (in conjunction with ATEX) Hazardous areas - Ex i Zone 1 gas [Ex ia IIC T1 ...T6 Gb] EAC (option) | Eurasian Economic Community |
|  | Hazardous areas Ex i Zone 1 gas [1 Ex ib IIC T3/T4/T5/T6] |  |
|  | INMETCO (option) Hazardous areas Ex i Zone 1 gas [Ex ib IIC T3 ...T6 Gb] | Brazil |
|  | NEPSI (option) Hazardous areas Ex i Zone 1 gas [Ex ib IIC T3 ~ T6] | China |
|  | KCs - KOSHA (option) Hazardous areas - Ex i Zone 1 gas [Ex ib IIC T4 ...T6] | South Korea |
| - | PESO (option) Hazardous areas | India |
|  | GOST (option) | Russia |
|  | KazInMetr (option) Metrology, measurement technology | Kazakhstan |
| - | MTSCHS (option) Permission for commissioning | Kazakhstan |
|  | BelGIM (option) Metrology, measurement technology | Belarus |
|  | UkrSEPRO (option) Metrology, measurement technology | Ukraine |
|  | Uzstandard (option) Metrology, measurement technology | Uzbekistan |

| 1 | 2 | 3 | 4 |
|---|---|---|---|
| Type | Operating temperatures of the thermocouple |  |  |
| IEC 60584-1 | ASTM E230 |  |  |
| Class 2 | Class 1 | Standard Special |  |
| K | -40 ... +1,200 °C | -40 ... +1,000 °C | 0 ... 1,260 °C |
| J | -40 ...+750 °C | -40 ...+750 °C | 0 ... 760 °C |
| E | -40 ...+900 °C | -40 ...+800 °C | 0 ... 870 °C |
| N | -40 ... +1,200 °C | -40 ... +1,000 °C | 0 ... 1,260 °C |
| T | -40 ...+350 °C | 0 ... 370 °C |  |

### Output signal 4 ... 20 mA, HART ® protocol, FOUNDATION™ Fieldbus and PROFIBUS ® PA

| Transmitter (selectable versions) | Model T16 | Model T32 | Model T53 | Model T91.10 |
|---|---|---|---|---|
| Data sheet | TE 16.01 | TE 32.04 | TE 53.01 | TE 91.01 |
| Output |  |  |  |  |
| ■ 4 ...20mA | x | x |  | x |
| ■ HART ® protocol |  | x |  |  |
| ■ FOUNDATION™ Fieldbus and PROFIBUS ® PA |  |  | x |  |
| Input |  |  |  |  |
| ■ Thermocouples IEC 60584-1 | K, J, E, N,T | K, J, E, N,T | K, J, E, N,T | K, J,T |
| Explosion protection | Optional | Optional | Standard | - |

| Measuring insert length l 5 inmm | Tolerance inmm |
|---|---|
| 75 ... 825 | +2 0 |
| > 825 | +3 0 |

| 1 | 2 | 3 | 4 |
|---|---|---|---|
| Measuring insert diameter Ø d inmm | Index per DIN 43735 | Tolerance inmm |  |
| 3 1) | Standard | 30 | 3 ±0.05 |
| 6 | Standard | 60 | 6 0 -0.1 |
| 8 (6 mmwith sleeve) | Standard | - | 8 0 -0.1 |
| 8 | Standard | 80 | 8 0 -0.1 |
| 1/8 inch (3.17 mm) 1) 1/4 inch (6.35 mm) 3/8 inch (9.53 mm) | Option, on request | - | - |

| Certification type | Measurement accuracy | Material certificate |
|---|---|---|
| 2.2 test report | x | x |
| 3.1 inspection certificate | x | x |
| DKD/DAkkS calibration certificate | x | - |

## Documents

- [ds_te6511_en_co_d865ae](https://avigan.com.ua/wika/docs/wika/ds_te6511_en_co.pdf) — PDF, 722 KB, en