# Temperature dry-well calibrator — CTD9350

> Temperature dry-well calibrator premium version with integrated measuring instrument, suitable for laboratories, workshops, power plants, and chemical industry applications.

- Catalog code: ctd9350
- Model: CTD9350
- Category: Calibration technology › Portable temperature calibrators
- Product page: https://avigan.com.ua/wika/en/products/ctd9350
- Price: on request
- Supplier: AVIGAN (TOV VIKA Prylad), official WIKA partner in Ukraine

## Overview

Range of applications Whether in laboratories, workshops or on-site, the CTD9350 series of temperature dry-well calibrators can meet any calibration ...

## Applications

- Bio and pharmaceutical industries
- Food industry
- Power plants and plant construction
- Measurement and control laboratories in the chemical industry
- Demanding calibrations in production and laboratory

## Special features

- Easy operation via intuitive, user-friendly menus
- Large, easy-to-read touchscreen
- Short response times due to optimised control
- Improved accuracy due to homogeneous dry-well temperature
- Creation of calibration tasks incl. preparation of a certificate

## Specifications

| Parameter | Value |
|---|---|
| model | CTD9350-165 |
| product family | CTD9350 |
| short description | Temperature dry-well calibrator premium version with integrated measuring instrument, suitable for laboratories, workshops, power plants, and chemical industry applications. |
| subcategory | dry-well |
| certifications | RoHS; EMC EN 61326; Low voltage EN 61010; EMC directive EN 61326; Low voltage directive EN 61010; CE |
| ambient temp | 0 … 50 °C |
| ip rating | IP20 |
| weight kg | 13.5 |
| languages in datasheet | en; de |
| calibration type | dry_block |
| temperature ranges | -35 … 165 °C; -35 … 700 °C |
| accuracy or uncertainty | ±0.075 % of reading, min. ±0.3 K |
| traceability | DAkkS |
| output signal | mA; V; resistance; thermoelectric voltage; current (0/4...20 mA); RTD simulation; TC simulation; Pt100; Pt1000; thermocouple |
| portable | true |

## Datasheet tables

| Basic information | CTD9350-700 | CTD9350-165 |
|---|---|---|
| Temperature range | T amb ... 700 °C [T amb ... 1.292 °F] | -35 ...+165 °C [-31 ...+329 °F] |
| Accuracy | → See table &#x27;Accuracy specifications&#x27; |  |
| Temperature stability 1) | → See table &#x27;Accuracy specifications&#x27; |  |
| Metal block |  |  |
| Immersion depth | 145 mm[5,71 in] |  |
| Insert dimensions | Ø29 x 150 mm[Ø1,14 x 5,91 in] | Ø28 x 150 mm[Ø1,10 x 5,91 in] |
| Dry-well material | Aluminium-Bronze 2.0966 | Brass 2.0375 |
| Functions |  |  |
| Menu functions | ■ Calibration without certificate ■ Calibration with certificate ■ Remote control ■ Data export to a USB stick |  |
| User settings | User-defined data is indicated on the test certificate |  |
| Dimensions (W x D x H) |  |  |
| Calibrator without carrying handle | 210 x 300 x 330mm [8.27 x 11.81 x 12.99 in] | 210 x 300 x 380mm [8.27 x 11.81 x 14.96 in] |
| Height of carrying handle | 50 mm[1,97 in] |  |
| Weight | 10 kg [22.1 lbs] | Approx. 13.5 kg [29.77 lbs] |

| Digital display instrument | CTD9350-700 | CTD9350-165 |
|---|---|---|
| Display | Bright colour touchscreen (7"), laminated safety glass |  |
| Display range | 0 ... 700 °C [32 ... 1,292 °F] | -50 ...+165 °C [-58 ...+329 °F] |
| Resolution | 0.001 °C |  |
| Units | Adjustable via menu ■ °C ■ °F ■ K |  |
| Menu languages | Adjustable via menu ■ English ■ German |  |

