# Pressure controller — CPC7000

> Pneumatic high-pressure version

- Catalog code: cpc7000
- Model: CPC7000
- Category: Calibration technology › Pressure controllers
- Product page: https://avigan.com.ua/wika/en/products/cpc7000
- Price: on request
- Supplier: AVIGAN (TOV VIKA Prylad), official WIKA partner in Ukraine

## Overview

Version The CPC7000 pneumatic high-pressure controller always provides a suitable calibration solution on account of its accuracy class. Its outstanding ...

## Applications

- Automotive and avionics industry
- Industry (laboratory, workshop and production)
- Transmitter and pressure gauge manufacturers
- Calibration service companies and service industry
- Research and development laboratories

## Special features

- Pressure ranges: 0 ... 700 bar [0 ... 10,000 psi]
- Control speed 30 s
- Control stability 0.008 % FS
- Accuracy down to 0.01 % IS (IntelliScale)

## Specifications

| Parameter | Value |
|---|---|
| model | CPC7000 |
| product family | CPC |
| short description | Pneumatic high-pressure controller with touchscreen interface, autoranging capability, and high accuracy for calibration of pressure measuring instruments. |
| subcategory | high-pressure controller |
| certifications | CE; RoHS; EAC; GOST; BelGIM |
| ambient temp | 15 … 45 °C |
| ip rating | None |
| weight kg | 40 |
| languages in datasheet | en; de; es; fr; it; nl; pl; pt; ro; sv; ru; el; ja; zh |
| calibration type | electronic |
| pressure ranges | 100 … 400 bar; 400 … 700 bar; 101 … 401 bar; 401 … 701 bar |
| temperature ranges | 0 … 70 °C |
| accuracy or uncertainty | 0.01 % FS or 0.01 % IS-50 (depending on option) |
| traceability | A2LA; DKD; DAkkS |
| output signal | pressure |
| medium | nitrogen |
| portable | false |

## Datasheet tables

### Reference pressure sensors

| Pressure range | CPR8000 | CPR8050 |
|---|---|---|
| Accuracy 1) | Standard:0.01%FS 2) Optional: 0.01 %IS-50 3) | 0.01%FS 2) |
| Gauge pressure | 100 ... 400 bar 4) 1,500 ... 6,000 psi 4) | 400 ... 700 bar 6,000 ... 10,000 psi |
| Absolute pressure | 101 ... 401 bar 4) 1,515 ... 6,015 psi 4) | 401 ... 701 bar 6,015 ... 10,015 psi |
| Calibration interval | 365 days | 365 days |
| Optional barometric reference |  |  |
| Function | The barometric reference can be used to switch pressure types 5) , absolute <=> gauge.With gauge pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry out an absolute pressure emulation. |  |
| Measuring range | 552 ... 1,172 mbar abs. (8 ... 17 psi abs.) |  |
| Accuracy 1) | 0.01 %of reading |  |
| Pressure units | 38 and two freely programmable |  |

| Basic instrument | 2 |
|---|---|
| Instrument |  |
| Instrument version | Standard: desktop case Option: 19" rack-mounting kit |
| Dimensions | see technical drawings |
| Weight | approx. 40 kg (88.2 lbs) incl. all internal options |
| Warm-up time | approx. 15 min |
| Display |  |
| Screen | 8.9" colour LC display with resistive touchscreen |
| Resolution | 4 ... 6 digits depending on range and units |
| Connections |  |
| Pressure connection | Standard: 5 ports with M16 x 1.5 female with sealing cone Optional: 2 ports with 9/16-18 UNF female Optional barometer:M12 x 1 female |
| Filter elements | All pressure ports have 20-µm filters. |
| Permissible pressure media | Nitrogen class 2.8 or better |
| Overpressure protection | Burst disc up to 1,000 bar (14,500 psi) |

### Basic instrument

| 1 | 2 |
|---|---|
| Permissible pressure |  |
| Supply port | 107 …110 %FSat least 30 …50bar (435 ... 725 psi) over nominal pressure, whichever is greater |
| Measure/Control port | max.105%FS |
| Voltage supply |  |
| Power supply | AC 100 ... 120 / 200 ... 240 V, 50/60 Hz |
| Power consumption | max.140VA |
| Permissible ambient conditions |  |
| Storage temperature | 0 ... 70 °C (32 ... 158 °F) |
| Humidity | 5 …95%r.h.(non-condensing) |
| Compensated temperature range | 15 ... 45 °C (59 ... 113 °F) |
| Mounting position | horizontal |
| Control parameters |  |
| Control stability | 0.008 %FSin a 25 ml test volume |
| Control time | 30 s 6) |
| Control range | 0 ...100%FS |
| Minimum control pressure | 1 bar (14.5 psi) |
| Test volume | 0 ... 50 ccm |
| Communication |  |
| Interface | Ethernet, IEEE-488, USB, RS-232 |
| Command sets | Mensor,WIKA SCPI |
| Response time | approx. 100 ms |
| Internal program | up to 24 sequences with up to 99 steps each |

| Logo | Description | Country |
|---|---|---|
|  | EU declaration of conformity ■ EMCdirective 7) EN 61326-1 emission (group 1, class A) and immunity (industrial application) ■ Low voltage directive ■ RoHS directive | European Union |
|  | EAC (option) ■ Pressure equipment directive ■ Low voltage directive | Eurasian Economic Community |
|  | GOST (option) Metrology, measurement technology | Russia |
|  | BelGIM (option) Metrology, measurement technology | Belarus |

### Certificate

| Calibration 8) | Standard:A2LA calibration certificate Option:DKD/DAkkS calibration certificate |
|---|---|
| Recommended recalibration interval | 1 year (dependent on conditions of use) |

| Cal-Template (light version) | Cal-Template (full version) | Log-Template (full version) |
|---|---|---|
| ■ Semi-automatic calibration | ■ Fully automatic calibration | ■ Live measurement recording for a certain period of time with selectable interval, duration and start time |
| ■ Calibration certificate creation 3.1 per DIN EN 10204 ■ Calibration reports can be exported to Excel ® template or XML file ■ Calibration of gauge pressure measuring instruments with absolute pressure references and vice versa ■ Generation of calibration certificates with no limitations on measuring points |  |  |

### Ordering information for your enquiry for a single license

| WIKA-CAL-LZ-Z-Z | WIKA-CAL-CZ-Z-Z | WIKA-CAL-ZZ-L-Z |
|---|---|---|
| Ordering information for your enquiry for a pair license: |  |  |
| Cal-Template (light version) together with Log-Template (full version) | WIKA-CAL-LZ-L-Z |  |
| Cal-Template (full version) together with Log-Template (full version) | WIKA-CAL-CZ-L-Z |  |

| Accessories for CPC7000 | 2 | Order code |
|---|---|---|
|  | Description | CPX-A-C7 |
|  | 19" rack-mounting kit | -2- |
|  | Barometric reference Measuring range 552 ... 1.172 mbar abs. (8 ... 17 psi abs.) Accuracy to 0.01 %of reading | -3- |
|  | for barometric reference, voltage supply and software | -5- |
|  | Calibration adapter and voltage supply with interface cable RS-232 | -6- |
|  | Adapterset M12 on 9/16-18 UNF female | -1- |
| Ordering information for your enquiry: |  |  |
| 1. Order code: CPX-A-C5 2. Option: | ⇓ [ ] |  |

## Documents

- [ds_ct2763_en_co_00085d](https://avigan.com.ua/wika/docs/wika/ds_ct2763_en_co.pdf) — PDF, 681 KB, en