SiemensETIJohnson ControlsIFMSICKWIKABosch RexrothELKO EPHYUNDAIABBVEGAKSR KUEBLERAlfa Laval

P399 Electronic Pressure Transducer

Вид документаБюлетень виробу
ВиробникJCI
Код документаp399-electronic-pressure-transducer
Група обладнанняP399 Electronic Pressure Transducer
Сторінок1
Мова документаанглійська
Розмір файлу0.0 МБ

Текст документа

Code No. LIT-1900018
                                                                                                                                                  Issued July 1, 2006

P399
Electronic Pressure Transducer
Description                                                     Pressure connections are available in one of
                                                                two standard styles:
The P399 Series Electronic Pressure
Transducer is a compact, economical, rugged,                    •  1/4 in. SAE female flare fitting (with
direct-mount pressure transducer designed to                       Schrader valve depressor)
produce a low voltage analog signal                             • 1/8 in. NPT male fitting
(0.5 to 4.5 VDC nominal ) based on the                          The wiring harness is purchased separately or
sensed pressure. They are designed for use                      included with the P399xxx-1-200C.
in commercial refrigeration and air-
conditioning applications.                                      Features
The P399 Transducer features welded                                                                                                                P399 Series
                                                                •   affordable
                                                                                                                                        Electronic Pressure Transducers
stainless steel construction with                               •   rugged stainless steel construction
environmentally sealed electronics. They are                        provides a durable assembly and
resistant to the effects of wide temperature                        eliminates potential refrigerant loss due to                 •   reliable, repeatable performance and long
swings, high humidity, condensation, and                            O-ring failures - resists Electromagnetic                        operating life minimizes service and
icing. They are suitable for use with all non-                      Interference (EMI) and damage due to                             replacement costs
corrosive refrigerants as well as ammonia.                          shock, vibration, and pressure pulsations                    •   available in several pressure ranges (up to
P399 Transducers are available in several                       •   environmentally sealed electronics                               750 psis) to provide a single line of
pressure ranges (up to 750 psis), covering                          withstand the effects of adverse conditions                      transducers for all refrigeration and air
most common refrigeration and air                                   found in typical equipment rooms                                 conditioning application needs
conditioning applications.

Selection Chart                                                                                  Accessories
Code Number             Description                                                              Code Number             Description
P399AAA-1C              Pressure Transducer, 0-100 psis, 1/8 in. NPTM pressure                   WHA-P399-200C           Wire Harness, 6 1/2 ft (2.0 m.) length with pigtails
                        connection                                                               WHA-P399-400C           Wire Harness, 13 ft (4.0 m.) length with pigtails
P399AAC-1C              Pressure Transducer, 0-100 psis, 1/4 in. SAE female flare                WHA-P399-600C           Wire Harness, 19-5/8 ft (6.0 m.) length with pigtails
                        pressure connection with valve depressor
P399BAA-1C              Pressure Transducer, 0-500 psis, 1/8 in. NPTM pressure                   Note:     Wiring harness must be ordered separately.
                        connection
P399BAC-1C              Pressure Transducer, 0-500 psis, 1/4 in. SAE female flare                Technical Specifications
                        pressure connection with valve depressor
                                                                                                                     P399 Electronic Pressure Transducer
P399CAA-1C              Pressure Transducer, 0-750 psis, 1/8 in. NPTM pressure
                        connection                                                               Overpressure                             3x upper range limit, short term
P399CAC-1C              Pressure Transducer, 0-750 psis, 1/4 in. SAE female flare                Burst Pressure                           5x upper range limit
                        pressure connection with valve depressor                                 Vacuum                                   30 microns max (0.03 mm Hg) short term
P399GAA-1C1             Pressure Transducer, 0-200 psis, 1/8 in. NPTM pressure                   Media Compatibility                      Noncorrosive refrigerants, refrigeration
                        connection                                                                                                        lubricating oils, ammonia
P399AAA-1-200C          Pressure Transducer, 0-100 psis, 1/8 in. NPTM pressure                   Supply Voltage                           5.00 ± 0.25 VDC ratiometric range,
                        connection, with 6-1/2 ft (2m) wire harness with pigtails                                                         6.0 VDC maximum
P399AAC-1-200C          Pressure Transducer, 0-100 psis, 1/4 in. SAE female flare                Current Draw                             5 mA maximum @ 5.0 VDC
                        pressure connection with valve depressor, with 6-1/2 ft                  Output Signal                            Ratiometric, 10% to 90% of supply voltage
                        (2m) wire harness with pigtails
                                                                                                                                          (0.5 to 4.5 VDC nominal)
P399BAA-1-200C          Pressure Transducer, 0-500 psis, 1/8 in. NPTM pressure
                                                                                                 Pressure Connections                     1/8 in. NPT male, direct mount
                        connection, with 6-1/2 ft (2m) wire harness with pigtails                                                         1/4 in. SAE female flare with Schrader valve
P399BAC-1-200C          Pressure Transducer, 0-500 psis, 1/4 in. SAE female flare                                                         depressor, direct mount
                        pressure connection with valve depressor, with 6-1/2 ft
                                                                                                 Temperature Storage                      -40 to 257oF (-40 to 125oC)
                        (2m) wire harness with pigtails                                          and Humidity Ambient                     -40 to 158oF (-40 to 70oC)
P399CAA-1-200C          Pressure Transducer, 0-750 psis, 1/8 in. NPTM pressure
                        connection, with 6-1/2 ft (2m) wire harness with pigtails                                   Refrigerant           -40 to 300oF (-40 to 149oC); 0 to 100% RH
P399CAC-1-200C          Pressure Transducer, 0-750 psis, 1/4 in. SAE female flare                Accuracy                                 ±1.5% Span, -4 to 212oF (-20 to 70oC);
                        pressure connection with valve depressor, with 6-1/2 ft                                                           ±2.0% Span, -40 to -4oF (-40 to -20oC)
                        (2m) wire harness with pigtails                                          Materials          Connector             Nylon, 30% glass filled
1. Not applicable to System 350™ Pressure Controls                                                                  Main Body             Stainless steel all welded construction
                                                                                                                                          (304L)
                                                                                                 Vibration                                10G, 20-2000 Hz maximum

The performance specifications are nominal and conform to acceptable industry standards. For applications at conditions beyond these specifications, consult the local Johnson Controls office.
Johnson Controls, Inc. shall not be liable for damages resulting from misapplication or misuse of its products. © 2006 Johnson Controls, Inc.                  www.johnsoncontrols.com

                                                                                       1 of 1

Позиції каталогу з цим документом

Документ виробника, опублікований як довідковий матеріал на обладнання, яке ми постачаємо. Правовласник може попросити прибрати його — приберемо.