| Accuracy specifications | CTD9350-700 | CTD9350-165 |
|---|---|---|
| Temperature range | T amb ... 700 °C [T amb ... 1,292 °F] | -35 ...+165 °C [-31 ...+329 °F] |
| Accuracy |  |  |
| External reference temperature probe | ±0.100 K 2) | ±0.100 K |
| Internal reference temperature probe | 0.075 %of reading, min. ±0.3 K | ±0.100 K |
| Temperature stability 1) |  |  |
| External reference temperature probe | ±0.020 K 2) | ±0.008 K |
| Internal reference temperature probe | ±0.100 K | ±0.015 K |
| Influence due to loading 1) |  |  |
| External reference temperature probe | ±0.020 K 2) | ±0.010 K |
| Internal reference temperature probe | ±0.300 K | ±0.300 K |
| Temperature distribution 1) |  |  |
| Axial homogeneity | ±0.300 K | ±0.100 K |
| Radial homogeneity | ±0.040 K | ±0.020 K |

| Accuracy specifications | CTD9350-700 | CTD9350-165 |
|---|---|---|
| Hysteresis | ±0.015 K | ±0.010 K |

| 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|
| Temperature control | CTD9350-700 | CTD9350-165 |  |  |
| Heating time | 19 min | from 20 °C to 690 °C [from 68 °F to 1,274 °F] | 14 min | from 20 °C to 165 °C [from 68 °F to 329 °F] |
|  |  |  | 16 min | from -35 °C to +165 °C [from -31 °F to +329 °F] |
| Cooling time | 85 min | from 700 °C to 30 °C [from 1,292 °F to 86 °F] | 13 min | from +20 °C to -30 °C [from +68 °F to -22 °F] |
|  |  |  | 11 min | from +165 °C to 20 °C [from 329 °F to 68 °F] |
| Stabilisation time | Dependent on temperature and temperature probe |  |  |  |

| Electrical connection | CTD9350-700 | CTD9350-165 |
|---|---|---|
| Operating voltage 1) | ■ AC 110 ... 115 V, 60 Hz 2) ■ AC 230 V, 50 Hz 2) | AC 100 ... 240 V, 50/60 Hz |
| Power consumption | 1,000W | 375W |
| Electrical safety | Overvoltage category (installation category) II, Pollution degree 2 in accordance with IEC-61010-1 |  |
| Fuse | 10 AH 250V slow blow fuse | 6.3 AH 250V slow blow fuse |
| Power cord | ■ For Europe ■ For USA/Canada ■ For Switzerland ■ For UK |  |

### Operating conditions

| Place of use | For indoor use only |
|---|---|
| Altitude | To 2,000 m[6,562 ft] above sea level |
| Operating temperature | 0 ... 50 °C [32 ... 122 °F] The ambient temperature influences the heating/cooling behaviour |
| Storage and transport temperature range | -10 ...+60 °C [14 ... 140 °F] |
| Relative humidity, condensation | < 80 %to 31 °C [88 °F], decreasing linearly to 50 %at 40 °C [104 °F] (non-condensing) |
| Mounting position | Upright / vertical standing |

### Communication

| Interface | ■ 3 x USB ■ Ethernet |
|---|---|
| Connectivity | ■ OPC UA ■ Serial communication ■ HTTP |
| Baud rate | 2400 |
| Measuring rate | 1 measured value per second |
| Internal program | Test items, test tasks and test points can be applied without limit |

### Output signal

| 1 | 2 |
|---|---|
| Analogue output |  |
| Voltage supply | DC 24V (can be activated via menu) |
| Load | Max.24mA |
| Switching function | NC,NO |

