## Overview

Aim – shoot – test For the calibration of infrared thermometers above ambient temperature, the model CTI5000 is suitable up to 500 °C [932 °F]. Due to its ...

## Applications

- Testing and calibration of non-contact temperature measuring instruments
- Heating source for works laboratories for the calibration of infrared thermometers
- Also suitable for on-site calibration

## Special features

- Testing of infrared thermometers across a wide temperature range; 50 ... 500 °C [122 ... 932 °F]
- Large surface area of the black body (∅ 57 mm [2.24 in]) for all types of pyrometers
- Bore to accept a reference sensor
- Low weight and compact design
- Simple operation

## Specifications

| Parameter | Value |
|---|---|
| model | CTI5000 |
| product family | CT |
| short description | Portable infrared calibrator for testing and calibration of non-contact temperature measuring instruments with a temperature range of 50 to 500 °C and fixed emissivity of 0.95. |
| subcategory | infrared calibrator |
| certifications | CE; EAC; RoHS |
| ambient temp | 5 … 35 °C |
| weight kg | 2.1 |
| languages in datasheet | en |
| calibration type | temperature_bath |
| temperature ranges | 50 … 500 °C |
| accuracy or uncertainty | ±0.1 K to 100 °C, ±0.2 K to 350 °C, ±0.4 K to 500 °C |
| portable | true |

## Datasheet tables

### Infrared calibrator

| Indication | 2 |
|---|---|
| Application range | 50 …500 °C [122 ... 932 °F] |
| Stability | ±0.1 K to 100 °C [212 °F] ±0.2 K to 350 °C [662 °F] ±0.4 K to 500 °C [932 °F] |
| Resolution | 0.1 K |
| Temperature control |  |
| Heating time | 30 min fromT amb. to 500 °C [T amb. to 932 °F] |
| Cooling time | 40 min from 500 °C to 100 °C [from 932 °F to 212 °F] |
| Stabilisation time | approx. 15 ... 20 min |
| Measuring surface |  |
| Emissivity | 0.95 |
| Diameter of the measuring surface | 57 mm[2.24 in] |
| Bore for temperature probe | suitable for 3 mm[0.12 in] probes with an immersion depth of 68 mm[2.68 in] |
| Permissible ambient conditions |  |
| Relative humidity | 35 ... 85 %r.h. (non-condensing) |
| Operating temperature | 5 ... 35 °C [41 ... 95 °F] |
| Voltage supply |  |
| Power supply | AC 230 V, 50/60 Hz |
| Power consumption | 275VA |
| Case |  |
| Dimensions [W x D x H] | 190 x 248 x 113 mm[7.5 x 9.8 x 4.5 in] |
| Weight | 2.1 kg [4.6 lbs] |

| Logo | Description | Country |
|---|---|---|
|  | EU declaration of conformity ■ EMCdirective ■ Low voltage directive ■ RoHS directive | European Union |
|  | EAC ■ EMCdirective ■ Low voltage directive | Eurasian Economic Community |
| - | MTSCHS Permission for commissioning | Kazakhstan |
|  | BelGIM Metrology, measurement technology | Belarus |

## Documents