### Electrical connection

| Number of channels | 2 |
|---|---|
| Resistance thermometer | 2 |
| Thermocouple | 2 |
| Current signal | 1 |
| Voltage signal | 1 |
| Switch test | 2 |
| Connection type |  |
| Resistance thermometer | 4 x 4-mm banana jacks |
| Thermocouple | 2 x thermocouple terminal (mini) |
| Current and voltage signal | 4 mmbanana jacks |
| Pin assignment |  |
| Resistance thermometer | ■ 2-wire connection ■ 3-wire connection ■ 4-wire connection |
| Measuring range |  |
| Resistance thermometer | Pt100 Pt1000 0 ...400Ω 0 ...4,000Ω |
| Thermocouple | -10 ...+100mV |
| Current signal | DC 0 ...24mA |
| Voltage signal | DC 0 ...12V |

| 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|
| Accuracies | Measuring range | Accuracy |  |  |
| Resistance thermometer |  |  |  |  |
| Pt100 | -200 ...+850 °C | [-328 ... +1,562 °F] | ±0.03 °C | [±0.05 °F] |
| Pt500 | -200 ...+850 °C | [-328 ... +1,562 °F] | ±0.12 °C | [±0.22 °F] |
| Pt1000 | -200 ...+850 °C | [-328 ... +1,562 °F] | ±0.06 °C | [±0.11 °F] |
| Ni100 | -60 ...+180 °C | [-76 ...+356 °F] | ±0.02 °C | [±0.04 °F] |
| Ni500 | -60 ...+200 °C | [-76 ...+392 °F] | ±0.08 °C | [±0.14 °F] |
| Ni1000 | -60 ...+200 °C | [-76 ...+392 °F] | ±0.04 °C | [±0.07 °F] |
| Cold junction | -200 ... +1,820 °C | [-328 ... +3,308 °F] | ±0.3 °C | [±0.54 °F] |
| Thermocouple |  |  |  |  |
| Type K | -160 ... +1,260 °C | [-256 ... +2,300 °F] | ±0.08 °C | [±0.14 °F] |
| Type J | -190 ... +1,200 °C | [-310 ... +2,192 °F] | ±0.07 °C | [±0.13 °F] |
| Type N | 0 ... 1,300 °C | [32 ... 2,372 °F] | ±0.13 °C | [±0.23 °F] |
| Type E | -200 ... +1,000 °C | [-328 ... +1,832 °F] | ±0.06 °C | [±0.11 °F] |
| TypeT | -130 ...+400 °C | [-202 ...+752 °F] | ±0.09 °C | [±0.16 °F] |
| Type R | 160 ... 1,760 °C | [320 ... 3,200 °F] | ±0.78 °C | [±1.40 °F] |
| Type S | 170 ... 1,760 °C | [338 ... 3,200 °F] | ±0.73 °C | [±1.31 °F] |
| Type B | 920 ... 1,820 °C | [1,688 ... 3,308 °F] | ±0.5 °C | [±0.90 °F] |

| Accuracies | Measuring range | Accuracy |
|---|---|---|
| Direct current | 0 ...24mA | 0.01 %of end value |
| DC voltage | 0 ...12V | 0.01 %of end value |

| Logo | Description | Region |
|---|---|---|
|  | EU declaration of conformity | European Union |
|  | EMCdirective EN 61326 emission (group 1, class A) and immunity (industrial application) |  |
|  | Low voltage directive EN 61010, safety requirements for electrical equipment for measurement, control and laboratory use |  |
|  | RoHS directive |  |

### Certificates

| 1 | 2 |
|---|---|
| Calibration |  |
| Integrated measuring instrument | ■ Without ■ 3.1 inspection certificate per EN 10204 for PT, TC, mAandV ■ DAkkS calibration certificate for PT, TC, mAandV |
| Reference thermometer | ■ Without ■ 3.1 inspection certificate per EN 10204 up to a maximum of 165 °C [329 °F] ■ DAkkS calibration certificate up to a maximum of 165 °C [329 °F] ■ 3.1 inspection certificate per EN 10204 for the rangeT amb ... 700 °C [T amb ... 1,292 °F] ■ DAkkS calibration certificate for the rangeT amb ... 700 °C [T amb ... 1,292 °F] |
| Calibrator 1) | ■ 3.1 inspection certificate per EN 10204 ■ DAkkS calibration certificate (traceable and accredited in accordance with ISO/IEC 17025) |
| Recommended calibration interval | 1 year (dependent on conditions of use) |

| 1 | 2 | 3 |
|---|---|---|
| Accessories for model CTD9350-165 1) | Order code |  |
| Description | CTX-A-KC |  |
|  | Transport case with trolley rack | -3- |
|  | Exteranl reference probe | -E- |
| Ordering information for your enquiry: |  |  |
| 1. Order code: CTX-A-KC 2. Option: | ⇓ [ ] |  |

| 1 | 2 | 3 |
|---|---|---|
| Inserts for model CTD9350-165 1) | Order code |  |
| Description |  | CTA9I-3R |
|  | Undrilled insert Dimensions:Ø28 x 150 mm[Ø1.1 x 5.91 in] Material: Brass 2.0375 | -N- |

| 1 | 2 | 3 |
|---|---|---|
| Inserts for model CTD9350-165 1) | Order code |  |
| Description | CTA9I-3R |  |
|  | Drilled insert Dimensions:Ø28 x 150 mm[Ø1.1 x 5.91 in] Drilling depth: 145 mm[5.71 in] Material: Brass 2.0375 | - |
|  | For thermometers with diameters up to 1.2 mm[0.05 in] Bore diameter: 1 x 1.5 mm[1 x 0.06 in] | -1- |
|  | For thermometers with diameters up to 1.6 mm[0.06 in] Bore diameter: 1 x 2 mm[1 x 0.08 in] | -2- |
|  | For thermometers with diameters up to 2.7 mm[0.11 in] Bore diameter: 1 x 3 mm[1 x 0.12 in] | -3- |
|  | For thermometers with diameters up to 3.2 mm[0.13 in] Bore diameter: 1 x 3.5 mm[1 x 0.14 in] | -4- |
|  | For thermometers with diameters up to 4.7 mm[0.19 in] Bore diameter: 1 x 5 mm[1 x 0.20 in] | -5- |
|  | For thermometers with diameters up to 6.3 mm[0.25 in] Bore diameter: 1 x 6.5 mm[1 x 0.26 in] | -6- |
|  | For thermometers with diameters up to 7.2 mm[0.28 in] Bore diameter: 1 x 7.5 mm[1 x 0.30 in] | -7- |
|  | For thermometers with diameters up to 8.2 mm[0.32 in] Bore diameter: 1 x 8.5 mm[1 x 0.33 in] | -8- |
|  | For thermometers with diameters up to 9.5 mm[0.37 in] Bore diameter: 1 x 10 mm[1 x 0.39 in] | -9- |
|  | For thermometers with diameters up to 3.2 mm[0.13 in] Bore diameter: 1 x 5.0 mmand 1 x 3.5mm [1 x 0.20 in and 1 x 0.14 in] | -I- |
|  | For thermometers with diameters up to 6.3 mm[0.25 in] Bore diameter: 1 x 5.0 mmand 1 x 6.5mm [1 x 0.20 in and 1 x 0.26 in] | -X- |
|  | Bore diameter: 1 x 3.2 mmand 1 x 6.3mm [1 x 0.13 in and 1 x 0.25 in] | -A- |
|  | Bore diameter: 2 x 3.2 mm, 1 x 4.2 mm, 1 x 6.3 mm, 1 x 8.4 mmand 1 x 9.9mm [2 x 0.13 in, 1 x 0.17 in, 1 x 0.25 in, 1 x 0.33 in and 1 x 0.39 in] | -B- |
|  | Customer-specific, special probes are possible on request. | -?- |
|  | Insert replacement tool | -J- |
| Ordering information for your enquiry: |  |  |
| 1. Order 2. Option: | ⇓ [ ] |  |

| 1 | 2 | 3 |
|---|---|---|
| Accessories for model CTD9350-700 1) | Order code |  |
| Description | CTX-A-KD |  |
|  | Transport case with trolley rack | -3- |
| Ordering information for your enquiry: |  |  |
|  | 1. Order code: CTX-A-KDC 2. Option: | ⇓ [ ] |

| 1 | 2 | 3 |
|---|---|---|
| Inserts for model CTD9350-700 1) | Order code |  |
| Description |  | CTA9I-3S |
| Material: | Undrilled insert Dimensions:Ø29 x 150 mm[Ø1.14 x 5.91 in] Aluminium-Bronze 2.0966 | -N- |
| Drilling Material: For thermometers with diameters | Drilled insert Dimensions:Ø29 x 150 mm[Ø1.14 x 5.91 in] depth: 145 mm[5.71 in] Aluminium-Bronze 2.0966 up to 1.2 mm[0.05 in] | -1- |
| 1 x 1.5 mm[1 x 0.06 in] For Bore 1 x 2 | thermometers with diameters up to 1.6 mm[0.06 in] diameter: mm[1 x 0.08 in] | -2- |
| For Bore 1 x 3 mm[1 thermometers | thermometers with diameters up to 2.7 mm[0.11 in] diameter: x 0.12 in] | -3- |
| For with diameters Bore diameter: 1 x 3.5 mm[1 x 0.14 in] | up to 3.2 mm[0.13 in] | -4- |
| For thermometers with diameters Bore diameter: 1 x 5 mm[1 x 0.20 in] | up to 4.7 mm[0.19 in] | -5- |
| For thermometers with diameters up to 6.3 mm[0.25 Bore diameter: 1 x 6.5 mm[1 x 0.26 in] | in] -6- |  |
| For thermometers with diameters up to 7.2 mm[0.28 Bore diameter: 1 x 7.5 mm[1 x 0.30 in] | in] | -7- |
| For thermometers with diameters up to 8.2 Bore diameter: 1 x 8.5 mm[1 x 0.33 in] | mm[0.32 in] | -8- |
| For thermometers with diameters up to 9.5 mm[0.37 Bore diameter: 1 x 10 mm[1 x 0.39 in] | in] | -9- |
| Bore 1 x 5.0 [1 x 0.20 in | For thermometers with diameters up to 3.2 mm[0.13 in] diameter: mmand 1 x 3.5mm and 1 x 0.14 in] | -A- |

| 1 | 2 | 3 |
|---|---|---|
| Inserts for model CTD9350-700 1) | Order code |  |
| Description | CTA9I-3S |  |
|  | For thermometers with diameters up to 6.3 mm[0.25 in] Bore diameter: 1 x 5.0 mmand 1 x 6.5mm [1 x 0.20 in and 1 x 0.26 in] | -X- |
|  | Bore diameter: 1 x 3.2 mmand 1 x 6.3mm [1 x 0.13 in and 1 x 0.25 in] | -A- |
|  | Bore diameter: 2 x 3.2 mm, 1 x 4.2 mm, 1 x 6.3 mm, 1 x 8.4 mmand 1 x 9.9mm [2 x 0.13 in, 1 x 0.17 in, 1 x 0.25 in, 1 x 0.33 in and 1 x 0.39 in] | -B- |
|  | Customer-specific, special probes are possible on request. | -?- |
|  | Insert replacement tool | -J- |
| Ordering information for your enquiry: |  |  |
| 1. Order code: 2. Option: | ⇓ [ ] |  |

## Documents

- [ds_ct4139_en_co_a6704c](https://avigan.com.ua/wika/docs/wika/ds_ct4139_en_co.pdf) — PDF, 753 KB, en

## Configurator

This product has an interactive configurator on its product page.

```
CTD9350- J -ZZ
```